Emilee Minalga1, Allison H Payne1, Robb Merrill1, Dennis L Parker1, and J. Rock Hadley1
1UCAIR, University of Utah, Salt Lake City, UT, United States
Synopsis
The use of dedicated RF coils can
increase image SNR in magnetic resonance guided focused ultrasound (MRgFUS). However,
shielding and eddy current effects of the ultrasound transducer ground plane (TGP)
can significantly degrade the increased SNR. Accounting for these
considerations when designing the ultrasound transducer can potentially
increase SNR throughout the imaging volume during MRgFUS procedures.
Purpose
In MRI, the highest
signal-to-noise ratio (SNR) is achieved by placing the radiofrequency (RF)
coils as close to the region of interest as possible. In MRgFUS treatments, the
transducer is often within the volume that would contain the ideal RF coil. RF eddy
currents in the TGP can have significant affects on the RF coil sensitivity and
SNR in the region of interest. Using an approach similar to a prior study to
reduce gradient induced eddy currents in a TGP [1], this work investigates the effects
of different TGP configurations on the perceived sensitivity and relative SNR of
nearby RF coil elements. The results from this work provide strategies for TGP
design that can reduce RF eddy currents and provide improved SNR for MRgFUS.Methods
The effects of TGP RF eddy
currents were evaluated using a single rectangular RF coil (20cmx13cm) wrapped
around a homogeneous cylindrical phantom and a simulated TGP. The phantom was a
homogenous (1.955g/L CuSO4) cylinder (12cm diameter
x 11cm height). The TGP
was simulated by placing copper tape on the face of a 3-D printed plastic
transducer (aperture: 13 x 10 cm, 10 cm radius of curvature). To create
different TGP configurations, the ground plane was broken into multiple rows
and columns of copper patch elements by removing thin copper strips (1 mm wide)
in the short and long axis directions (see Figure 1 and Table 1).
To minimize
gradient eddy currents and maximize RF eddy currents, the phantom, RF coil and
TGP were positioned as shown in Figure 1 for all SNR studies, with the RF loop
axis perpendicular to the B0 field of the scanner. For each ground plane
configuration, the RF coil loop was tuned and matched to 123 MHz with an
insertion loss better than -35 dB. Active and preamp detuning were better than
-35 dB and -20 dB, respectively.
The SNR studies were performed by acquiring 2D
GRE images (TR/TE=500ms/10ms, Flip angle=90°, FOV=256x256mm, 1x1mm resolution,
5mm slice thickness on a Siemens 3T PrismaFit scanner) of the phantom with the RF
loop and dividing the resulting image by the Standard Deviation of the noise in the oversampled region of the image. Results
SNR map results of various ground
plane conditions are shown in Figure 2. The lack of distortions between SNR
plots indicate that the gradient eddy currents were minimized. Relative SNR values for the different TGP
configurations are plotted in Figure 3. Using the single element ground plane
resulted in a 57% reduction in SNR compared to the SNR with no TGP. As the
number of TGP segments were increased, RF eddy currents were reduced resulting
in increased SNR.Discussion and Conclusion
This work demonstrates that use of
a single continuous TGP in the MRgFUS transducer can result in significant SNR
loss. This loss can be dramatically reduced by splitting the ground plane into smaller
segments.
As seen in Figures 2 and 3, increasing
the number of segments in the transducer ground plane increased the relative SNR
both at the center of the phantom and at the transducer face for this
particular transducer and coil configuration.
Since different transducer/sample configurations
can have varied effects on SNR, the ground plane configuration should be
evaluated at the beginning of the ultrasound transducer design process for a
given application. The reduction of RF eddy current effects in the imaging
volume will increase image SNR and enhance both anatomic and temperature
imaging during MRgFUS treatments. Acknowledgements
Funding was provided by NIH grant R01
CA172787References
1. Lechner-Greite SM, Hehn N, Werner B, Zadicario E, Tarasek M, Yeo D. Minimizing eddy currents induced in the ground plane of a large phased-array ultrasound applicator for echo-planar imaging-based MR thermometry. Journal of Therapeutic Ultrasound. 2016;4:4.
2.
Minalga
E, Payne A, Merrill R, et al. An 11-Channel Radio Frequency Phased Array Coil
for Magnetic Resonance Guided High Intensity Focused Ultrasound of the Breast. Magnetic
Resonance in Medicine. 2013;69(1):295-302.