We investigate how to improve the SNR of vessel-selective ASL images by off-resonance compensation and label rotation scheme optimization.
SSPCASL works by creating a set of labeling bands within the labeling plane by inclusion of additional gradients between RF pulses. These bands are rotated about the target vessel to define a small labeling region. The width of the labeling bands and the rotation scheme were modified to maximize labeling efficiency.
Bloch Simulations of a SSPCASL sequence were conducted to calculate a spatial map of inversion efficiency (“inversion profile”). Simulations assumed constant velocity field of 24 cm/s, 71x71 matrix, and vessel-selective gradient moment of 0.2 G/cm. Inversion profiles were compared for a range of rotation schemes.
In order to validate the simulations, the following experiment was performed on a 3T MR750 scanner (GE, Waukesha,WI). A simple flow phantom consisting of tygon tubes (diameter = 6 mm), was connected to a peristaltic pump in a closed-loop . Average flow velocity was 24 cm/s. Two tubes were setup in parallel, separated by 20 mm . SSPCASL images were collected with a 2D spiral acquisition (128x128 matrix, slice thickness 7 mm , FOV 20 cm), amplitude of labeling rotation gradients = 0.22 G/cm, labeling duration = 200 ms , post-labeling delay = 50 ms. After positioning the label spot on one of the two parallel tubes, several rotation schemes were tested and compared. A 3x3 region of interest (ROI) is assigned in the center of each vessels and the degree of inversion was measured.
The position of left and right carotid arteries were determined from TOF images in two human participants. SSPCASL images were collected with the following scanning parameters: 3D spiral acquisition, vessel selective rotation gradient amplitude 0.24 = G/cm, number of rotations = 64, matrix size = 64x64x20 , FOV = 24 cm, slice thickness = 6 mm labeling duration = 1400 ms, post-labeling delay = 1800 ms, background-suppression pulses. In order to test for off-resonance artefacts 3, SSPCASL images were collected using a range of phase corrections (0 – PI radians) to maximize inversion efficiency empirically. Optimal phase correction was then chosen for each carotid artery and used to collect each vascular territory map. Non-selective pCASL were also collected. Inversion efficiency was calculated relative to non-selective pCASL.
Simulation and experimental results showed that the inversion efficiency increases with the number of steps per rotation. However, the size of the labeling area also increases, decreasing the selectivity of the method. A smaller number of rotation steps should be chosen to optimize the selectivity profile. We are currently using the framework described here to further improve the vessel selectivity profile.
We also demonstrated in human volunteers that the off-resonance in the labeling plane can be estimated with an additional phase-calibration scan as in 3. The optimal phase correction resulted in improved labeling efficiency for SSPCASL, nearly matching the efficiency of non-selective PCASL.
[1] E. C. Wong, “Vessel-encoded arterial spin-labeling using pseudocontinuous tagging,” Magn. Reson. Med., vol. 58, no. 6, pp. 1086–1091, 2007.
[2] M. Helle, D. G. Norris, S. Rüfer, K. Alfke, O. Jansen, and M. J. P. Van Osch, “Superselective pseudocontinuous arterial spin labeling,” Magn. Reson. Med., vol. 64, no. 3, pp. 777–786, 2010.
[3] H. Jahanian, D. C. Noll, and L. Hernandez-Garcia, “B0 field inhomogeneity considerations in pseudo-continuous arterial spin labeling (pCASL): effects on tagging efficiency and correction strategy,” NMR Biomed, vol. 24, no. 10, pp. 1202–1209, 2011.