Christiaan G. Overduin1, Axel J. Krafft2, Michael Bock2, Hans JF Langenhuijsen3, Sjoerd F.M. Jenniskens1, Jurgen J. Fütterer1,4, and Tom W.J. Scheenen1
1Radiology and Nuclear Medicine, Radboud University Medical Centre, Nijmegen, Netherlands, 2Radiology - Medical Physics, Medical Center University of Freiburg, Freiburg, Germany, 3Urology, Radboud University Medical Centre, Nijmegen, Netherlands, 4MIRA Institute for Biomedical Engineering and Technical Medicine, University of Twente, Enschede, Netherlands
Synopsis
This study
assessed the feasibility of 2D UTE-based MR thermometry of frozen tissue during
in vivo MRI-guided cryoablation. Axial 2D UTE images were acquired at the end
of the first and second freeze cycle during an MRI-guided renal cryoablation
procedure. Measurable MR signal could be obtained from frozen tissue. MR
temperature maps were estimated using a relative signal level calibration
performed in ex-vivo porcine muscle. Our work demonstrates the feasibility of
2D UTE-based MR thermometry of frozen tissue during in vivo MRI-guided
cryoablation, which could be an important step towards clinical application of
this technique.
Introduction
MRI-guided
cryoablation is a promising minimally invasive treatment with applications in
liver, kidney, and prostate cancer1. To assure effective treatment, intraprocedural
temperature feedback is desirable but a non-invasive approach is currently
lacking. Previous studies have demonstrated measurable MR signal from frozen
tissue using ultrashort TE (UTE) MR imaging2,3. Recently, we have
shown the feasibility of 3D MR thermometry of cryoablation using UTE signal
intensity in an ex-vivo setup4. In vivo, a 3D approach presents
issues with out-of-volume magnetization due to non-selective excitation as well
as undersampling to avoid unacceptably long measurement times. Therefore, this
work investigated the feasibility of 2D UTE-based MR thermometry of frozen
tissue during in vivo MRI-guided cryoablation.Methods
Calibration experiments
An
MR-compatible cryoneedle (IceRod, Galil Medical) was inserted into ex-vivo
porcine muscle specimens (n=2) at room temperature on a 3T clinical MR system
(Magnetom Skyra, Siemens). Three fiber-optic probes (T1, Neoptix) were placed at
5mm intervals for temperature reference (Figure 1). Two freeze-thaw cycles of
10:3 min. were applied under continuous MR imaging using a single-slice dual
echo UTE sequence that utilizes half pulse RF excitation followed by a center-out
radial readout to achieve a shortest TE of 50µs. Imaging parameters included TR/TE1/TE2=32.4/0.05/1.93ms,
flip angle=20°, slice thickness=10mm, FOV=360mm, matrix size=192x192, no. of
spokes=384, averages=3 and acq. time=1:14min. Fat saturation was used and
spatial saturation bands were placed symmetrically on both sides of the imaging
slice to suppress out-of-slice magnetization5. Per temperature
probe, signal intensity (SI) values of the first echo image were recorded for
three voxels at same radial distance from the cryoneedle, normalized to the
maximum signal level observed in cooled unfrozen tissue directly adjacent to
the ice-ball and related to recorded temperatures. Data for temperatures <0°C
were fitted by a mono-exponential function.
In vivo measurements
2D UTE imaging was performed during an MRI-guided
renal cryoablation procedure on the same 3T MR system. After percutaneous
placement of three cryoneedles, two 10:3 min. freeze-thaw cycles were applied
under regular MR monitoring using T2-weighted half-fourier single-shot turbo
spin echo (HASTE) imaging. Axial 2D UTE images with same measurement parameters
regarding TR, TE and flip angle as in the calibration (matrix size reduced to 128x128)
were acquired during the first minute of freezing and at the end of the first
and second freeze cycle. For frozen tissue, signal values of the first echo
image were normalized to the maximum signal level in cooled but unfrozen
surrounding tissue. MR temperature maps were estimated based on the ex-vivo
calibration curve.
Results
In the
calibration experiments, normalized SI decreased mono-exponentially with
temperature for T<0°C, with the signal decay fitted by SI=0.73*e0.08T
(R2=0.91) (Figure 2). In vivo, measurable signal was observed within
the ice-ball in the shortest TE images (Figure 3a-b). Signal within the frozen
tissue is highlighted in the difference image between the first and second echo
(Figure 3c). Estimated temperature maps calibrated to the ex-vivo measurements
show temperature differences within the frozen zone at the end of the first and
second freeze cycle (Figure 4).Discussion
This work demonstrates
the feasibility of obtaining measurable MR signal from frozen tissue during in
vivo cryoablation using 2D UTE imaging. Image acquisitions were achieved during
free-breathing, within a clinically realistic window (<1min.) and with
acceptable spatial resolution (2.8x2.8mm). Although a higher resolution would
be preferable to reduce partial volume effects of the temperatures, a larger
voxel size was chosen to optimize signal-to-noise in the frozen tissue. Estimated
temperature maps were calculated based on relative signal level calibration in
porcine muscle. A previous study has found the temperature dependence of UTE MR
signal to be consistent between heart muscle, kidney and liver tissue, which
could potentially obviate the need for tissue-specific calibration3.
Alternatively, R2* may be used for MR thermometry of frozen tissue2,3.
However this would require acquisition of multiple images at different short TEs
to accurately estimate the relaxation rate, which may be impractical in a
clinical time frame. Multi-echo acquisitions are limited by minimum echo
spacing and in our data the second echo did not contain sufficient signal to
perform accurate T2* estimation. Finally, our images were affected by subtle
streaking. Increased FOV may help to reduce these artifacts in future acquisitions.
Imaging improvements and validation of our calibration to kidney and prostate
tissue are currently under investigation.Conclusion
2D
UTE-based MR thermometry of frozen tissue is feasible during in vivo MRI-guided
cryoablation. Further work addressing the accuracy of the MR-based temperature
estimates is required. Clinical application could allow insight into the
effective treatment zone during MRI-guided cryoablation procedures.Acknowledgements
No acknowledgement found.References
1.
Morrison
et al. JMRI 2008
2. Wansapura et al. Acad Radiol 2005
3.
Kaye
et al. JMRI 2010
4.
Overduin et al. JMRI 2016
5. Krafft et al. ISMRM 2015, abstract: 87