Francesco Santini1,2, Grzegorz Bauman1,2, and Oliver Bieri1,2
1Division of Radiological Physics, University of Basel Hospital, Basel, Switzerland, 2Department of Biomedical Engineering, University of Basel, Basel, Switzerland
Synopsis
CABIRIA
is a method for the simultaneous quantification of T1 and T2 values
of the myocardium. As other T1-quantification sequences, it relies on
a bSSFP readout, and it is thus susceptible to off-resonance effects.
In this work, we show an improvement over the CABIRIA method that
reduces the TR and therefore increases the robustness towards field
inhomogeneities. This method can potentially be extended to other
quantification techniques.
Introduction
Cardiac
relaxometry has attracted considerable interest in the last years for
diagnosis of both diffuse and focal pathologies of the myocardium.
Recently, a method for simultaneous quantification of T1 and T2,
named Cardiac Balanced Inversion Recovery with Interleaved
Acquisition (CABIRIA) was proposed1. CABIRIA is based on an
inversion pulse followed by a balanced steady state free precession
(bSSFP) readout, also common to a number of other cardiac T1
quantification methods2-4. Generally, all
bSSFP-based methods are off-resonance sensitive to some extent5,6,
leading to quantification inaccuracies. In this work, the repetition
time (TR) of the CABIRIA sequence is shortened using a recently
proposed SSFP optimization strategy for lung imaging7 to
reduce its sensitivity to field inhomogeneities.Methods
The
CABIRIA method is based on an adiabatic inversion pulse followed by a
continuous train of single-shot Cartesian bSSFP kernels encoding the
recovery curve. After a free recovery period, the acquisition is
repeated with a delay between the inversion and acquisition realized
with dummy TRs1. In this work, the bSSFP kernels of CABIRIA were
optimized by reducing the excitation duration and improving the
gradient switching patterns by using overlapping ramps and asymmetric
readout, in a similar way as described in7,8. This
optimized version, termed ultrafast-CABIRIA (uf-CABIRIA) was compared
to a conventional CABIRIA implementation in a phantom and in vivo.
The
phantom experiment was performed with a spherical homogeneous
phantom, and the shim was tweaked in order to obtain a linearly
increasing frequency shift across the phantom. The acquisition
parameters for uf-CABIRIA were: resolution 1.33x1.33x8mm3, matrix
size 256x208, bandwidth 1775Hz/px, flip angle 35°, total acquired
images 50. The achieved TR/TE for the uf-CABIRIA sequence was
2.1/0.7ms. The CABIRIA sequence was resolution-matched and had a
bandwidth if 1550Hz/px and a TR/TE of 2.9/1.3ms.
For
the in vivo experiment, the same acquisition parameters were used and
an additional CABIRIA acquisition with a lower resolution, and
therefore lower TR, was also acquired for comparison. The parameters
of the low-resolution CABIRIA were resolution 1.7x1.7x8mm3, matrix
size 192x162, bandwidth 1530Hz/px, TR/TE 2.7/1.19ms. A single slice
of a mid-ventricular short axis view of the heart was acquired in a
healthy volunteer.
T1
and T2 maps were calculated using the procedure described in1.
In
the phantom, a longitudinal profile of the acquired maps was
extracted and compared between the two different methods. The in vivo
images were visually compared and manual segmentation was performed
according to the American Heart Association 17-segment model. The
variation of the quantitative values across the segments was
analyzed.Results
The
uf-CABIRIA sequence allowed a TR reduction of 28%, consequentially
leading to an increase of the bSSFP passband by 40%.
The
phantom experiments showed a clear difference in sensitivity between
the two sequences, the uf-CABIRIA sequence having a larger area of
homogeneity and especially a more consistent behavior in T2
quantification (fig. 1).
In
the in vivo acquisition, an off-resonance-related bias is apparent
for the CABIRIA T1 and T2 maps in the anterior and anterolateral
regions, being more pronounced in T2. The uf-CABIRIA shows the same
but damped inhomogeneity pattern. Similar, from the shortened TR, the
low resolution CABIRIA scan is less affected by inhomogeneities but
suffers from significant partial volume effects (fig. 2).
Segmentation
confirms the visual inspection of decreased values in the anterior
and anterolateral segments for CABIRIA that become less pronounced
for uf-CABIRIA (fig. 3).Discussion
The
optimizations introduced to the uf-CABIRIA acquisition markedly
enhanced T1 and T2 mapping in cardiac regions affected by B0
inhomogeneity, namely for the anterior and anterolateral segments
that are close to the airspace of the lungs. Decreasing the
resolution in order to reduce the TR of a conventional CABIRIA
acquisition similarly redices the off-resonance sensitivity, but at
the cost of increased partial volume effects affecting precision.
Overall, no artifacts from the TR acceleration were visible in the
images, thus recommending uf-CABIRIA for cardiac imaging, especially
at higher field strengths.Conclusion
CABIRIA,
and possibly other bSSFP-based cardiac relaxometry sequences, such as
MOLLI, can benefit from a shortening of the TR using a recently
proposed optimization strategy for ufSSFP imaging to improve
quantification robustness in the presence of B0 inhomogeneities.Acknowledgements
This
work was partly supported by Siemens Healthineers.References
1. Santini F,
Kawel-Boehm N, Greiser A, Bremerich J, Bieri O. Simultaneous T1 and
T2 quantification of the myocardium using cardiac balanced-SSFP
inversion recovery with interleaved sampling acquisition (CABIRIA).
Magn. Reson. Med. 2015;74:365–371. doi: 10.1002/mrm.25402.
2. Messroghli
DR, Radjenovic A, Kozerke S, Higgins DM, Sivananthan MU, Ridgway JP.
Modified Look-Locker inversion recovery (MOLLI) for high-resolution
T1 mapping of the heart. Magn. Reson. Med. Off. J. Soc. Magn. Reson.
Med. Soc. Magn. Reson. Med. 2004;52:141–146. doi:
10.1002/mrm.20110.
3. Chow K,
Flewitt JA, Green JD, Pagano JJ, Friedrich MG, Thompson RB.
Saturation recovery single-shot acquisition (SASHA) for myocardial
T(1) mapping. Magn. Reson. Med. 2014;71:2082–2095. doi:
10.1002/mrm.24878.
4. Piechnik SK,
Ferreira VM, Dall’Armellina E, Cochlin LE, Greiser A, Neubauer S,
Robson MD. Shortened Modified Look-Locker Inversion recovery
(ShMOLLI) for clinical myocardial T1-mapping at 1.5 and 3 T within a
9 heartbeat breathhold. J. Cardiovasc. Magn. Reson. Off. J. Soc.
Cardiovasc. Magn. Reson. 2010;12:69. doi: 10.1186/1532-429X-12-69.
5. Kellman P,
Hansen MS. T1-mapping in the heart: accuracy and precision. J.
Cardiovasc. Magn. Reson. 2014;16:2. doi: 10.1186/1532-429X-16-2.
6. Kellman P,
Herzka DA, Arai AE, Hansen MS. Influence of Off-resonance in
myocardial T1-mapping using SSFP based MOLLI method. J. Cardiovasc.
Magn. Reson. 2013;15:63. doi: 10.1186/1532-429X-15-63.
7. Bieri O.
Ultra-fast steady state free precession and its application to in
vivo (1) H morphological and functional lung imaging at 1.5 tesla.
Magn. Reson. Med. Off. J. Soc. Magn. Reson. Med. Soc. Magn. Reson.
Med. 2013. doi: 10.1002/mrm.24858.
8. Bauman G,
Santini F, Pusterla O, Bieri O. Pulmonary relaxometry with inversion
recovery ultra-fast steady-state free precession at 1.5T. Magn.
Reson. Med. 2016. doi: 10.1002/mrm.26490.