Zarko Celicanin1,2, Alina Giger2,3, Grzegorz Bauman1,2, Philippe Cattin2,3, and Oliver Bieri1,2
1Division of Radiological Physics; Department of Radiology, University of Basel Hospital, Basel, Switzerland, 2Department of Biomedical Engineering, University of Basel, Basel, Switzerland, 3Medical Image Analysis Center, University of Basel, Basel, Switzerland
Synopsis
Treatment
planning relies on accurate organ motion modeling. Temporally-resolved volumetric
imaging (4DMRI) of the lungs using a recently developed ultra fast steady state
free precession sequence was attempted. No artifacts were present in the lungs,
while image stacking was accurate with no significant image sorting issues.
Introduction
Treatment
planning in interventional MRI and radiotherapy relies vitally on accurate
organ motion modeling (1). To this end, different methods have been devised in
the past to acquire temporally-resolved three-dimensional images (4DMRI) of the
moving organs. One of the most sophisticated methods is based on the 2D multi-slice
interleaved acquisitions of data and navigator slices in order to capture even subtle
motion of organs during therapy, organ drifts (2).
Balanced steady-state
free precession (bSSFP) is a promising MR pulse sequence to image organ motion
due to its temporal resolution and high signal-to-noise ratio, while in-flow
effect increases the signal of blood vessels, which is desirable in motion
tracking algorithms. However, imaging with bSSFP-type of sequences in the lungs
is challenging due to banding artifacts, for which riddance short TRs are
required. Recently, a pulse sequence for lung imaging, termed ultra-fast
balanced steady state free precession (ufSSFP) was proposed (3), offering repetition
times close to 1ms, thus artifact-free imaging at 1.5T. Here, the 4DMRI imaging
method is adapted to offer ufSSFP repetition times from which lung motion
models can be derived, to be used in radiotherapy or interventional MRI
treatments of the lung.
Methods
A 4DMRI sequence
was implement on a clinical 1.5T MR system (gradient field maximum amplitude of
40 mT/m and slew rate of 200 mT/(m × ms)). Excitation pulses, gradient
switching patterns, and partial echoes were adjusted to reduce echo and repetition
time as described in (3). The acquisition scheme was interleaved, i.e. data and
navigator slices were acquired one after another as in (2). Data slices were
positioned to cover the whole lungs, while the navigator was placed in the
right part of the lungs to capture the main head-feet motion. Data and
navigator slices were of sagittal orientation.
MR signal was
received using a body coil made of 6 channels, together with spine coils built-in
in the patient table. The adjustment of shim option was set to a default or
tune up. All data analysis and visualization were performed offline.
The following
imaging parameters were used: TE/TR = 0.88/1.5 ms, bandwidth = 1300 Hz/pixel,
flip angle = 34°, number of slices = 52, parallel acceleration factor = 2,
number of measurements = 50, total imaging time = 10 min, field of view = 288 ×
420 mm2, in-plane resolution = 2.19 × 2.19 mm2, imaging
matrix = 132 × 192, slice thickness = 8 mm, acquisition time per image = 124.5
ms resulting in a sampling frequency of 8.03 Hz. In vivo experiments were
performed on three healthy volunteers.Results
Exemplary images
of data and navigator slices are shown in Fig. 1a and 1b, respectively. While
the lungs are without any artifacts, there are still some remaining banding artifacts
in the lower part of the liver and around the heart (blue arrows in a), caused
by the magnetic field inhomogeneities. The horizontal red line in Fig. 1b, i.e.
navigator, provides a visual clue of the
organ motion amplitude.
In Fig. 2, a 3D
stack of images at the end of the expiration are shown in different
orientations.
Fig. 3a and 3b show
3D stacks at the end of the expiration and in the inspiration, respectively.Discussion
The ultra short
echo and repetition time of the ufSSFP provide imaging free of banding
artifacts in the lungs. The signal obtained from the lung tissue using this
method had a high signal-to-noise ratio and contrast-to-noise ratio.
Shortening of
repetition time yielded the reduction of total imaging time per image, which translated
into high sampling frequency of the breathing motion that could potentially increase
accuracy of organ motion models.
Images are
accurately assigned to theirs corresponding respiratory phases using the 2D
navigator, which could be asserted from viewing the 3D stacks from a coronal
and axial view.
Repeated
acquisition of the navigator, which was kept at the same position caused signal
saturation that is visible on the 3D stacks. While not of a significant issue, in
order to avoid it, one could image the lungs segmented as to avoid overlapping
of the navigator and data slices volume.
External
tracking methods could be attempted in future to provide additional or even complete
tracking information, e.g. using a different imaging modality.Conclusion
4DMRI imaging
method was successfully demonstrated in the lungs providing artifact-free
images with high temporal resolution. Image stacking was accurate with no
significant image sorting issues.Acknowledgements
We thank Swiss National Science Foundation (SNF) for financing this research. Grant number: 320030_163330
References
1. Nguyen D et
al. Computerized triplet beam orientation optimization for MRI-guided Co-60
radiotherapy. Med Phys. 2016;43:5667; 2. von Siebenthal M et al. 4D MR imaging
of respiratory organ motion and its variability. Phys Med Biol.
2007;52:1547-1564; 3. Bieri O. Ultra-fast
steady state free precession and its application to in vivo 1H
morphological and functional lung imaging at 1.5 Tesla. Magn Reson Med.
2013;70:657-663.