Giorgia Grisot1,2, Joseph M. Mandeville2, and Anastasia Yendiki2
1Harvard-MIT Health Sciences and Technology, Massachusetts Institute of Technology, Cambridge, MA, United States, 2Athinoula A. Martinos Center for Biomedical Imaging, MGH, Charlestown, MA, United States
Synopsis
The time constraints of in vivo diffusion MRI require a
compromise to be made between spatial and angular resolution. Given the lack of
ground truth on the configuration of human brain connections, determining the
optimal operating point along this trade-off is still an open problem. We use
high-SNR ex vivo human data at microscopic resolution to study the effect of
spatial and angular resolution on dMRI tractography accuracy. Our findings show
that voxel size has a much more dramatic effect on tractography reconstruction of
challenging white matter configurations than the number of gradient directions.
Purpose
The time constraints of in vivo diffusion MRI (dMRI) require
a compromise to be made between spatial resolution and angular resolution. The
optimal operating point along this trade-off is not known, and it is likely
dependent on the specific white-matter (WM) architecture of interest. The lack
of ground truth on the configuration of WM bundles in the human brain makes it
challenging to determine optimal acquisition parameters for reconstructing
these bundles with dMRI tractography. Here, we take advantage of the high
signal-to-noise ratio (SNR) that is feasible only ex-vivo with very long scans at
ultra-high field strengths to acquire dMRI data on human brain samples at
microscopic resolutions, allowing us to investigate the limits of dMRI
performance. We present results from a human WM area with multiple crossing and
splitting bundles that we have previously found to cause errors in tractography
when compared to chemical tracing in non-human primates1,2. We image it at 9.4T and show
the effects that reducing the spatial or angular resolution has on tractography
accuracy.Methods
We identify a critical area with crossing and splitting
pathways (Figure 1), which we have found to cause errors in prior studies
comparing dMRI tractography and chemical tracing in macaques2. We image a 3x2x2cm block of this
area from an ex vivo human brain (Figure 1) in a small-bore 9.4T MRI system
with maximum gradient=480mT/m. We collect a dMRI dataset using a 2-shot EPI
sequence with δ=15ms, Δ=19ms, 514 directions, 0.25mm resolution and bmax=40000s/mm2.
From this ultra-high resolution dataset, we generate datasets with varying
number of gradient directions (122, 256, and 514) and spatial resolutions (0.25mm,
0.5mm, 1mm, and 2mm). We fit the generalized q-sampling image (GQI) model3 to
each dataset and perform deterministic tractography4 with 100000 subvoxel seeds
in DSIstudio. Using the two ROIs (ROI1 and ROI2) shown in Figure 1, we select the streamlines we use in our analysis (Figure 3). Then, using additional inclusion and exclusion masks, we isolate, from that subset of streamlines, six unique
bundles (Figure 2) in the original dataset and assess how well they can be
reconstructed as we reduce the spatial and angular resolution.Results
We evaluate
tractography accuracy on each dataset by computing the fraction of incorrect tractography
streamlines. A streamline is deemed “incorrect” if a) it goes through the
inclusion masks for one of the six bundles shown in Fig. 2, but also diverges
outside the corresponding bundle, or b) is not neuroanatomically sound1 (e.g. streamlines
cutting through the putamen instead of splitting into either extreme capsule or
internal capsule). Figure 3 shows the portion of streamlines we used for
this analysis and Figure 4 shows which fraction of those streamlines is deemed incorrect. We assess tractography accuracy by calculating the ratio of the number of incorrect streamlines over the total number of streamlines that go through ROIs 1 and 2 of Figure 1; we perform this analysis across datasets with varying spatial resolution and
angular resolution (Figure 5).Discussion
Our findings show that spatial resolution has a larger
effect on tractography reconstruction than angular resolution. Across all the
datasets with each different number of diffusion directions, an increase in
voxel size resulted in an increase in incorrect streamlines, with a dramatic
accuracy loss at the 2mm resolution that is commonly used in vivo (Fig. 5). Also,
the number of errors generally decreased when adding more directions, and this
gain was substantial at the 2mm resolution but much less so at higher spatial
resolutions. Conclusion
We used ex-vivo human data acquired at an ultra-high field
strength to study the effect of spatial and angular resolution on dMRI
tractography. Our results indicate that voxel size has a much more dramatic
effect than the number of gradient directions on the accuracy of the
reconstruction of a challenging WM area in the human brain. These results are
specific to the crossing and splitting pathways dorsal to the striatum and the
internal capsule, and the analyses methods used here. Our future work will
analyze tractography accuracy for pathways from additional areas, as well as under
varying SNR levels. Ultimately, our goal is to determine whether there is a trade-off
between SNR, spatial and angular resolution that is feasible in vivo and that yields
satisfactory tractography performance.Acknowledgements
This research was carried out in whole or in part at the Athinoula A. Martinos Center for Biomedical Imaging at the Massachusetts General Hospital, and was funded by NIH P50 MH106435-01, R01 EB021265, R01 5RO1MH04557325, NBIB P41EB015896 and the NIH Blueprint for Neuroscience Research (T90DA022759/R90DA023427). This work also involved the use of instrumentation supported by the NIH Shared Instrumentation Grant Program and/or High-End Instrumentation Grant Program; specifically, grant number(s) S10RR016811, S10RR023401, S10RR019307, S10RR019254, S10RR023043.References
1. Schmahmann, J. D.
& Pandya, D. N. Fiber Pathways of the Brain. New York Oxford University
Press 1, (2006).
2. Grisot, G. et al. Optimization of acquisition parameters
for diffusion MRI using chemical tracing. ISMRM proceedings, (2016).
3. Yeh, F. C., Wedeen, V. J. & Tseng, W. Y. I.
Generalized q-sampling imaging. IEEE Trans. Med. Imaging 29, 1626–1635 (2010).
4. Yeh, F. C., Verstynen, T. D., Wang, Y.,
Fernández-Miranda, J. C. & Tseng, W. Y. I. Deterministic diffusion fiber
tracking improved by quantitative anisotropy. PLoS One 8, (2013).