4D Flow-MRI offers unique possibilities for the diagnosis of pulmonary hypertension. A validation of the sequence on a digital broadband 3T MR system is essential before introduction into clinical routine. This study compared 4D Flow–MRI derived stroke volume (SV) and maximal flow velocity (Vmax) in the pulmonary arteries of 23 healthy volunteers to 2D phase contrast MRI, right ventricular stroke volume (RVSV), conservation of mass-analysis (COM), and to a static phantom for additional error analysis. Results revealed clinically acceptable differences with a trend of phantom correction to improve results except for the COM-analysis.
MRI scans: The pulmonary arteries of 23 healthy volunteers (14f, 38±18y, BMI 21.5±3.0 kg/m², HR 65±9/min) were examined after IRB approval and written informed consent on a digital broadband 3T system (Philips Ingenia dStream, R5.18, The Best, Netherlands) with a 20 channel body surface coil. A non-contrast-enhanced, retrospectively ECG-gated, SENSE accelerated 4D phase contrast (PC) sequence with referenced velocity encoding in all directions (4D Flow-MRI), adaptive respiratory gating and imaging parameters Venc=120cm/s and FA=7° was reconstructed to an isotropic resolution of 2mm and 20 time frames. 2D PC sequences aligned orthogonal to the main (MPA), left (LPA) and right (RPA) pulmonary artery were acquired with 40 heart phases during breathholds with Venc=100cm/s (MPA) and 120cm/s (LPA and RPA). A CINE bSSFP sequence transversal to the body axis covering the entire right ventricle was acquired for quantification of RVSV. 2D and 4D PC sequences were repeated after each volunteer on a static phantom.
Data analysis: Flow data were analyzed using GTFlow (v2.2.6, GyroTools, CH). Geometric information from 2D planes were used to exactly match the 4D plane positions (Fig. 1) and were subsequentially transferred to phantom data for error determination. Stroke volume (SV) and maximal velocity (Vmax) were evaluated. RVSV was determined per clinical standard according to the Simpson method using Extended MR WorkSpace (v2.6.3.5, Philips).
Statistical analysis: Results were given as mean ± standard deviation. Bland-Altman analysis and Student’s t-test were performed to compare the different methods. COM-analysis postulated SV in the MPA to equal the sum of SV in RPA and LPA (SVMPA = SVLPA + SVRPA).
1. Reiter G, Reiter U, Kovacs G et al. Blood Flow Vortices Along the Main Pulmonary Artery Measured with MR Imaging for Diagnosis of Pulmonary Hypertension. Radiology. 2015;275(1):71-9.
2. Barker AJ, Roldan-Alzate A, Entezari P et al. Four-Dimensional Flow Assessment of Pulmonary Artery Flow and Wall Shear Stress in Adult Pulmonary Arterial Hypertension: Results from Two Institutions. Magn Reson Med. 2015;73(5):1904-13.
3. Chernobelsky A, Shubayev O, Comeau CR et al. Baseline Correction of Phase Contrast Images Improves Quantification of Blood Flow in the Great Vessels. J Cardiovasc Magn Reson. 2007;9(4):681-5.
4. Giese D, Haeberlin M, Barmet C et al. Analysis and Correction of Background Velocity Offsets in Phase-Contrast Flow Measurements Using Magnetic Field Monitoring. Magn Reson Med. 2012;67(5):1294-302.
5. Frydrychowicz A, Wieben O, Niespodzany E et al. Quantification of Thoracic Blood Flow Using Volumetric Magnetic Resonance Imaging with Radial Velocity Encoding: In Vivo Validation. Invest Radiol. 2013;48(12):819-825.
6. Bollache E, Van Ooij P, Powell A et al. Comparison of 4d Flow and 2d Velocity-Encoded Phase Contrast MRI Sequences for the Evaluation of Aortic Hemodynamics. Int J Cardiovasc Imaging. 2016;32(10):1529-41.
7. Lankhaar JW, Hofman MB, Marcus JT et al. Correction of Phase Offset Errors in Main Pulmonary Artery Flow Quantification. J Magn Reson Imaging. 2005;22(1):73-9.
Fig. 1: 3D Streamline visualization at peak systole.
Positioning of 2D and 4D contours in main (MPA), right (RPA) and left (LPA) pulmonary artery is demonstrated.
Tab. 1: Stroke volume (SV) and maximal velocity (Vmax) in 4D and 2D measurements of the main pulmonary artery
avg diff: average difference, uncorr: uncorrected, corr: phantom corrected, bold type = significant
Tab. 2: Stroke volume (SV) of the main pulmonary artery in 4D and right ventricular stroke volume (RVSV)
avg diff: average difference, uncorr: uncorrected, corr: phantom corrected
Tab. 3: Conservation of mass analysis
SV: stroke volume, MPA: main pulmonary artery, RPA: right pulmonary artery, LPA: left pulmonary artery, avg diff: average difference, uncorr: uncorrected, corr: phantom corrected, bold type = significant