Qiang Zhang1, Jia Ning1, Shuo Chen1, Zhensen Chen1, Xihai Zhao1, Chun Yuan2, and Huijun Chen1
1Center for Biomedical Imaging Research, Department of Biomedical Engineering, School of Medicine, Tsinghua University, Beijing, People's Republic of China, 2Department of Radiology, University of Washington
Synopsis
The purpose of this study is to compare the MRA image contrast of SNAP
and TOF sequences in in-vivo cerebral artery images and explore the explanation
of the observation by theoretical simulation. We found TOF MRA showed better artery-background
contrast than SNAP MRA in proximal large cerebral artery for all cases, but worse
in distal small cerebral artery for most of the cases. The theoretical simulation
considering the blood velocity and the blood travel distance in imaging slab further
validated this observation, suggesting the contrast difference between SNAP and
TOF were related to blood velocity and travel distance.
Introduction:
MR 3D time of flight (TOF)
has been widely used as a non-contrast MR angiography (MRA) technique [1]. Recently,
SNAP technique [2] has
been proposed as a new non-contrast MRA technique, which can detect both
luminal stenosis and intra-plaque hemorrhage in patients with atherosclerosis.
A recent study showed that SNAP may have better small artery delineation
capability than TOF [3].
However, the study only focused on middle cerebral artery (MCA) and its
branches and lacked theoretical analysis of the MRA signal and contrast. This
study aims to explore the MRA contrast difference between 3D TOF and SNAP using
in-vivo data and theoretical simulation. Methods:
In-vivo
contrast comparison: In this retrospective study, SNAP and
TOF of 30 stroke patients (19 males, mean$$$\pm$$$SD age: 45.73$$$\pm$$$13.36
years) were included for analysis. All images were acquired on a 3T Philips
scanner (Achieva, Philips Healthcare, Best, The Netherlands). The imaging
parameters of TOF and SNAP were shown in Table 1. The coverage of TOF in axial
direction was cropped as the same with SNAP and the in-plane resolution of both
images was resampled to 0.5mm. Then MRA were generated by minimum/maximum
intensity projection along axial direction for SNAP/TOF. The centerlines of a proximal
large cerebral artery and a distal small artery were blindly delineated in the TOF
and SNAP MRA (Fig. 1) by one reviewer. Then the mean intensity ($$$I_{artery}$$$)
along the centerline was calculated. The background regions of interest (ROIs)
were also drawn on both SNAP and TOF MRA images. The arteries within the background
ROI were excluded by setting an intensity threshold. Then, the mean intensity
of the left pixels in background ROI ($$$I_{background}$$$) was calculated. Thus,
the contrast to noise ratio (CNR) can be given as $$$abs(I_{artery}- I_{background})/noise$$$,
where the $$$noise$$$ was calculated as
the standard deviation of background ROI excluding arterial pixels. The CNRs of
the proximal large cerebral artery and distal small cerebral artery between
SNAP and TOF MRAs were compared using paired t-test. Simulation: In the
traditional simulation model of SNAP and TOF, the blood was usually assumed fresh
when imaging [2, 4]. In this study, a
more realistic simulation was performed by considering the multiple excitations
of the blood spins under different flow velocities and different travel distances
(L1) that the spin has flown through within the imaging slab. The image
contrast between artery and background in the MRA image is calculated as $$$abs(I_{artery}-
I_{background})$$$, assuming same noise level for SNAP and TOF MRA. The
following parameters were used for simulation: blood T1=1550ms, blood T2=275ms [5], white matter T1=832ms,
white matter T2=110ms [6], blood velocity ranging
from 10 cm/s to 60cm/s, and L1 ranging from 5mm to 300mm.Result:
The
MRA CNR comparison of the in vivo images is shown in Fig. 2. The CNR of
proximal large cerebral arteries were significantly higher than distal small
cerebral arteries in both SNAP (p<0.001) and TOF (p<0.001). For proximal
large cerebral artery, the CNR of TOF MRA was larger than that of SNAP MRA in
all subjects (p<0.001). However, for distal small cerebral artery, the CNR of
SNAP MRA was larger than that of TOF MRA in most of cases (24/30), and the difference
wasn’t statistically significant (4.9 vs 4.4, p=0.521). Fig. 1 also shows two
case in which SNAP and TOF have better distal small artery delineation than the
other one respectively. The simulation results further explained the in-vivo
results (Fig. 3). In proximal arteries, where blood travel distance is short, TOF
MRA has better CNR at all simulated blood velocity (Fig. 3b). On the other
hand, in distal arteries, where blood travel distance is long, the SNAP MRA has
better CNR when blood velocity is high enough, but TOF MRA may have better CNR
when blood velocity is low (Fig. 3b). Overall, the arteries with shorter blood
travel distance has better CNR (Fig. 3a).Discussion and Conclusion:
This
study showed the difference of MRA CNR between SNAP and TOF in in-vivo experiment,
and further explained the cause of difference via simulation. This study found
that SNAP MRA have lower CNR in large proximal cerebral arteries, and usually
have higher CNR in small distal cerebral arteries. More importantly, these
results suggested that the CNR difference between TOF MRA and SNAP MRA may be
an indicator of the blood flow velocity with known cerebral artery position.
For example, our simulation results indicated that the blood flow velocity may
be low in those six patients whose distal small arteries had better CNR in TOF
MRA than SNAP MRA.Acknowledgements
No acknowledgement found.References
1. Heiserman, J.E., et al.,
Intracranial vascular stenosis and
occlusion: evaluation with three-dimensional time-of-flight MR angiography.
Radiology, 1992. 185(3): p. 667-673.
2. Wang, J., et al., Simultaneous Non-contrast Angiography and
intraPlaque hemorrhage (SNAP) imaging for carotid atherosclerotic disease
evaluation. Magnetic resonance in medicine : official journal of the
Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in
Medicine, 2013. 69(2): p. 337-345.
3. Wang, J., et al., In Vivo Validation of Simultaneous
Non-Contrast Angiography and intraPlaque Hemorrhage (SNAP) Magnetic Resonance
Angiography: An Intracranial Artery Study. PLoS ONE, 2016. 11(2): p. e0149130.
4. Dumoulin, C.L., et al., Three-dimensional time-of-flight magnetic
resonance angiography using spin saturation. Magnetic Resonance in
Medicine, 1989. 11(1): p. 35-46.
5. Noeske, R., et al., Human cardiac imaging at 3 T using phased
array coils. Magnetic Resonance in Medicine, 2000. 44(6): p. 978-982.
6. Wansapura, J.P., et al., NMR
relaxation times in the human brain at 3.0 tesla. Journal of Magnetic
Resonance Imaging, 1999. 9(4): p.
531-538.