Li Guo1, Zhongbiao Xu1, Yingjie Mei1,2, Wenxing Fang3, Chenguang Zhao3, Wufan Chen1, Yanqiu Feng1, and Feng Huang4
1Guangdong Provincial Key Laborary of Medical Image Processing, School of Biomedical Engineering, Southern Medical University, Guangzhou, People's Republic of China, 2Philips Healthcare, Guangzhou, People's Republic of China, 3Philips Healthcare (Suzhou), Suzhou, People's Republic of China, 4Neusoft Medical System, Shanghai, People's Republic of China
Synopsis
IRIS corrects the
motion-induced inter-shot phase errors for multi-shot diffusion-weighted
imaging by extracting the phase information from an additional navigator data,
which may cause the distortion mismatch between the image and navigator data. To
solve the distortion mismatch issue in IRIS without using B0 field map, we propose
to extract the coil sensitivities and phase information from navigator data by using
an eigen-analysis scheme. The performance of the proposed method is
demonstrated in both phantom and in vivo data sets.
Purpose
Multi-shot EPI
has been widely used for high spatial
resolution
diffusion-weighted MR imaging. However, motion-induced phase errors among shots
might result in ghost artifacts which degrade the image quality for diagnosis.
MUSE [1] and IRIS [2] have been proposed to correct these motion-induced phase
variations in image space. Since MUSE uses SENSE reconstruction [3] of each shot
to extract the phase information, the applicability of MUSE is limited by the
number of shots due to SENSE g-factor. Compared with MUSE, IRIS has no such
limitation since the phase information is from an additional navigator. Therefore,
IRIS is preferred, especially when a large number of shots are required. However,
IRIS has a potential problem due to the usage of navigator: the distortion of the
image and the navigator data might be mismatched as discussed in IRIS paper. To
solve this issue, original IRIS uses a B0 field map to correct the navigator distortion
which takes extra scan time. Recently, a k-space reconstruction method,
SEPARATE [4], was published to avoid the necessary of B0 field map. However,
the k-space method intrinsically has lower signal-noise-ratio (SNR) than image
domain scheme. The goal of this study is to develop an image domain
reconstruction approach to solve the distortion mismatch issue in IRIS without
using B0 field map, nor decreasing the SNR for multi-shot acquisition.Theory and Methods
An eigen-analysis
based scheme for coil sensitivity maps (CSM) extraction, which is noticed to be
more robust than conventional image space scheme, was recently validated in ESPIRIT
[5]. Inspired by the scheme, two approaches are proposed in this work to
improve IRIS. First, a virtual coil concept is used. Data from different shots
are treated as virtual coil elements with the coil sensitivities modulated by motion-induced
phase errors. Hence, the final image can be generated using typical SENSE reconstruction.
Second, the eigen-analysis scheme is used to extract the modulated CSM that
contain coil sensitivities and inter-shot phase variations from IRIS navigator
data. The proposed method is named as eigen-analysis IRIS (eIRIS).
To validate the proposed
method, both phantom and in vivo data sets were acquired on a Philips Multiva
1.5T scanner (Philips Healthcare, Suzhou, China), using an eight-channel head
coil and a dual refocusing spin-echo multi-shot EPI pulse sequence [2]. The
second 180◦ refocusing
pulse was used to acquire IRIS navigator. For the phantom acquisition, the scan
parameters include: number of shot (NS) =4, SENSE acceleration factor (R) =1, field
of view (FOV) =230×230mm2, voxel size=1.20×1.20×5mm3,
b=1000s/mm2, number
of signal averages (NSA) =1, repetition time/echo time (TR/TE) =2036.4ms/87.4ms
and echo-train=34 for image-echo, TE=135ms and echo-train=28 for navigator-echo.
For the in vivo acquisition, the scan parameters include: NS=4, R=1, FOV=240×240mm2,
voxel size=1.88×1.88×5mm3, b=500s/mm2, NSA=1, echo-train=30,
TR/TE=1500.0ms/81.3ms for image-echo, TE=120.0ms for navigator-echo. For
comparison, we implemented conventional IRIS without B0 correction (cIRIS) and SEPARATE.
Results
Figs.1 and 2 provide
the results of phantom and in vivo data, respectively. The numbers in the
figures are the quantitative measures of SNR within a region of interest (the red
box in each image). They were calculated by dividing the mean of the signal by its standard
deviation. Due to the distortion mismatch, cIRIS results in images with clear
artifacts, as shown by the white arrows at images. SEPARATE avoids the
artifacts in cIRIS, but reduces the SNR and introduces some residual aliasing
artifacts. eIRIS results in the lowest noise/artifact level.Discussion
In this work, an eigen-analysis
scheme is used to extract the modulated CSM to correct phase inconsistency in
IRIS. Compared to cIRIS, eIRIS avoids the requirement of B0 field map, while
results in the artifact-free images when there are distortion mismatch between
navigator and image data. It hints that the eigen-analysis scheme for CSM
extraction may be insensitive to the distortion. Compared to SEPARATE, it results
in lower noise/artifact level due to the inheritance of the advantages of SENSE
by the adoption of virtual coil elements concept. Conclusion
eIRIS solves the
distortion mismatch issue in IRIS without using B0 field map, nor decreasing
the SNR for multi-shot acquisition. Moreover, the CSM in eIRIS is from the
navigator, thus it has the potential to avoid the motion-induced mismatch
between reference scan and imaging data.References
[1] Chen
N-k, et. al. NeuroImage 2013; 72:41-47. [2] Jeong H-k., et. al. MRM 2013; 69:793-802.
[3] Pruessmann K.P., et. al.
MRM 1999; 42:952-962. [4] Ma, Xiaodong, et.
al. ISMRM 2015; p2799. [5] Uecker, M, et. al. MRM 2014; 71:990-1001.Acknowledgements
No acknowledgement found.References
No reference found.