Oliver Kraff1, Britta M Hüning2, Andreas Deistung1,3, Viktor Pfaffenrot1, Andreas K Bitz4, Mark E Ladd1,4, and Harald H Quick1,5
1Erwin L. Hahn Institute for MRI, University Duisburg-Essen, Essen, Germany, 2Clinic of Pediatrics I, Neonatology, University Hospital Essen, Essen, Germany, 3Medical Physics Group, Institute of Diagnostic and Interventional Radiology, Friedrich Schiller University Jena, Jena, Germany, 4Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany, 5High Field and Hybrid MR Imaging, University Hospital Essen, Essen, Germany
Synopsis
In
this retrospective study, log files of the RF safety watchdog system were
evaluated from 30 head exams performed at 7T in conjunction with the widely
available 32-channel Nova Medical head coil. Head local SAR as reported by the
PALI and the start/end times of each sequence were compared
for eight different sequence types. This study aimed at providing insights into
the expected RF exposure in UHF neurological protocols and may become useful in
simulations to study realistic exposure scenarios, as well as to determine
which sequences may be acceptable if a SAR reduction is required for a 7T exam.
Target Audience:
Radiologists
and physicists interested in MR safety and ultra-high field imaging.Purpose:
For
safety assessments of radiofrequency (RF) transmit coils or in combination with
implanted medical devices (IMD), temperature measurements and thermal
simulations are typically performed at maximum specific absorption rate (SAR) levels.
While from a conservative point of view this approach is reasonable, it rarely
reflects actual imaging procedures in terms of different SAR loads per sequence
and run times. In 2015, Langman et al. [1] reported that in neurological exams
at 1.5T high SAR sequences (> 2 W/kg) run for approximately 6-12 minutes of
a 32-minute exam, with a median head SAR of 1.3 W/kg. Furthermore, the authors
reported a median inter-sequence downtime of 23 seconds. At ultra-high fields (UHF)
of 7 Tesla and higher, SAR is a limiting factor and large effort is
put into in reducing the SAR load of clinically important sequences such as
turbo-spin echo (TSE) or time-of-flight (TOF). On the other hand, low flip
angle gradient echo sequences, for example susceptibility weighted imaging
(SWI), have demonstrated diagnostic value at 7T and have become key components
of head protocols, despite typically long acquisition times. Furthermore,
workflow at 7T still demands manual adjustments and more interaction from the operator,
which prolong inter-sequence downtimes. In this regard, this study aimed at
providing insights into the expected RF-exposure in UHF head protocols.Methods:
In this
retrospective study, log-files of the MR system’s (Magnetom 7T, Siemens
Healthcare GmbH, Germany; software version VB17) RF safety watchdog (RFSWD) were evaluated from 30 head exams. All measurements were performed with a widely available 1-channel transmit/32-channel receive head coil (Nova Medical
Inc., Wilmington, MA). For each subject, at least one T1-weighted structural
scan was included (MP(2)RAGE) while the rest of the protocol varied according
to the scope of the individual research study (e.g. vascular diseases,
cerebellar development, spectroscopy, fMRI). Table 1 gives an overview of the applied
sequences. The log-files were parsed to extract head local SAR (most critical
aspect) as reported by the power absorption limiter (PALI) for the 360 s
checking interval, as well as the start/end times of each sequence. The PALI values also
reflect different coil loading and transmitter adjustments. The sequences were
grouped according to their SAR load: low
SAR was defined as head local SAR of up to 4 W/kg, medium SAR reflects values greater than 4 W/kg and less than 8
W/kg, and high SAR was defined as head
local SAR of 8 W/kg and higher.Results:
All
sequence runs remained within limits set for the normal operating mode, i.e. 10
W/kg local SAR [2]. The mean number of actual (structural, functional, or
spectroscopic) sequences per subject was 5.4. On average, 2.5 scans per subject had
a low SAR profile, 1.7 had a medium SAR profile, and 1.2 scans were high SAR level (none for spectroscopic and fMRI studies). In 17/30 (57%)
datasets at least one high SAR sequence was included. Figure 1 shows the mean
cumulative duration for the three levels of SAR load. In detail, Figure 2
provides an overview of the SAR load per type of sequence. The two major
constituents of a UHF neurological protocol, SWI and MP(2)RAGE, never exceeded
4 W/kg and had the longest duration. On the other hand, the 2D TSE yielded
highest head local SAR values at minimum run times. The mean total acquisition
time was 1 h 2 min 9 sec ± 11
min 38 sec, whereof on average 4 min 52 sec ± 1
min 35 sec could be attributed to preparations (adjustments, localizer scans, B1
mapping for transmitter adjustments) prior to the very first structural scan.
The mean inter-sequence
downtime was 30 seconds, while for MR spectroscopy this was increased to 11 min
51 sec due to repeated B0 shimming.Discussion:
This
analysis of the head exams at 7T showed that more than half of the subjects
were exposed to at least one high SAR sequence. However, the mean cumulative
time of the high SAR load per subject was only a quarter of the total exposure
duration. For more than half of the imaging duration low SAR sequences were
typically applied. The results obtained at 7T were similar to the results
reported by Langman et al. for 1.5T [1]. The presented analysis may become
useful in simulations to study realistic exposure scenarios, as well as to determine
which sequences may be acceptable for a subject with implants that requires a
SAR reduction for a safe 7T exam.Acknowledgements
No acknowledgement found.References
[1] Langman
DA, et al., What is the SAR for Routine Clinical MRI Exams at 1.5T?, Proc. 23rd
ISMRM (2015), abstract 302.
[2] IEC.
60601-2-22 Medical electrical equipment – Part 2-33. Edition 3.2, 2015.
[3] Wrede
KH, et al., Caudal image contrast inversion in MPRAGE at 7 Tesla: problem and
solution. Acad Radiol. 2012;19(2):172-8.
[4] Dammann
P, et al., The venous angioarchitecture of sporadic cerebral cavernous
malformations: a susceptibility weighted imaging study at 7 T MRI. J Neurol
Neurosurg Psychiatry. 2013;84(2):194-200.
[5] Poser
BA, et al., Three dimensional echo-planar imaging at 7 Tesla. Neuroimage. 2010;51(1):261-6.
[6] Johst
S, et al., Time-of-flight magnetic resonance angiography at 7 T using venous
saturation pulses with reduced flip angles. Invest Radiol. 2012;47(8):445-50.