Jun Chen1, Phillip J Rossman1, Kevin J Glaser1, and Richard L Ehman1
1Radiology, Mayo Clinic, Rochester, MN, United States
Synopsis
A
suitable quality assurance (QA) phantom and image acquisition and processing
procedures were developed for confirming the proper function and longitudinal stability of
these MRE systems. The goals of this educational poster are to demonstrate the
QA phantom, the longitudinal testing protocol, and the utility of detecting a problem
caused by discontinuous motion.
Background
To date, commercially
available, FDA-cleared, MR Elastography (MRE) systems have been installed on
approximately 800 MRI units around the world for clinical applications and
research activities. The major subsystems required to perform hepatic MRE are 1)
an acoustic driver device to generate vibrations in the liver, 2) an imaging
sequence capable of recording the tissue vibrations into the phase of MR images
(“wave images”), and 3) an algorithm to process the wave images to generate images
of tissue elasticity (“elastograms”). While the imaging sequence and inversion
algorithm subsystems do not change once installed, human error and hardware
problems can prevent motion from being produced or can cause discontinuous
motion in the acoustic driver subsystem, which alters the measured wave
information in ways that can corrupt the elastograms. We have developed a
suitable quality assurance (QA) phantom and image acquisition and processing
procedures for confirming the proper function and longitudinal stability of
these MRE systems. The goals of this educational poster are to demonstrate the
QA phantom, the longitudinal testing protocol, and the utility of detecting a problem
caused by discontinuous motion.Method: QA Protocol
Phantom Setup:
The MRE QA phantom is made of
Polyvinyl Chloride (PVC) in a 12.5-cm high,15.5-cm diameter cylinder container
with a lid (Figure 1). The MRE QA phantom setup uses a standard patient
liver MRE passive driver, the standard elastic belt used to secure the driver
to the patient, a phantom-specific friction cloth to prevent slipping between
the driver and the phantom, and a standard torso RF coil array (Figure 1).
Imaging Parameters:
The phantom imaging
parameters have been optimized for GE, Siemens and Philips MRI systems according
to the phantom relaxation times, chemical spectrum and geometry. The imaging
parameters for each vendor are available upon request. For example, on a 1.5-T
whole-body MR imaging system (GE Signa HDxt, GE Healthcare, Waukesha, WI), the
imaging parameters of the GRE MRE sequence is as follows: coronal imaging, slice
thickness = 10 mm, FOV = 20 cm, acquisition matrix = 256x64, TE = 18.6 ms, TR
= 50 ms, flip angle = 25º, bandwidth =
31.25 kHz, parallel imaging acceleration factor = 1, choose the middle peak of the
spectrum during prescan, MRE active driver power = 10%, driver frequency = 60
Hz, driver cycles = 3, motion-encoding-gradient (MEG) frequency = 75 Hz, MEG
amplitude = 3 G/cm, 4 MRE time points, and MEG direction = through plane (Z).
Region of Interest (ROI): Use the MRE MR
magnitude image to define a circular ROI in the middle of the phantom with a
diameter half that of the phantom’s diameter. The edges of the phantom should be avoided due to edge artifacts in the MRE image
processing (Figure 2). The mean and standard deviation of the pixel values in the ROI on
elastograms are reported as the phantom stiffness (typically in units of Pa or
kPa).
Longitudinal Schedule and Pass Criteria:
The MRE QA phantom exams should be scheduled every six
months for each site. The longitudinal pass criterion uses the difference in
the measured stiffness at two consecutive time points (Figure 3). The pass criterion for
the current exam is a stiffness measurement difference ≤ 10%. The threshold of
10% is used to account for the possibility of the PVC phantom stiffening over
time (up to 10% per year).
Example of Failed MRE QA Exam
During hepatic MRE
scans, the driver motion is required to be continuous because discontinuous
motion can cause errors in the elastograms. Motion discontinuities normally
happen when the duration of the active driver is shorter or longer than the TR
of the imaging sequence. Usually the motion duration and TR are precalculated and
synchronized so the motion is always continuous. However, if the TR and the driver duration are
inadvertently set to incorrect values, then the synchronization between the MRE
hardware and the imaging sequence can be disrupted, causing the vibrational
motion to be discontinuous.
We demonstrate a failure situation in which we intentionally set the
motion duration 0.1 ms longer than the TR (50 ms) so the synchronizing triggers
from the imaging sequence are missed every other TR causing motion
discontinuity (Figure 4). In a 5.9
kPa PVC phantom, this discontinuity caused a stiffness measurement difference
of -14.7%, which did not meet the pass criterion (Figure 5). Conclusion
A MRE system longitudinal QA protocol is key to verifying
the integrity of the whole hepatic MRE system over time and insuring that the
MRE system is working properly. This is
critical when using MRE to longitudinally monitor liver disease progression or
treatment efficacy. Acknowledgements
NIH EB 001981References
No reference found.