Glen Morrell1 and Vivian S. Lee1
1Radiology and Imaging Sciences, University of Utah, Salt Lake City, UT, United States
Synopsis
A prospectively gated free-breathing renal BOLD sequence was compared to conventional breath hold renal BOLD. The new sequence gives consistently better SNR than the conventional sequence. RMS error of T2* estimation was investigated as a measure of accuracy. The free-breathing renal BOLD sequence allows flexible tradeoff of imaging time with SNR and can improve the accuracy of renal BOLD T2* estimation.
Purpose
To
evaluate the accuracy of a novel free-breathing prospectively navigated method
of renal BOLD compared to conventional breath-hold renal BOLD. Introduction
Renal
BOLD MRI shows promise for the evaluation of oxygen levels in the kidneys. Hypoxia
is thought to play a central role in the development and progression of chronic
kidney disease. A noninvasive method for
measuring renal oxygen levels would have great scientific and clinical
impact. Renal BOLD results have been
exciting but contradictory, possibly because of limitations of the underlying
MRI methods. The kidneys move with
respiration, limiting human renal BOLD to a single breath hold with limited SNR,
resolution, and coverage. Methods are
needed to break the breath hold barrier to allow a flexible tradeoff between
imaging time and SNR and resolution. We
recently introduced a free-breathing prospectively gated renal BOLD sequence (FB-BOLD)
which can extend imaging time beyond a single breath hold (1). In the current work we evaluate this method
compared to conventional breath hold renal BOLD (BH-BOLD) for the case of
single slice 2D renal BOLD.Methods
The FB-BOLD sequence is illustrated in Figure 1. The sequence runs with constant repetition
time. A slice-selective navigator is
acquired at the beginning of each sequence repetition and analyzed in real time
to bin the image data based on the phase of the respiratory cycle, with
feedback to the running sequence to acquire the appropriate next phase encoding
step. When one bin contains a complete
k-space the sequence is finished. The
resulting BOLD source images are free of respiratory motion artifact. Since data acquisition takes place during
free breathing, there is no constraint on total image time. For 2D single-slice renal BOLD MRI, this
freedom can be used to acquire multiple signal averages, thus increasing
SNR. We recruited 10 healthy volunteers
for this IRB-approved HIPAA-compliant study.
Each subject underwent conventional BH-BOLD with a single coronal image
obtained through the largest cross-section of the kidneys, and FB-BOLD with
identical sequence parameters but longer imaging time to allow four signal
averages. Sequence parameters for this
study included GRE acquisition, matrix size 256x256 over a 360x360mm FOV, 8mm
slice thickness, 300 Hz pixel bandwidth, TE=(5,10,15,20,25,30,35,40)msec,
FA=25, fat saturation, and 1 signal average for breath hold and 4 for FB-BOLD. BH-BOLD TR was 58 msec while the FB-BOLD TR
was 95 msec, reflecting the extra time necessary for navigation and real-time
sequence feedback. Total BH-BOLD imaging
time was 15 sec compared to about 3:00 minutes for FB-BOLD. Efficiency of the FB-BOLD sequence is about
50-60% (i.e. 50-60% of the collected data is used in the final images). Image data was analyzed by manual segmentation
of right and left kidney. SNR and mean
squared error of T2* fitting were quantified for right and left kidney separately.Results
Figure
3 compares SNR between the BH-BOLD and FB-BOLD sequences. The FB-BOLD sequence gives consistently
higher SNR by an average factor of 1.87 for the right kidney and 1.94 for the left kidney. Figure 4 shows the RMS error of T2* estimation
averaged over the kidney. T2* estimation
error for FB-BOLD is 0.65 that of BH-BOLD for the right kidney, 1.25 for the
left kidney.Discussion
The
FB-BOLD sequence shows expected increase in SNR made possible by breaking the
breath hold barrier with free breathing acquisition. Increased SNR is expected to result in more
accurate T2* estimation. Interestingly,
the RMS error of T2* estimation is lower in the right kidney with the FB-BOLD
sequence but increased in the left kidney.
This error gives some indication of the accuracy of T2* estimation but
must be interpreted with caution, since local B0 inhomogeneity can cause the
actual signal decay to deviate significantly from an ideal exponential (2). The
difference between right and left kidney likely reflects the placement of the
navigator plane across the right kidney and diaphragm.Conclusion
The
free-breathing prospectively navigated renal BOLD sequence achieves expected
improvements in SNR compared to a conventional breath hold renal BOLD sequence. Navigation improves the accuracy of T2*
estimation of the kidney nearest the navigator but may not improve T2* estimation
of the contralateral kidney.Acknowledgements
This work was supported by NIH/NIDDK grant R01 DK063183References
1. Morrell G et al.:
Prospectively navigated multi-echo GRE sequence for improved 2D BOLD imaging of
the kidneys. 21st Meeting ISMRM. Salt Lake City, 2013 pp 1569.
2. Hernando D et al.:
R*(2) mapping in the presence of macroscopic B(0) field variations. Magn Reson Med, 68: 830-840, 2012.