Andrew David Scott1,2, Sonia Nielles-Vallespin1,3, Pedro Ferreira1,2, Zohya Khalique1, Dudley Pennell1,2, and David Firmin1,2
1Cardiovascular Biomedical Research Unit, Royal Brompton Hospital, London, United Kingdom, 2National Heart and Lung Institute, Imperial College London, London, United Kingdom, 3National Heart Lung and Blood Institute, National Institutes of Health, Bethesda, MD, United States
Synopsis
Cardiac diffusion
tensor imaging (cDTI) is a novel non-invasive method of interrogating myocardial
microstructure that has seen a recent surge in interest. Many of the most
interesting clinical results were obtained using stimulated echo acquisition
mode (STEAM) imaging at multiple cardiac phases. Recently however, spin-echo cDTI
with second order motion compensated diffusion gradients (M012-SE) was
proposed. In this study we report results of a comparison of M012-SE and STEAM
imaging in multiple cardiac phases at 3T in 15 healthy subjects with matched
sequence parameters. While M012-SE provides comparable quality data in systole,
STEAM is the more reliable technique in diastole.
Introduction
Stimulated echo
acquisition mode (STEAM) cardiac diffusion tensor imaging (cDTI) is robust, reproducible1,2
and has recently produced interesting clinical results3. However
STEAM is signal to noise inefficient and the influence of strain on the
measured diffusion is not fully understood. Alternatively, spin-echo (SE) with
acceleration and velocity compensated diffusion gradients (M012) has been
demonstrated and compared to STEAM at 1.5T in systole using high-performance gradient
systems4,5. In previous work we compared STEAM to M012-SE6
at 3T for multiple cardiac time points, but subsequent investigations discovered
a slice thickness inconsistency between the sequences. Here we compare
the sequences using matched slice thicknesses and sequence parameters on a
standard clinical 3T scanner at three cardiac phases.Methods
STEAM and M012-SE
sequences were implemented with identical EPI readouts and matched slice
thicknesses(figure 1). Imaging was performed on a Siemens Skyra 3T scanner
(40mT/m, 200T/m/s) at 2.8x2.8x8mm3 (1.4x1.4x8mm3
reconstructed), SENSE x2, read field of view (FOV) 360mm and phase FOV 135mm
(using a slice selection gradient on the second RF pulse). Both sequences used binomial
(1-2-1) water excitation, bmain=450smm-2, bref=150smm-2
and 6 directions. STEAM imaging used TE=23ms, TR=2RR-intervals and 8 averages.
M012-SE imaging used TE=75ms, TR=1RR-interval and 16 averages. Imaging was
performed in an agar phantom (T1=1090ms, T2=51ms, 50 averages) to the compare relative
SNR of the two sequences to theoretical values5 and in 15 volunteers (9 male, 24 [20–36]years, median [range]). In-vivo data was acquired in a mid-ventricular short-axis
slice during breath holding with diffusion encoding timed to end-systole,
diastasis and the systolic sweet-spot7. Previous analysis of DENSE
strain data acquired in 13 volunteers identified the systolic sweet-spot as 150ms
from the R-wave8. Data was processed using in-house Matlab software
to produce pixel-wise maps of helix angle (HA), absolute sheetlet angle based
on the second eigenvector (E2A), transverse angle (TA), mean diffusivity (MD),
fractional anisotropy (FA). A linear regression of HA with transmural depth was
used to compare HA between sequences. HA maps were also visually scored 0-4
based on <50%, 50-75%, 75-95% and >95% respectively of the myocardium
demonstrating the expected transmural variation. The standard deviation of TA
and the R2 of the linear regression of transmural HA variation were
used as additional data quality measures based on the assumptions of a linear
transmural HA distribution and an approximately constant TA.Results
The measured SNR ratio
(M012-SE/STEAM, mean signal/temporal standard deviation) of 1.87 in the agar phantom
was similar to the theoretical value of 1.97. cDTI data was of sufficient
quality for analysis in all STEAM acquisitions, but 1 systolic, 1 sweet-spot and
4 diastolic M012-SE acquisitions scored 0 and were excluded from further
analysis. Figure
2 shows example parameter maps and figure 3
compares average left ventricular values. MD is lower (p<0.01) in all phases
and FA is higher at the sweet-spot and in diastole using STEAM (p<0.01). There
was a reduction in E2A mobility (systole-diastole) using M012-SE; E2A was
higher in systole and lower in diastole using STEAM (p<0.01 at all phases).
Transmural HA gradient was significantly different at the sweet-spot
(p<0.01). Figure 4 shows the HA map visual scores, which were significantly
lower for M012-SE at the sweet-spot (median STEAM score 3 vs. M012-SE score 2, p=0.004).
HA map visual score was found to positively correlate with RR-interval for
M012-SE in diastole (Spearman Rho=0.64, p=0.009). Figure 5 shows a higher TA
standard deviation at the sweet-spot and in diastole and a reduced R2
of the linear regression of HA with transmural depth at the sweet-spot
(p<0.01) using M012-SE, which is suggestive of poorer quality data5.Discussion
The previously
described5 systolic differences in MD and FA between STEAM and M012-SE primarily
due to differences in diffusion time (~1000ms for STEAM vs. ~30ms for M012-SE) are
maintained at 3T in systole, diastasis and the sweet-spot, when imaging
parameters are matched between sequences. Data were of comparable quality
between the sequences in systole, but data quality measures were better using
STEAM at the sweet-spot despite the higher SNR available using M012-SE. M012-SE
was unsuccessful in 27% of diastolic acquisitions, due to either missed ECG
triggers as a result of the longer diffusion encoding or an insufficient
duration of stasis, which could not accommodate the longer diffusion gradients required by M012-SE. Conclusion
There are systematic
differences between cDTI parameters obtained using STEAM and M012-SE. Although
both sequences can be used to provide cDTI data at peak systole, further work
is required to improve the reliability of M012-SE in diastole. Patient studies
are required to assess whether the differences between the sequences can
provide complementary microstructural information.Acknowledgements
Not applicable.References
1.
Reese MRM
1995.DOI:10.1002/mrm.1910340603.
2.
Nielles-Vallespin
MRM 2012. DOI:10.1002/mrm.24488?
3.
Ferreira JCMR 2014. DOI:10.1186/s12968-014-0087-8
4.
Stoeck MRM
2015. DOI:10.1002/mrm.25784
5.
Von
Deuster MRM 2015. DOI:10.1002/mrm.25998
6.
Scott ISMRM
2016. P3115.
7.
Tseng MRM
1999. DOI: 10.1002/(SICI)1522-2594(199908)42:2<393::AID-MRM22>3.0.CO;2-F
8.
Scott SCMR/ISMRM Workshop on
Quantitative CMR 2016. W17.