Julie Kabil1,2, Alexia Missoffe1,2, Pierre-André Vuissoz1,2, Cédric Pasquier1,2,3, and Jacques Felblinger1,2,4,5
1Université de Lorraine, IADI, Nancy, France, 2INSERM, U947, Nancy, France, 3Healtis, Nancy, France, 4CHRU Nancy, France, 5INSERM, CIT-1433, Nancy, France
Synopsis
Patients carrying
multiple leads like abandoned pacemaker leads are still denied MRI scans, as their
situation is not currently covered by the existing safety guidelines. Therefore,
to assess the impact of lead coupling regarding radiofrequency-induced heating,
temperature measurements were performed on different types of simplified leads
and coupling factors were introduced to quantify and evaluate the phenomenon. The
lead coupling can have a significant impact on temperature and can either
induce higher or lower temperatures compared to the case when the leads are
alone, and thus should be considered in future MRI safety standards.
Purpose
Radiofrequency (RF)
safety in MRI for medical devices like pacemakers1 is of great
importance as the worldwide population is aging and requiring more MRI scans2.
There is a need to assess the safety of patients carrying multiple leads3-4,
as it is not yet covered by the existing MRI compatibility standards5.
Thus, the purpose of this study is to experimentally observe the lead coupling regarding
RF-induced heating with different simplified leads, to have a better insight on
the phenomenon occurring and thus approach safety guidelines suitable for every
patient carrying multiple leads.Methods
Two 40-cm long
simplified leads were made of straight hardened steel wires of 1.5 mm diameter with a 0.375
mm-thick polyolefin insulation (heat-shrink tube). As these hardened steel
wires which were easily available and perfectly straight were also magnetic, two
other leads were made the exact same way except with non-magnetic and
non-perfectly straight stainless steel wires, to check if this property
affected the RF-induced heating and coupling. To verify the occurrence of a
coupling phenomenon in different lead models, two other leads of the same
length were made using a coaxial cable with its metallic shield removed; the stainless
steel wire diameter was 0.92 mm and the PTFE insulation thickness was 1 mm. All
leads were left bare along 5 mm at their extremity to represent the
electrodes. In each set of leads, one lead was also left bare at the other
extremity over a length of 2 mm, called the “uncapped” lead and the other,
insulated, was the “capped” one, to have different termination conditions. Different configurations were studied for each
set of leads (Figure 1), embedded at the right side of an ASTM gel phantom6
and supported by a dedicated 3D-printed plastic support and smaller
3D-printed plastic guides along the leads (Figure 2). When both leads were together, they were spaced 2.5 mm. The experiments were
conducted so that there was no lead or temperature probe displacement during
the measurements on the same lead (alone or in presence of the other lead). A fiber
optic temperature probe (NEOPTIX T1, Neoptix, Quebec City, Canada) was
positioned orthogonally on the same 3D-printed support to ensure the accuracy
of measurements at the extremity of the leads. A 3’24 duration FSE-XL ASTM6 RF-heating sequence was then performed
on the phantom inside a 1.5 T GE Signa HDx MRI scanner (General Electric
Healthcare, Milwaukee, WI, USA) and the temperature was acquired in real time
with the temperature probe software. The experiments were repeated, some
of them several months apart, to assess the repeatability of the observations. The
first leads are referred to as leads #1, the non-magnetic ones used to check the
influence of this property as leads #2, and the leads made from coaxial cables as
leads #3. A coupling factor CF was calculated for each lead as the ratio
between the temperature elevation of the leads when they are both present and when
they are alone. Results
The temperature elevations and coupling factors are reported in Table 1. The
mean value for CFcapped in the #1 leads set is 0.63±0.05 (CI95%),
and the mean value for CFuncapped in the same leads set is 0.93±0.07
(CI95%). Depending on the experiment, CFuncapped values can
be higher or less than 1. Discussion
The difference
in absolute temperature elevations between two measurements for the same leads
can mainly arise from the position of the temperature probe at the extremity of
the leads in the gel phantom. However, the experimental setup ensured a proper
repeatability of the coupling factors values. The magnetic property of leads #2
does not seem to have an influence on the temperature elevations, thus leads #1
experiments are relevant to study lead coupling regarding radiofrequency. The
coupling factors values on these simplified leads suggest that coupling can
have a significant impact on temperature elevation, up to 40%. Moreover, the
different CFuncapped values show that lead coupling can either induce
higher or lower temperatures: this highlights the complexity of this phenomenon
and indicates that the presence of another lead can have an effect
on RF-induced heating, and therefore should be studied carefully. Conclusion
This study highlights the complex nature of lead coupling in MRI, and shows
that the observations about lead coupling inducing higher or lower temperatures
could be essential for MRI compatibility safety standards of medical devices
when considering several leads. Experiments on real medical devices leads (for
example pacemakers’) will be performed to verify the impact of lead coupling in
more realistic leads. Acknowledgements
The authors thank the Région Lorraine and FEDER for financial support.References
1. Boilson BA, Wokhlu A, Acker NG, Felmlee JP,
Watson RE, Julsrud PR, et al. Safety of magnetic resonance imaging in patients
with permanent pacemakers: a collaborative clinical approach. Journal of
Interventional Cardiac Electrophysiology. 2012 Jan;33(1):59–67.
2. Philips, data on file. Based on Millennium research group reports
RPUS21LJ15; RPUS20SP15; RPGL12CR14.
3. Higgins JV, Gard JJ, Sheldon SH, Espinosa RE, Wood CP, Felmlee JP, Cha
YM, Asirvatham SJ, Dalzell C, Acker N, Watson RE, Friedman PA. Safety and
Outcomes of Magnetic Resonance Imaging in Patients with Abandoned Pacemaker and
Defibrillator Leads. Pacing and Clinical Electrophysiology, 2014;37: 1284–1290.
4. Mattei E, Gentili G, Censi F, Triventi M, Calcagnini G. Impact of capped
and uncapped abandoned leads on the heating of an MR-conditional pacemaker
implant: MR-Conditional Pacemaker Implant and Abandoned Lead. Magn Reson Med.
2015 Jan;73(1):390–400.
5. ISO/TS 10974:2012, Assessment of the safety of magnetic resonance
imaging for patients with an active implantable medical device.
6. ASTM F2182-11a, Standard Test Method for Measurement of Radio Frequency
Induced Heating On and Near Passive Implants During Magnetic Resonance Imaging,
2011.