Amin Douiri1, Stefan Scholz1, Wolfgang Görtz1, Jakob Kreutner1,2, and Gregor Schaefers1,2
1MR:comp GmbH, Gelsenkirchen, Germany, 2MRI-STaR GmbH, Gelsenkirchen
Synopsis
In this study
the suitability of a miniature accelerometer for vibration measurements
according to ISO/TS 10974 was investigated. Gradient-induced vibration of a
titanium disc was measured both with an electric accelerometer and an optical displacement
sensor and the obtained datasets were compared. Taking the characteristics of
the setup into account, the results indicate that the used type of
accelerometer is suitable to meet the specifications given in ISO/TS 10974.
Introduction
Vibratory forces caused by the interaction of
the gradient-induced eddy current magnetic moments of an active implantable
medical device (AIMD) with the static magnetic field (B0) are a
well-known safety hazard in MR imaging. Following the specifications given in
ISO/TS 10974 Clause 10, the use of vibrometers or other optical measurement
systems is recommended [1] which requires a complex setup and calibration in
order to generate reliable measurement results. In this investigation a
comparative vibration measurement of a titanium disc as a dummy device under
test (DUT) was conducted to assess the suitability of an adhesive miniature
accelerometer for measuring gradient-induced vibration. This approach has the
advantage of a significantly lower noise level (see FIG. 1) and
secondly excludes artefacts caused by an insufficient alignment. The measurement
on the other hand could be influenced by the extra weight of the sensor and its
fixation. The goal of this study was to investigate under what circumstances an
electric accelerometer can replace an optical measurement setup given the specifications
in ISO/TS 10974. An optical displacement sensor was used to generate reference
values.
Method
Measurements were
conducted in a clinical 1.5T Avanto and a 3T Verio MR scanner (Siemens
Healthineers, Germany).The DUT (Titanium disc, diameter = 46.7 mm; weight =
4.14 g) was suspended by strings (aligned parallel to the X-axis) and placed in
coronal orientation inside the MR bore at a location with the highest
achievable gradient field in Y-direction. The setup was fixed onto the patient
table decoupled from the vibrations of the MR scanner. The accelerometer “MRacc”
(MRI-Tec, Germany, height = 3.5 mm; diameter = 2.2 mm; weight = 0.18 g) was fixed
on top of the main surface of the titanium disc while the optical displacement
sensor “MRI-VRS 100” (MRI-Tec, Germany) was placed approximately 3 mm
below the DUT’s surface (FIG.
2). EPI sequences were
applied and both the displacement and the acceleration were measured. The gradient
field exposure level was measured simultaneously using a 3-planes search coil
according to IEC 60601-2-33 [2]. Data was recorded using a data card and
software (National Instruments, USA) with a sampling rate of 100 kSa/s. Additional
measurements without the DUT were conducted to investigate induced vibration into
the accelerometer and the whole test setup. The measured acceleration signal was numerically
double-integrated and compared to the data obtained from the displacement
sensor. Both datasets were spectrally analyzed with FFT and the resulting
frequency spectra were again compared.
Results and Discussion
The mean deviation of the detected peak frequencies
between the displacement sensor and the accelerometer was 0.06% with a maximum
deviation of 0.5% at 685 Hz. Detected peak frequencies of both sensors are
plotted versus each other in FIG. 3. A very high
correlation between the measurements can be seen (r²=1). Mean discrepancy
between measured displacement amplitude and the double-integrated acceleration
was 27.7% with a maximum deviation of 39.4%. For all sequences the optical
sensor provided higher amplitudes compared to the integrated results of the
accelerometer (see FIG. 4 and FIG. 5).This was expected
due to the optical sensor being positioned at the opposite side of the DUT as
the accelerometer (see FIG. 2). The additional
weight of the accelerometer and its fixture causes a shift of the rotational
axis away from the center line of the disc towards the accelerometer thus
resulting in a higher deflection detected by the optical sensor. Vibration of
the test setup was determined at 350 Hz (the largest deflection was measured at
this frequency with the 1.5 T Avanto) with a detected displacement of 87 nm which
was 0.82% of the DUT’s deflection at the same sequence. Therefore the vibration
of the setup was considered negligible.Conclusion
This investigation shows that the presented
miniature accelerometer is suitable for measurements in accordance with the
specifications given in ISO/TS 10974. This can be concluded by the excellent
frequency response of the sensor. The small dimensions of the sensor compared
to generic vibrometer setups additionally enable the accessibility of higher
gradients inside the MRI bore. For precise measurements however careful
considerations of the influence of the sensor and its fixture on the mass
moment of inertia of the DUT are necessary. The approach to measure vibration
with an electronic accelerometer instead of an optical sensor should therefore
be included in the recommendations given in ISO/TS 10974 Clause 10.6.3.1.Acknowledgements
No acknowledgement found.References
[1] Technical specification ISO/TS 10974 “Assessment of the
safety of magnetic resonance imaging for patients with an active implantable
medical device”, 2nd Edition, Blender Version 3.
[2] Technical specification
IEC 60601-2-33 IEC. 2010. Ed.3.1. Medical Electrical Equipment - Part 2-33:
"Particular requirements for the basic safety and essential performance of
magnetic resonance equipment for medical diagnosis."