Seyedmorteza Rohani Rankouhi1 and David Norris2
1Erwin L. Hahn Institute for Magnetic Resonance Imaging, Essen, Germany, 2Donders Institute for Brain, Cognition and Behaviour, Nijmegen, Netherlands
Synopsis
The most
commonly used editing MRS method for measuring GABA is MEGA which works at TE=68 ms and any odd multiples thereof. Here we present
a J-difference editing method which makes it possible to measure GABA signal at
variable TEs independent of the effect of J-evolution by preserving the two side
peaks of GABA in antiphase states at any echo time above 34ms. We also show an application of this editing technique to be measurement of T2 relaxation time of GABA in vivo.
Purpose
The most
commonly used editing MRS method for measuring GABA is MEGA1. This
editing method works at TE=68 ms and any odd multiples thereof. Here we present
a J-difference editing method which makes it possible to measure GABA signal at
variable TEs independent of the effect of J-evolution by preserving the two side
peaks of GABA in antiphase states at any echo time above 34ms.Methods
The antiphase editing sequence is shown in
Figure1 which involves two interleaved modes, Mode1 and Mode2. A single narrow
bandwidth Gaussian pulse is used for editing. As shown in Figure1, the timing conditions
for Mode1 is T1-T2 = 34ms=1/4J and the condition for Mode2 is T2–T1=34ms=1/4J
ms. By having these two timing conditions at any TE, the two side peaks of GABA
will be in antiphase state relative to the central peak (90 and -90 degrees out
of phase relative to the central peak). The difference between the two modes
would be that the phase of the two side peaks will be reversed relative to each
other. Therefore, taking the difference of the two modes will remove central
peak and the two side peaks will be doubled. Compared to the MEGA method where
the two side peaks are in phase in both ON and OFF modes, in antiphase editing
method, the two side peaks are 180 degrees out of phase relative to each other.
The antiphase technique is only viable because the separation of the two
sidepeaks is 15Hz, so for commonly encountered shim conditions where the
linewidth is comparable to or narrower than the separation of the sidepeaks, the
signal cancellation can be acceptable. The
antiphase editing sequence was implemented in a sLASER sequence using a single
Gaussian pulse with duration of 15ms and bandwidth of 120Hz. Phantom and in
vivo scans were performed on a 7T system (Magnetom 7T, SIEMENS Healthcare GmbH,
Germany) with 32 channel Rx and 1 channel Tx head coil (Nova Medical, NY). The
editing performance of the sequence was first tested on a 30 mMol GABA phantom
at six different TEs of 77,87,97,107,117 and 127ms. In vivo data were then
acquired from one healthy subject who participated with ethical approval from institutional
ethics committee. Anatomical reference was acquired using 3D MPRAGE. B0 shimming
was performed using FASTESTMAP2. Single voxel MRS data were
collected at TEs = 77,87,97,107,117and 127ms from 20x20x20 mm3 voxel
positioned at occipital region using antiphase editing sequence. Data were
analyzed using JMRUI software3 and MATLAB4.Results
Figure2 shows acquired difference spectra from
phantom at 6 different TEs. It shows the two side peaks of GABA being in
antiphase state independent of TE demonstrated in the real spectra and decay of
GABA signal and elimination of the central peak demonstrated in the magnitude
spectra. For in vivo scan, while no GABA signal was observed in the difference
spectrum at TE=77 ms, results at other five TEs are shown in Figure3. GABA at 3
ppm is indicated in the spectra in this figure. In addition to GABA, there are
some other coedited peaks observable in the difference spectra indicated with
letters A-D in each resultant spectrum. Figure4 shows T2 relaxation time of
GABA in phantom and in vivo calculated by mono-exponential fitting to be 62 ms
and 49 ms respectively.Discussion
The
editing method we present here measures GABA signal independent of the effect
of J evolution. The antiphase editing method brings the danger of the two
signals canceling each other if they are closer than the linewidth. For GABA,
the two side peaks are around 15 Hz away from each other, but the method is not
useful for editing metabolites like lactate where the side peaks are just 7 Hz
away or for J coupled metabolites with doublet of doublet pattern. For
instance, a characteristic of
antiphase difference spectrum compared to standard MEGA spectrum is the absence
of Glx at 3.75 ppm indicated in Figure3 as D which is because Glutamate being
doublet of doublet. In addition to GABA there are some coedited signals
observable in the antiphase editing difference spectrum indicated as A,B, and C
in Figure3 which do not show up in the MEGA technique.Conclusion
The antiphase editing method provides the possibility
of measuring GABA at variable TEs independent of the effect of J evolution on
signal intensity and hence provides the possibility of measuring T2 relaxation
time of GABA as an application.Acknowledgements
This work was funded by the Helmholtz Alliance ICEMED – Imaging and Curing Environmental Metabolic Diseases, through the Initiative and Networking Fund of the Helmholtz Association.References
1. Mescher
M, et al. Simultaneous in
vivo spectral editing and water suppression. NMR Biomed. 1998;11(6):266-72.
2. Gruetter
R, Tkác I. Field mapping without reference scan using asymmetric echo-planar techniques.
Magn Reson Med. 2000;43(2):319-23
3. Naressi
A, et al. Java-based graphical user interface for
the MRUI quantitation package. MAGMA. 2001;12(2-3):141-52.
4. MATLAB
Release 2016b, The MathWorks, Inc., Natick, Massachusetts, United States.