Isabelle Saniour1, Gwenaël Gaborit2,3, Lionel Duvillaret3, Anne-Laure Perrier2, and Olivier Beuf1
1Univ. Lyon, CREATIS ; CNRS UMR 5220 ; INSERM U1206 ; INSA-Lyon ; UJM-Saint-Etienne ; Université Lyon1, Villeurbanne, France, 2Univ. Savoie-Mont-Blanc, IMEP-LAHC, Le Bourget-du-Lac, France, 3Kapteos, Sainte-Hélène-du-Lac, France
Synopsis
High-field MR systems bring
additional safety issues regarding the patient. One of main concern is the heating
effect caused by the high-frequency and high-power signals that can be picked
up by the conductors. It is thus required to identify these regions and to
quantify the electrical E-field knowing
that it is proportional to the square of the temperature elevation. For a
direct measurement, an optical probe was used to perform a mapping of the E-field distribution inside a birdcage
of a preclinical MRI. Experiments and simulations were found similar with E-field concentrations mainly located
close to the capacitors.
Introduction
The spatial resolution of MR images could be improved by
developing sophisticated radiofrequency (RF) transceiver coils or
by increasing the static magnetic field B0-strength1.
However, the fast and successful progresses of MR scanners could be the
reason for additional safety issues regarding the patient2, 3. The
specific absorption rate (SAR) is proportional to the square of B0 thus local heating is
caused by the transmission of high-frequency and high-power electrical E-field along conductors or passive
components e.g. capacitors. This requires a fast and an accurate assessment of the E-field concentrations in order to control the
SAR-parameter during MRI exam. Detecting of the distribution of the E-field inside MRI or near metallic
implants was already evaluated by dipole-based
probes inserted in a phantom4, 5 and numerical simulations6.
But, the metallic structure of the dipole probe may alter the electromagnetic
fields and thus limits the accuracy of the measurements. In this study, we propose an optical-based probe for E-field measurements and we
present a comparison between the simulated E-field distribution and the experimental map of the E-field using an optical-based probe inside a MRI birdcage.Methods
A finite element method
simulation was performed using COMSOL Multiphysics. The birdcage consists of two circular
conductive loop (end rings) where the capacitors are placed and nineteen
conductive elements referred to as rungs (Fig 1). The capacitors of the coil
are defined using lumped elements. The birdcage is linear thus only one
lumped port boundary is used for RF excitation (P = 138W). The birdcage was placed inside a cylindrical RF
shield. The experimental measurement
of the E-field was performed in a 4.7-T
preclinical MRI (Bruker) using an optical-based probe. The E-field probe operates based on Pockels effect; it consists of
isotropic crystal that changes its optical proprieties according to the
variation of the E-field. A laser
diode was simultaneously applied to the crystal. The polarization of the laser changes
with the applied E-field.
Optoelectronic devices were used to analyze the modulation of the polarization and
insure the optical-to-electrical conversion of the signal to be then visualized
by a spectrum analyzer. The probe measures the radial component of the E-field. It was calibrated outside the
MRI to determine the proportionality factor between the EO signal and the E-field. The proximal side of the birdcage
corresponds to the plane Z=0. On this plane, we have chosen nine measurement
points as shown in Fig 2. For
these nine points, we have measured the E-field
every 1cm by moving the probe along z-axis. The measurements points have been
used to plot 2D maps of the E-field
using Matlab language. For these experiments, a FLASH sequence with hermite shape RF pulse was
used.Results
The simulated results
show that the E-field is not uniform within
the birdcage. Figure 3 exhibits clearly the positions of the E-field concentrations regarding the
geometry of the birdcage. The high-concentrations areas (around 1900 V/m) are mainly
located near the lumped port and the capacitors. The concentrations-value
gradually decreases along the copper rungs to have the lowest value (100 V/m) in
the middle of the birdcage rungs/legs. Experimental maps are shown in the Fig 4. The E-field high-concentrations are located
at Z-values equal to 5 cm and 11 cm for all the slices. The E-field was measured with a factor of
proportionality equal to 417714.9 m-1. The
maximum value was around 1900 V/m and decreases progressively by approaching to
the center. Discussion
The
E-field measurements and simulations
were performed in air in order to determine the distribution of the E-field inside a birdcage and specifically near its walls without
any limitations due to the dimensions of a phantom4, 5. The experimental maps of the E-field are very close to the
simulation in terms of distribution and value. The high concentrations of the E-field
are in proximity of the birdcage wall and the weakest concentrations are in the
center. The simulation shows that the E-field high-concentrations are located
near lumped port and capacitors. The distance between the two rings is 56 mm,
therefore the high-concentrations in the experimental maps corresponds exactly
to the locations of the capacitors. Conclusion
We
proved that the optical-based probe has the ability to measure in air the E-field inside a birdcage within
picoseconds. This probe would also be used inside a phantom or a biological
tissue to control the SAR-parameter and could definitely replace thermal
sensors commonly used as an alternative method for SAR assessment but cause additional
measurement errors due to the conviction and the diffusion of the temperature in
the medium nearby. Acknowledgements
This work was
supported by the Rhône-Alpes region
and performed within the framework of LABEX-PRIMES (ANR-11-LABX-0063) of Lyon
University, and the program “Investissements d’Avenir” (ANR-11-IDEX-0007)
operated by the French National Research Agency.References
1. L.
Quettier, et
al. Expected magnetic field quality from the preliminary measurements
performed during the manufacturing of the Iseult/Inumac whole-body 11.7-T MRI magnet. IEEE Trans. Appl. Supercond. 2016;26(4):1–4.
2. P. Röschmann.
Radiofrequency penetration and absorption in the human body: Limitations to
high-field whole-body nuclear magnetic resonance imaging.
Med. Phys.1987;14(6):922–931.
3. D. I.
Hoult
et al. Sensitivity and power deposition in a high-field imaging experiment. J.
Magn. Reson. Imaging. 2000;12(1):46–67.
4. A.
Kangarlu, et al. Electric field
measurements and computational modeling at ultrahigh-field MRI. Magn.
Reson. Imaging. 2007; 25(8):1222–1226.
5. P.
Nordbeck, et
al. Spatial distribution of RF-induced E-fields and implant heating in MRI. Magn. Reson. Med. 2008; 60(2):312–319.
6. B. Park, et al. Analysis of conservative
and magnetically induced electric fields in a low-frequency birdcage coil. Journal of Electromagnetic Analysis and
Applications. 2013;5:271–280.