Joshua Yung1, Yao Ding1, Ken-Pin Hwang1, Carlos E Cardenas2,3, Hua Ai1, Michael A Boss4, Thomas L Chenevert5, Clifton Fuller6, and R Jason Stafford1
1Imaging Physics, The University of Texas MD Anderson Cancer Center, Houston, TX, United States, 2Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX, United States, 3The University of Texas Graduate School of Biomedical Sciences, Houston, TX, United States, 4Physical Measurement Laboratory, National Institute of Standards and Technology, Boulder, CO, United States, 5Radiology, University of Michigan, Ann Arbor, MI, United States, 6Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, TX, United States
Synopsis
The purpose of this study was to determine the quantitative
variability of apparent diffusion coefficient values across a large fleet of MR
systems. Using a NIST traceable magnetic resonance imaging diffusion phantom,
imaging was reproducible and the measurements were quantitatively compared to
known values. Significant differences in identical phantoms were not observed,
but uncertainty in the measurements was seen at low apparent diffusion
coefficient values. The same trend was observed when the diffusion phantoms were
imaged in 20 different MR systems. The characterization of ADC variability for these
systems provides an improved quality control for quantitative diffusion weighted
imaging.
Introduction
The purpose of this study was to determine the quantitative
variability of apparent diffusion coefficient (ADC) values across a large fleet
of MR systems.Methods
A NIST traceable, quantitative magnetic resonance imaging
(MRI) diffusion phantom (High Precision Devices, Inc, Boulder, CO) was imaged on 20 different MR systems. The phantom 194 mm in diameter holds 30
mL vials of polymer polyvinylpyrrolidone (PVP) in aqueous solution. The vials
have a range of concentrations of PVP allowing for a range of ADC values.
Increasing concentrations of the polymer cause the ADC values to be reduced. The
13 vials consist of three vials of pure water and two vials each of a PVP
solution at 10, 20, 30, 40 and 50% PVP by mass fraction. To eliminate the
temperature sensitivity of diffusion, prior to scanning, the phantom was filled
with ice water to equilibrate the phantom at 0°C
according to the provided phantom instructions.
Protocols consistent with the sample protocols developed by
QIBA/RSNA were used on 1.5T and 3.0T MR systems from different vendors, GE and
Siemens, and different platforms. The
resulting images were analyzed with QIBA DWI phantom analysis software that
generated ADC maps from b-values of 0, 500, 900, 2000 s/mm2 with a
user-directed region-of-interest (ROI) placement (Figure 1). The ADC values
were compared to the reference values provided by the phantom manufacturer.
With multiple phantoms in the institution, the phantoms were imaged on the same
MR system to determine consistency among phantoms. Phantoms were then imaged in
the different MR systems in the institution. Results
The error of the ADC values for each of the PVP
concentrations for three different phantoms is shown in Figure 2. ADC values
among the three phantoms showed good agreement and were not significantly
different from one another. At 40% PVP concentration or lower, corresponding to
ADC values of 235x10-5 mm2/s and higher, errors in ADC
measurements were .088. The low ADC vials showed larger errors and variation.
When comparing all MR systems, similar trends were observed
(Figure 3). ADC measurements were all similar to one another except for the low
ADC values (high PVP concentration). Also a positive error or overestimation of
the ADC value was found for most MR systems, except for the low ADC vials. Conclusion
The large variation in the 50% PVP concentration vials among
all three phantoms was consistent. In the other vials of PVP concentrations,
the MR systems showed good agreement and low error to one another and among the
different phantoms. Future investigations will be performed to study
differences in magnetic field strength, vendor, MR system models, gradients,
and bore size. With the increasing usage of diffusion weighted imaging in the
clinic, the characterization of ADC variability for MR systems provides an improved
quality control over the MR systems. Acknowledgements
No acknowledgement found.References
1 Diffusion Phantom Model 128, High Precision
Devices, Inc, Boulder, CO