Minimally invasive interventions benefit from image guidance during instrument positioning. To ensure high image quality, MR-guided interventions are preferably performed in closed bore systems using MR markers and passive tracking sequences to localize and monitor the interventional device. In this work we accelerate a passive marker tracking sequence designed for needle-guided procedures using a simultaneous multi-slice excitation. With this technique the acquisition time is reduced by 33%, which results in a higher temporal fidelity while maintaining targeting accuracy.
The POCC tracking sequence1-3 detects the position of a passive, cylindrical marker filled with contrast agent and features a central hole for needle insertions. In its current implementation, the sequence tracks the marker’s position and orientation from two parallel slices (T1-weighted FLASH) oriented perpendicular to the marker’s symmetry axis so that the marker appears in a ring-like cross section. The position of the ring is automatically detected via a POCC algorithm and the calculated position information is used to subsequently align a targeting image (bSSFP) parallel to the marker’s symmetry axis, i.e. the planned needle trajectory. The tracking and targeting images are continuously acquired to automatically follow manipulations of the marker in real-time.
In this work the two parallel tracking slices are acquired simultaneously (i.e., acceleration of the tracking part by a factor of 2) by using the simultaneous acquisition method POMP2. The POMP excitation pulse is calculated from the superposition of two slice selective sinc-shaped pulses in the time domain, in which the phase of one of these pulses is modulated with a factor of $$$exp(-iφt) (φ=2πγGΔz; γ:$$$ gyromagnetic constant, $$$G:$$$ slice selection gradient). The phase modulation induces a spatial shift $$$Δz$$$ between the two slices. The signals from both slices are separated in the image domain by an additional ±180° phase-cycling which shifts the signal from one slice in phase encoding direction. POMP excitation (pulse duration = 0.8ms, time-bandwidth product = 1.6) was implemented on a 1.5T system (Siemens Symphony) and the slice profiles were measured in a homogeneous phantom (Fig. 1). Afterwards, the POMP pulse was inserted into the POCC tracking sequence (Fig. 2) replacing the conventional serial acquisition of the two tracking slices. The two versions of the tracking sequence were then compared during needle insertions in a 2% agarose gel that contained 24 fiducial targets (diameter: 7.8±0.6mm). A customized marker holder fixed to the patient table was used to manually manipulate the marker via a ball joint (Fig. 3). To compare both sequences the following experimental work-flow was done for each target:
(i) positioning of the marker in a center (i.e. neutral) position,
(ii) online targeting with either the existing (TA=2.3s for one cycle) or the POMP-based (TA=1.5s for one cycle) tracking sequence until the planned needle trajectory was aligned with the target (the two sequences were alternated with every other target),
(iii) needle insertion into the target under HASTE guidance using the planned needle pathway according to the respectively preceding tracking sequence.
The targeting duration and the lateral distance of the needle pathway to the target’s center point using reformatted views of high-resolution data set (3D GRE, 0.6×0.6×0.6mm3) were measured for each target (Fig. 4a-c). Finally, the sequence was tested in vivo by targeting different structures in the abdomen of a volunteer.
1. de Oliveira A, et al. Automatic passive tracking of an endorectal prostate biopsy device using phase-only cross-correlation. Magn Reson Med 59:1043–1050 (2008)
2. Krafft AJ, et al. Passive marker tracking via phase-only cross correlation (POCC) for MR-guided needle interventions: Initial in vivo experience. Physica Medica 29:607-614 (2013)
3. Zamecnik P, et al. Automated Real-time Needle-Guide Tracking for Fast 3-T MR-guided Transrectal Prostate Biopsy: A Feasibility Study. Radiology 273:879-886 (2014)
4. Glover G H J. Phase-offset multiplanar (POMP) volume imaging: A new technique. Magn Reson Imag 1:457–461 (1991)