Saipavitra Venkateshwaran Murali Manohar1, Ioannis Angelos Giapitzakis1, Tamas Borbath1, Matti Gaertner2, and Anke Henning1,3
1High Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Tuebingen, Germany, 2Department of Psychiatry, Charité - Universitätsmedizin Berlin, 3Institute of Physics, Ernst-Moritz-Arndt University Greifswald, Greifswald, Germany
Synopsis
J-resolved semi-LASER with maximum-echo sampling is
optimized at 9.4T and compared with the same implementation at 3T in terms of
SNR and spectral resolution. SODA scheme is appreciated for the sequence rather
than the MC scheme. SNR at 9.4T (t1 steps: 85) was approximately 5.8
times greater than at 3T (t1 steps: 100) and strongly coupled peaks
are well-resolved. However, the trade-off between SNR and spectral resolution
is explained as lactate (1.32 ppm), a weakly-coupled metabolite, is better
resolved at 3T. Higher band-width AFP pulses helped in almost vanishing
the J-refocused peaks which made the J-resolved peaks clearly distinguishable.
A few interesting downfield peaks and the doublet of NAA (7.82ppm) are
observed.
Purpose
Enhancing the signal-to-noise ratio(SNR) and
the spectral resolution have always been the ambition in the field of in vivo 1H
magnetic resonance spectroscopy. Spectroscopic studies at ultra-high field
benefit from both increased spectral resolution and improved SNR. A distinct
method to reduce spectral signal overlap is 2D spectroscopy, which spreads the
spectral information along two frequency axes1-5. Exploiting these two
complimentary approaches, a J-resolved semi-LASER sequence with maximum-echo
sampling(MES) scheme was optimized for the application in human brain at 9.4T.
Also a qualitative comparison between J-resolved semi-LASER at 3T1 and 9.4T in
terms of SNR and spectral resolution was done.Methods
J-resolved semi-LASER experiments
were performed on a Siemens Magnetom 9.4T and a Siemens Trio 3T whole-body MRI
scanner(Erlangen, Germany). A voxel of 2x2x2 cm3 was chosen
in the occipital lobe with a mixed gray and white matter content at 9.4T and in
the pre-frontal cortex at 3T. Four-channel transceiver coil6 was
used at 9.4T and was shimmed using the FASTEST MAP7. Power
calibration was performed for the selected voxel. At 3T and 9.4T, the shortest TE
was 24ms, the step size was ∆TE=2ms. 50 or 85 echo time increments were
recorded at 9.4T whereas at 3T 100 steps5,8 were recorded. Bandwidth
in the direct dimension was 4000Hz and 2000Hz at 9.4T and 3T, respectively.
The sampling points were 4096 in the direct dimension. MES scheme was
implemented(Figure 1). Metabolite cycling(MC)9 scheme was tested
for the sequence with 8 and 16 phase-cycling steps and the resulting spectral
quality was compared to experiments with SODA10 water suppression(Short
duration water suppression using optimized flip angles). Also a non-water
suppressed experiment was performed with otherwise identical scan parameters in
order to allow for Eddy current correction11.The sequence was
optimized by performing scans on phantom and 8 healthy volunteers and later 7 healthy
volunteers participated in the study at 3T(n=1), 9.4T with 50(n=5) and 85
steps(n=1), respectively, all of them gave informed signed consent as approved
by the local ethics board.Results
The MES scheme gave the peak
tails a tilt in the 2D spectra and therefore avoided overlap of spectral peaks
and peak tails from high intensity metabolites and water. SODA water suppression,
yielded a suppression factor of 99.47%
at 9.4T and 99.23% at 3T, gave minimal ghosting artifacts whereas MC
gave stronger ghosting artifacts in the 2D spectra for 8 phase-cycling steps,
which reduced with 16-step phase cycle. Due to the high pulse bandwidth and
minimal chemical shift displacement artifact, J-refocused peaks were barely
visible in either the 2D spectra at 3T or at 9.4T(Figure 3,4). J-resolved
peaks could be well-distinguished at 9.4T(Figure 2,3) while 3T J-resolved semi-LASER data exhibited a stronger spectral overlap. However, the
lactate peak at 1.32 ppm could clearly be resolved at 3T but not at 9.4T due to
a lower number of sampling points in the indirect dimension(Figure 3,4) which
represented a trade-off between SNR and spectral resolution at 9.4T. At only 9.4T it was possible to observe some of the down-field peaks12 like
NAA(7.82ppm) and hCs(8.08ppm). The doublet of NAA at 7.82 ppm was
prominently noticeable at 9.4T(Figure 2).Discussion
J-resolved semi-LASER with MES scheme was investigated
for the first time at 9.4T and compared with 3T results. Compared to previous
1D MRS studies at 9.4T13 low number of averages per TE in J-resolved
MRS restricts the number of possible phase cycling steps(herein 8), which
appeared to be insufficient to suppress ghosting artifacts introduced by
metabolite cycling. Nevertheless, SODA10 provided an excellent water
suppression despite B1+ inhomogeneity at high field
strengths and at 3T. Unwanted coherence pathways were suppressed during
the SODA water suppression scheme using the most suitable combination of
gradients. The compromise between SNR and spectral resolution of the J-resolved
spectra was investigated. The weakly coupled metabolites like lactate could be
better resolved at 3T rather than at 9.4T because of a larger TE range due to
the Carr-Purcell behavior(CP)14,15 of the semi-LASER sequence(Figure
4). Though 9.4T showed a higher SNR(Table 1) and strongly-coupled peaks such as
NAAasp and mI were well-resolved, T2 relaxation times are
shorter at 9.4T16 than 3T. So, optimizing the TE range to avoid loss
of signal while providing sufficient time for J-evolution is required in order
to achieve both higher SNR and well-resolved weakly-coupled peaks. The
appearance of the downfield peaks showed that J-resolved semiLASER at 9.4T
could be a potential method to help assigning unknown peaks by understanding
their J-coupling properties, although the optimal TE range for downfield MRS
should be smaller as their T2 values are shorter17.
Acknowledgements
This work was supported by the Horizon 2020 CDS-QUAMRI grant. The authors thank Simone Grimm, Evgeniya Kirilina, Yan Fan, Anna Stippl,
Ana Herrera from Charité - Universitätsmedizin Berlin for help in this work.
References
1. Lin
M, Kumar A, Yang S. Two-Dimensional J-Resolved LASER and Semi-LASER
Spectroscopy of Human Brain. Magn Reson Med 2014; 71(3):911-920.
2. Thomas M, Yue K, Binesh N, et al. Localized
two-dimensional shift correlated MR spectroscopy of human brain. Magn Reson Med
2001; 46:58–67.
3. Andronesi O, Ramadan S, Mountford C, Sorensen A. Low-power
adiabatic sequences for in vivo localized two-dimensional chemical shift
correlated MR spectroscopy. Magn Reson Med 2010; 64:1542–1556.
4. Lange T, Trabesinger A, Schulte R, Dydak U,
Boesiger P. Prostate spectroscopy at 3 Tesla using two-dimensional S-PRESS. Magn
Reson Med 2006; 56:1220–1228.
5. Schulte
R F, Lange T, Beck J, Meier D, Boesiger. Improved two-dimensional J-resolved
spectroscopy. NMR Biomed 2006; 19(2): 264-270.
6. Pfrommer A, Avdievich N and Henning A (May-12-2014): Four
Channel Transceiver Array for Functional Magnetic Resonance Spectroscopy in the
Human Visual Cortex at 9.4 T, Joint Annual Meeting ISMRM-ESMRMB 2014, Milano,
Italy.
7. Gruetter
R, Tkac I. Field mapping without reference scan using asymmetric echo-planar
techniques. Magn Reson Med 2000; 43(2):319-23.
8. Fuchs
A, Boesiger P, Schulte RF, Henning A. ProFit revisited. Magn Reson Med. 2014; 71(2):458-68.
9. Giapitzakis
IA, Nassirpour S, Avdievich N, Kreis R, Henning A. Metabolite cycled single
voxel 1H spectroscopy at 9.4 T. 23rd Annual Meeting and Exhibition of ISMRM 2015.
10. Giapitzakis
IA, Nassirpour S, Henning A. Short duration water suppresion using optimised
flip angles (SODA) at ultra high fields.32nd Annual Scientific Meeting ESMRMB
2015.
11. Klose U. In vivo proton spectroscopy in presence
of eddy currents. Magn Reson Med. 1990; 14(1):26-30.
12. Govindaraju V, Young K, Maudsley AA. Proton
NMR chemical shifts and coupling constants for brain metabolites. NMR Biomed.
2000; 13(3):129-53.
13. Giapitzakis IA, Shao T, Avdievich
N, Kreis R, Henning A. Metabolite cycled semi-LASER in human brain at 9.4T: In-vivo
results. ISMRM Workshop on MR
Spectroscopy, Lake Constance 2016.
14. Allerhand A, Thiele E. Analysis of
Carr—Purcell Spin-Echo NMR Experiments on Multiple-Spin Systems. II. The Effect
of Chemical Exchange. J Chem Phys. 1966; 45(3):902-16.
15. Deelchand DK, Henry PG, Marjanska M. Effect of
carr-purcell refocusing pulse trains on transverse relaxation times of
metabolites in rat brain at 9.4 Tesla. Magn Reson Med. 2015; 73(1):13-20.
16. Deelchand DK, Van de Moortele PF, Adriany G,
Iltis I, Andersen P, Strupp JP, Vaughan JT, Ugurbil K, Henry PG. In vivo 1 H
NMR spectroscopy of the human brain at 9.4 T: initial results. J Magn Reson.
2010; 206(1):74-80.
17. Fichtner ND, Henning A, Zoelch N, Boesch C,
Kreis R. Eludication of the Downfield Spectrum of Human Brain at 7 T Using
Multiple Inversion Recovery Delays and Echo Times. Magn Reson Med 2016.
18. Scheenen TW, Klomp DW, Wijnen JP, Heerschap A.
Short echo time 1H-MRSI of the human brain at 3T with minimal chemical shift
displacement errors using adiabatic refocusing pulses. Magn Reson Med. 2008; 59(1):1-6.
19. Edden RAE, Barker P B. If J doesn’t evolve, it
won’t J-resolve: J-PRESS with bandwidth-limited refocusing pulses. Magn Reson
Med 2011; 65(6): 1509-14.