Yoshie Omiya1, Shintaro Ichikawa1, Utaroh Motosugi1, and Hiroshi Onishi1
1University of Yamanashi, Chuo, Japan
Synopsis
We evaluated the feasibility
of magnetic resonance elastography (MRE) for uterine fibroids and the effect of
imaging sequences (i.e., spin-echo echo-planar imaging [SE-EPI] and
gradient-echo [GRE]) on stiffness measurements. All MRE were performed in two
cross-sectional planes (i.e., axial and sagittal). SE-EPI-MRE showed lower
fibroid stiffness values than GRE-MRE. Each MRE method is a valid and reliable
technique for measuring the stiffness of uterine fibroids.
Background and Purpose
Magnetic resonance elastography (MRE) is a new technique for assessing tissue
stiffness. MRE has been established as a noninvasive diagnostic tool for the
assessment of hepatic fibrosis, and applied for other organs. However, there
are only a few reports of MRE for uterine fibroids (1). The purpose of this
study is to evaluate the feasibility of MRE for uterine fibroids and the effect
of imaging sequences (i.e., spin-echo echo-planar imaging [SE-EPI] and
gradient-echo [GRE]) on stiffness measurements.Methods
This retrospective study was approved by the
institutional review board; the requirement for informed patient consent was
waived. Twenty-two
patients (mean age, 45.4
years (38-52 years)) with 23 fibroids were enrolled. The mean size of the
fibroids was 90.1 ± 23.1 mm. Location of the fibroids was intramural (n = 19), submucosal
(n = 3), or subserosal (n = 1). All patients underwent MRE examinations with a 1.5-T MR system via two methods
(SE-EPI and GRE) using a pneumatic vibrator which was placed on the lower
abdominal wall and fixed with an elastic band. All MRE were performed in two cross-sectional
planes (i.e., axial and sagittal). A two-dimensional postprocessing method was
used for GRE-MRE, while a three-dimensional postprocessing method was used for
SE-EPI-MRE. Table 1 summarizes the MR sequence parameters. Two radiologists measured
the stiffness of fibroids. We manually placed as large a region of interest as
possible to exclude crosshatching of areas inadequate for measurement on the stiffness
map. Central slices were selected for SE-EPI-MRE. Interobserver agreement was assessed by intraclass correlation
coefficients (ICC).Results
The mean stiffness values of the fibroids were
as follows: GREaxial, 4.43 ± 1.48 kPa; GREsagittal, 4.32
± 1.78 kPa; SE-EPIaxial, 5.44 ± 2.10 kPa; and SE-EPIsagittal,
5.58 ± 2.33 kPa (Figure 1). The mean stiffness values of the fibroids measured
by SE-EPIaxial were significantly higher than those measured by GREaxial
(P = 0.0018) and GREsagittal
(P = 0.0012). The mean stiffness
values of the fibroids measured by SE-EPIsagittal were significantly
higher than those measured by GREaxial (P = 0.0138) and GREsagittal (P < 0.0001). There were no
significant differences between GREaxial and GREsagittal
(P = 1.0000) or SE-EPIaxial
and SE-EPIsagittal (P
= 1.0000).
The ICC
between the 2 observers was excellent for all methods (ρ = 0.903-0.981, Figure
2). Figure 3 presents representative clinical case.Discussion
We demonstrated that stiffness values measured by SE-EPI-MRE were
significantly higher than those measured by GRE-MRE. In liver MRE,
signal-to-noise ratio of SE-EPI-MRE is higher than that of GRE-MRE (2). This
might have led to the difference between the stiffness of uterine fibroids.Conclusion
SE-EPI-MRE showed lower fibroid stiffness values than GRE-MRE. Each MRE
method is a valid and reliable technique for measuring the stiffness of uterine
fibroids.Acknowledgements
No acknowledgement found.References
(1) Stewart EA, et al. Fertil
Steril 2011;95:281-284.
(2) Mariappan YK, et al. Radiology 2016 Aug 10:160153. [Epub ahead
of print]