In this study, we demonstrate the feasibility of cardiac cine MRI at 0.35T using bSSFP and compared the results with bSSFP at 1.5T.
Cardiac cine images were acquired in 6 healthy volunteers at
both 0.35T and 1.5T using a bSSFP cine sequence. Images were acquired for a
range of flip angles (FA) from 50-130°. All parameters were identical at both
field strengths (TR/TE=4/2ms, voxel size 1.25x1.25x7mm3, FOV=320mm,
readout bandwidth 780 Hz/pixel, parallel imaging: GRAPPA). SNR of the
blood-pool and myocardium was calculated from the ratio of the average tissue
signal intensity to the standard deviation of a background region outside the
body. Blood-myocardium contrast-to-noise ratio was calculated from the
blood-pool and myocardium SNR difference. SNR and CNR values at 0.35T were
corrected using the correction factor described below. The k-space data were
all fully sampled and no parallel imaging was used.
Our cardiac MRI was carried out on the ViewRay 0.35T system and on a 1.5T Siemens Avanto system (Siemens Medical Systems, Erlangen, Germany). Because the 0.35T system uses surface receive coils that have been modified to be compatible with the radiotherapy system, we sought to quantify any possible influence on SNR due to coil differences between the 0.35T and 1.5T systems. SNR was measured on both systems in an ACR phantom using a proton-density-weighted FLASH sequence (parameters: TR/TE: 500/3 ms, flip angle (FA): 15°, readout bandwidth: 610 Hz/pixel, voxel size: 1.9x1.9x10mm3) and 30 measurements were acquired. An SNR map was generated by calculating the ratio of the average intensity to the standard deviation over the 30 images on a voxel-by-voxel basis. An average SNR value was then calculated over the entire phantom for each system. If coil performance is identical between the two systems, then SNR ratio of the two systems would be the ratio of the field strengths: 1.5/0.35 = 4.29. However, if the coil performance is not identical, then the SNR values at 0.35T would have to be corrected by a correction factor given by:
Correction factor = (SNR1.5T x 0.35) / (SNR0.35T x 1.5)
The SNR in the ACR phantom was 119.3 at 1.5T and 16.4 at 0.35T, giving a ratio of 7.26 instead of 4.29. Therefore, the SNR values at 0.35T was corrected by a factor of 1.69.
Figure 1 top row shows bSSFP cine images acquired on the 0.35T system for FAs from 50° to 130°. For comparison with cine images at 1.5T, Figure 1 bottom row shows cine images at 1.5T for FAs 50° to 90°. A bSSFP cardiac cine acquisition with FA >90° is generally not possible at 1.5T due to SAR limits. It appears that an FA of about 90° to 110° at 0.35T has improved blood-myocardium contrast and general image quality compared to lower flip angles. Figure 2 shows cine images of multiple cardiac phases at both field strengths. In general, while the 0.35T images appear grainier due to reduced SNR, the left ventricle myocardium walls can be clearly delineated with good contrast. Figure 3 shows the blood-pool SNR, myocardium SNR and blood-myocardium CNR at multiple flip angles in both systems.
1. Wu et al., PLOS one, May 2016, doi:10.1371/journal.pone.0154711
2. Rutt et al., JMRI, 1996, 1:57-62