Hui Chen1, Xiaohai Ma1, Lei Zhao1, Xiaoyong Zhang2, Guoxi Xie2, Tianjing Zhang3, and Zhanming Fan1
1Beijing Anzhen Hospital, Beijing, People's Republic of China, 2Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, People's Republic of China, 3MR Collaborations NE Asia, Siemens Healthcare, Beijing, China
Synopsis
Traditional
CMR sequences are useful tools for assessing cardiac structure and function, However,
the poor image quality and motion artifacts caused by arrhythmia may hamper the
diagnostic quality of CMR images. We hypothesized that the optimized temporal parallel
acquisition technique (TPAT) may
improve this situation by allowing free-breathing of subjects and meanwhile
provides a similar diagnostic values for radiologists
.
Purpose
Traditional
CMR sequences are useful tools for assessing cardiac structure and function, However,
the poor image quality and motion artifacts caused by arrhythmia may hamper the
diagnostic quality of CMR images. We hypothesized that the optimized temporal parallel
acquisition technique (TPAT) may improve this
situation by allowing free-breathing of subjects and meanwhile provides a
similar diagnostic values for radiologists.Methods
Experiment: Thirteen patients
with arrhythmia and 13 patients with normal heart rhythm were were included in
this study. CMR scan was performed using a 3.0T system (MAGNETON Verio, Siemens
Healthcare, Erlangen, Germany). Cine images were acquired by bSSFP sequence
with conventional retrospective ECG-triggering method and the optimized real-time
TPAT technique
in the same planes. The optimized TPAT sequence allows for
free-breathing scans, which is beneficial for those with a poor breath-hold. Parameters of the
conventional scan including:
voxel size 1.3×1.3×5.0mm3, TR/TE 41.1/1.51ms, FOV 340×289 mm2, scan time per slice 12 s, temporal resolution
41.1ms; The
parameters of TPAT scan were:
voxel size 2.9×2.1×8.0mm3, TR/TE 60.48/1.12ms FOV 340× 289 mm2, scan
time per slice 2.6 s, temporal resolution 60.48ms, TPAT factor 4.
Image analysis: For quantitative analysis, LV global
myocardial longitudinal strain (GLS), LV circumferential strain and radial
strain (GCS and GRS) were measured by dedicated software (QStrain version 2.1, Medis, Leiden, the
Netherlands), LV volumes and function were measured using Argus software
(Siemens Healthcare, Erlangen, Germany). Specifically, GLS
was calculated from 4-chamber, 3-chamber and 2-chamber orientations. GCS and GRS
were calculated in the short axis orientation. LV volumes and
function were measured in the short axis orientation.
Besides, apparent contrast-to-noise ratio (CNR) between LV myocardium and blood
pool was also calculated. For
qualitative analysis, two radiologists subjectively scored the image quality
(IQ) using a four-point grading scale (1=very good IQ, no artifacts affecting
cardiac anatomy; 2=good/average IQ, artifacts slightly interfering with cardiac
anatomy; 3=below-average IQ, artifacts moderately affecting cardiac anatomy;
4=poor IQ with artifacts severely affecting cardiac anatomy) (1). Wilcoxon signed-rank
test was used to compare the measured samples between the conventional method and
optimized TPAT technique in both
patient groups with/without arrhythmia.
Results
There
was no
significant difference between the conventional method and TPAT technique in
evaluation of LV volumes and function (left ventricular end diastolic volume, EDV; left ventricular end systolic volume, ESV; left ventricular ejection
fraction, LVEF; left ventricular stroke volume, LVSV; left ventricular cardiac
output, LVCO and Mass), it’s also the same for longitudinal, circumferential
and radial strains in both patient groups (Table 1). For the images of patients
with arrhythmia obtained by TPAT technique, there
were 6 cases of grade 1 (46%), 6 cases of grade 2 (46%) and 1 case of grade 3 (8%), while
for those obtained by conventional method, there were 4 cases of grade 2 (31%),
8 cases of grade 3 (62%), and 1 case of grade 4 (7%). The IQ of TPAT
technique was higher than that of conventional method in patients with arrhythmia
on the whole. (Figure 1). The CNR values in patients with arrhythmia obtained
by TPAT were significantly higher than those by
conventional method. (10.0±3.4 vs 8.2±2.4, p = 0.009).Discussion
For
subjects with normal heart rhythm, there were no obvious difference in CNR of
images in TPAT and conventional methods. However, the IQ of conventional method
was higher than that of TPAT. For patients with arrhythmia, CNR of images in
TPAT sequence was higher than conventional cine sequence. No matter in subjects
with normal heart rhythm or patients with arrhythmia, there were no significant
differences between conventional and optimized TPAT methods in LV
volumes, function as well as
longitudinal, circumferential, radial strains. Nevertheless, TPAT can be performed
on patients without need of breath holding and use of ECG triggering, which has
greater clinical significance than conventional method. Conclusion
With the optimized TPAT technique, the cardiac cine can provide better image quality and shorter scan time in arrhythmia in this preliminary study. Nevertheless, it
has the potential advantages of allowing free-breathing scans.Acknowledgements
This work was supported in part by Capital Health Research and Development of Special (2016-4-2063) and Natural Science Foundation
of China (81671647).
References
1.Kaasalainen T. Cardiac MRI in patients with cardiac pacemakers: practical
methods for reducing susceptibility artifacts and optimizing image quality. Acta Radiol. 2016 Feb;57(2):178-187.