Lorenzo Di Sopra1, Davide Piccini1,2, Simone Coppo3, Jessica A.M. Bastiaansen1, Matthias Stuber1,4, and Jérôme Yerly1,4
1Department of Radiology, University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland, 2Advanced Clinical Imaging Technology, Siemens Healthcare AG, Lausanne, Switzerland, 3Department of Radiology, Case Western Reserve University, Cleveland, OH, 4Center for Biomedical Imaging (CIBM), Lausanne, Switzerland
Synopsis
The performance of
motion-resolved whole-heart MR imaging strongly depends on the quality of
cardiac- and respiratory-gating signals. While navigators or self-navigation
can be used to account for respiratory motion, ECG is a mainstay for
synchronizing data acquisition with the cardiac cycle. We tested whether physiological
motion information, directly extracted from k-space-center, can replace
respiratory navigators and ECG signals. The proposed solution was applied in 9 healthy
volunteers and results were compared to those obtained with the ECG-signal. Correlation
between R-wave time-stamps from the ECG and the cardiac self-gating signal was
excellent, while image quality and coronary artery conspicuity remained
unchanged.
INTRODUCTION
Cardiac and
respiratory motion-resolved whole-heart MR imaging techniques have recently
been introduced and enable simultaneous anatomical and functional assessment of
the heart from one single free-breathing scan1,2. However, these
techniques still require the use of external ECG devices to synchronize data
acquisition1,3 with the cardiac cycle. Unfortunately, this adds to
the patient setup time on the one hand and sometimes leads to artifactual
trigger points because of the magneto hydrodynamic effect on the other. In this
study, we aim to address these limitations by investigating self-gating strategies
that directly extract information about physiological motion from the center of
k-space. More specifically, we compare the timing of the ECG and the cardiac self-gating
signal and the quality of 5D whole-heart images independently reconstructed using
these two synchronization signals.METHODS
Data were acquired in
N=9 healthy volunteers on a 1.5T clinical MRI scanner (MAGNETOM Aera, Siemens
Healthcare) using a prototype non-ECG-triggered 3D golden angle radial bSSFP
sequence4. The segmented acquisition used a Phyllotaxis radial
sampling pattern5, where each segment was preceded by a chemically
selective radiofrequency pulse for fat suppression, 10 linearly increasing RF
preparation pulses, and one readout always oriented along the superior-inferior
(SI) direction (Fig.1a,b). Two automated algorithms were compared for the
extraction of cardiac and respiratory self-gating (SG) signals. The first
method consists of Fourier transforming (FT) all SI readouts and applying
principal component analysis (PCA) to the corresponding image space projections2 (Fig.1c).
The second method directly extracts the SG signals from the modulation of the central
coefficient of the k-space SI readouts6 (K0-modulation)(Fig.1b). The
principal components and the modulated K0-signals were bandpass filtered
(Fig.1d) to isolate the cardiac and respiratory motion signals by automatically
detecting the subject’s specific frequency components (Fig.2). The principal component
(for the PCA) or the coil element (for the K0-modulation) yielding the highest energy
within the subject’s specific frequency bands were automatically selected. The
filtered cardiac signal was then interpolated for finer detection of the signal
peaks (Fig.3a). These peaks were considered as SG triggers points and compared with
the R-wave time stamps from the ECG, which was recorded for reference during data acquisition. The number
of skipped ECG triggering points was also recorded and reported for each
dataset. The time intervals between two consecutive SG cardiac triggers were
compared to their ECG counterparts using linear regression analysis and
Bland-Altman plots. Of the two SG algorithms, only the most precise (i.e.,
smallest deviation from the ECG triggering points intervals) was considered for
the final SG reconstruction. Cardiac SG and R-wave time stamps derived from the
ECG were separately used to sort the acquired data into cardiac phases of 50 ms
duration, while the SG respiratory signal, extracted from the K0-modulation,
was used to resolve the data into 4 respiratory motion states in both cases. A
k-t sparse SENSE algorithm allowed the reconstruction of undersampled 5D
datasets7 (x-y-z-cardiac-respiratory dimensions). The reconstructed
images were first compared by measuring vessel length and sharpness8
of the right coronary artery (RCA) and left anterior descending artery (LAD),
then they were visually evaluated.RESULTS
SG cardiac and
respiratory motion signals were successfully extracted in all 9 volunteers with
both algorithms (Fig.4a). Some ECG triggering points were missed in 5/9
volunteers (volunteers 1-5, Fig.4a). The SG cardiac intervals deviated from the
reference ECG RR-intervals by an average of 28.8±12.5 ms and 31.3±15.0 ms for the
K0-modulation and PCA algorithms, respectively (p=0.49). However, if only
datasets with a maximum 2% of missing points are considered (volunteers 5-9),
these values decrease to 21.0±4.6 ms and 23.4±4.7 ms (p=0.53), respectively (Fig.3b-4a).
Because of the slightly better precision, the K0-modulation algorithm was
chosen for the final SG reconstruction. The measurements of vessel length and
sharpness (Fig.4b) and visual comparison of the overall image quality (Fig.5) showed good agreement between SG and the ECG-gated approach.DISCUSSION AND CONCLUSION
In the proposed
framework for automated signal extraction, the SG cardiac triggers proved to be highly correlated with the reference ECG signal, especially with the
K0-modulation approach. Unlike the ECG signal that is susceptible to skipped
triggering points, the SG strategy is immune to magnetic perturbations. These
preliminary results suggest that external ECG devices may no longer be
necessary for gating of simultaneous functional and anatomical 3D scans of the
heart. This would have positive implications on the ease-of-use and patient
setup times, while the frequent need for repositioning of the electrodes is
avoided altogether. However, further quantitative validations are now mandatory
and the performance of this approach in patients with arrhythmias or frequent
ectopic beats remains to be determined.Acknowledgements
The authors would like
to acknowledge Dr. Florian Knoll from NYU School of Medicine for support with
the GPU implementation of the 3D NUFFT. This work was partly supported by the
Swiss National Science Foundation grants 320030_143923 and 326030_150828.References
1. Feng, L., Coppo, S., Piccini, D., Lim,
R. P., Stuber, M., Sodickson, D. K. and Otazo, R. (2015), Five-Dimensional
Cardiac and Respiratory Motion-Resolved Whole-Heart MRI. Presented to ISMRM
23rd Annual Meeting & Exhibition, 01 June 2015, Toronto, Ontario, Canada
2. Pang, J., Sharif, B., Fan, Z., Bi, X.,
Arsanjani, R., Berman, D. S. and Li, D. (2014), ECG and navigator-free
four-dimensional whole-heart coronary MRA for simultaneous visualization of
cardiac anatomy and function. Magn. Reson. Med., 72: 1208–1217.
doi:10.1002/mrm.25450
3. Coppo, S., Feng, L.,
Piccini, D., Chaptinel, J., Bonanno, G., Vincenti, G., Schwitter, J., Otazo,
R., Sodickson, D. K. and Stuber, M. (2015), Improved free-running
self-navigated 4D whole-heart MRI through combination of compressed sensing and
parallel imaging. Presented to ISMRM 23rd Annual Meeting & Exhibition, 01
June 2015, Toronto, Ontario, Canada
4. Coppo, S., Piccini, D., Bonanno, G.,
Chaptinel, J., Vincenti, G., Feliciano, H., van Heeswijk, R. B., Schwitter, J.
and Stuber, M. (2015), Free-running 4D whole-heart self-navigated golden angle
MRI: Initial results. Magn. Reson. Med., 74: 1306–1316. doi:10.1002/mrm.25523
5. Piccini, D., Littmann, A.,
Nielles-Vallespin, S. and Zenge, M. O. (2011), Spiral phyllotaxis: The natural
way to construct a 3D radial trajectory in MRI. Magn. Reson. Med., 66:
1049–1056. doi:10.1002/mrm.22898
6. Larson, A. C., White, R. D., Laub, G.,
McVeigh, E. R., Li, D. and Simonetti, O. P. (2004), Self-gated cardiac cine
MRI. Magn. Reson. Med., 51: 93–102. doi:10.1002/mrm.10664
7. Feng, L., Axel, L., Chandarana, H.,
Block, K. T., Sodickson, D. K. and Otazo, R. (2016), XD-GRASP: Golden-angle
radial MRI with reconstruction of extra motion-state dimensions using
compressed sensing. Magn. Reson. Med., 75: 775–788. doi:10.1002/mrm.25665
8. Etienne, A., Botnar, R. M., van
Muiswinkel, A. M.C., Boesiger, P., Manning, W. J. and Stuber, M. (2002),
“Soap-Bubble” visualization and quantitative analysis of 3D coronary magnetic
resonance angiograms. Magn. Reson. Med., 48: 658–666. doi:10.1002/mrm.10253