Shimul Chandra Saha1, Alena V. Shchelokova2, Ioannis Sotiriou1, Alexey P. Slobozhanyuk2,3, Maria Koutsoupidou1, Elena Cellitti1, Helena Cano-Garcia1, Pavel A. Belov2, Andrew Webb4, George Palikaras1, and Efthymios Kallos1
1MediWiSe| Medical Wireless Sensing Ltd, London, United Kingdom, 2Department of Nanophotonics and Metamaterial, ITMO University, St. Petersburg, Russian Federation, 3Nonlinear Physics Center, Australian National University, Canberra, Australia, 4C.J. Gorter High Field Magnetic Resonance Center, Leiden University, Leiden, Netherlands
Synopsis
We present in vivo results from a metasurface structure comprising an array of brass
wires embedded in a high epsilon and low loss medium. The metasurface was used
to scan humans in a 1.5T MRI scanner and demonstrated enhancement of the signal-to-noise
ratio up to 200% in the area-of-interest close to the metasurface.
Introduction
All
clinical scanners at 1.5 Tesla use a body coil for transmission, which provides
a uniform global excitation field (B1+). However, there
are many clinical applications where increasing the local excitation field would be advantageous both in terms of
reducing the specific absorption rate (SAR), which is often maximum in areas of
the body distant from the imaging region of interest, and also reducing signal
from areas which produce motion artifacts, e.g. abdominal motion in spine
scanning. The use of metamaterials made from split ring resonators1
has been shown to improve the local transmit efficiency , but these types of
structures are very complicated to construct and fine tune1.
Dielectric pads made of high permittivity ceramics2 have also been
used to improve field uniformity and SNR, but typically work at higher fields.
In this work we have developed a metasurface which consists of a simple flat
array of thin non-ferrous metal wires embedded in a low loss dielectric. This
metasurface was made of brass wires embedded in deionized water and contained
in an acrylic box. The device has been used to image human knees in a 1.5T
scanner, and showed an increase in the transmit field efficiency of up to 200%.Methods
Lab grade deionized water was used as the high
permittivity material (ϵr = 80 at 64 MHz). A 14×2 wire matrix with
length 37.2 cm was assembled on thin hollow 3D-printed ABS plastic holders
(Fig. 1), and the resonance frequency of the first eigenmode of the metasurafce3
was optimized by simulating in 3D electromagnetic software (CST Microwave
Studio). The resonant structure was contained in a watertight & mechanically
robust box made of 5 mm thick acrylic sheets filled with DI water (Fig. 1). The
metasurface was used for scanning on a
Philips Ingenia 1.5T system. The transmit efficiency and signal-to-noise ratio
(SNR) were assessed in different set ups, (i) body coil transmit, four channel
array receive without metasurface, and (ii) body coil transmit, four channel
array receive with metasurface. The resonant frequency of the metasurface eigenmode was measured using a pick-up loop both
outside and inside the magnet, and minimal detuning was detected. T1-weighted
gradient echo images of the knee were acquired with the following parameters:
TR 9.5 ms,
TE 4.8 ms, data matrix 208 x 168 x 75, spatial resolution 0.96 x 0.96 x 2 mm, and
acquisition time 1 minute 30 seconds.Results, discussion and conclusion
A total of six human knee scans were
acquired and representative results are shown in Fig. 2. In order to demonstrate the local transmit
field increase, a low flip angle (FA) (4 degrees) gradient echo sequence was
run with and without the metasurface, using the body coil in transmit and four channel array in receive mode. In
addition, a scan with power (FA-25) optimized for the case without the metasurface
was also performed. It can be seen that
the image with the metasurface (Fig. 2c) has much higher SNR (Fig. 2a). The pixel-by-pixel
image intensities are shown in Fig. 3, which demonstrate a maximum 350%
increase with the metasurface. In order
to quantify the improvement, the SNR for all cases are shown in Fig. 4. An
average of 200% improvement is seen in the knee area close to the metasurface
averaged over the six subjects. The improvement of SNR is due to the increase in receive sensitivity by the
metasurface. The SAR of a knee phantom4 was
simulated in CST microwave studio using the appropriate body coil5. For 1W peak power, the maximum SAR with and
without metasurface is 0.044 and 0.0018 W/kg respectively (Fig. 5),
representing a factor-of-twenty higher with the metasurface in place. Since the
transmit power required is reduced by a factor of almost forty, this means that
the SAR can be halved with the metasurface for similar scan performance. In the
current work the improvement was evident mostly at the centre part of the
device and closer from the top surface. Future
design would be made of ceramics and a flexible structure which can conform to
the torso or head and enhance the performance at much greater depths.Acknowledgements
This
work was financially supported in part by Innovate UK under Project 710771
& Russian Science
Foundation under grant 15-19-20054. Authors would like to
thank technical & research staff (Paul de Bruin and Rita Schmidt) of C.J.
Gorter High Field Magnetic Resonance center for their help.References
1.
Lopez M A, Freire M J, Algarin
J M, Behr V C, Jakob P M, and Marques R. Nonlinear split-ring
metamaterial slabs for magnetic resonance imaging, Applied Physics
Letters. 2011; 98:133508.
2.
Brink W M, and Webb A G. High permittivity pads reduce specific
absorption rate, improve B1 homogeneity and increase contrast-to-noise ratio
for functional cardiac MRI at 3T, Magnetic resonance in Medicine. 2014; 71: 1632-1620.
3.
Slobozhanyuk A P, Poddubny A N, Raaijmakers A J E,
van den Berg C A T, Kozachenko A V, Dubrovina I A, Melchakova I V, Kivshar Y S,
and Belov P A. Enhancement of Magnetic Resonance Imaging with Metasurfaces. Advanced
Materials. 2016; 28(9):1832-1838.
4.
Andreuccetti D, Fossi R and Petrucci C: An
Internet resource for the calculation of the dielectric properties of body tissues
in the frequency range 10 Hz - 100 GHz. Website at http://niremf.ifac.cnr.it/tissprop/.
5.
CST microwave studio. MRI model from Institute of Biometrics and Medical
Informatics, Otto von Guericke University Magdeburg, Germany.