Li Yang1, Mengsu Zeng2, Xuhao Song2, Caixia Fu3, and Xu Yan3
1Department of Radiology, Shanghai Institute of Medical Imaging, Zhongshan Hospital, Fudan University, Shanghai, People's Republic of China, 2Department of Radiology, Zhongshan Hospital, Fudan University, 3Siemens Healthcare, Shanghai, P.R. China
Synopsis
Intravoxel incoherent motion
(IVIM) model provides both pure water motion and microcirculation by using multiple
b values. IVIM imaging has been shown to be useful for assessment of liver
diseases, including liver fibrosis and hepatocellular carcinoma. However, the clinical
implementation of IVIM imaging is limited by long acquisition time. We compared
IVIM imaging with different b-values to determine the combination of b-values
in IVIM imaging that allows the relatively short acquisition time to obtain
reproducible values of the IVIM parameters in patients with hepatocellular
carcinoma. Our results showed at least 10b-values should be used in IVIM imaging for the
assessment of hepatocellular carcinoma, and 5b-values IVIM imaging might
increase errors in the perfusion-related f and D* values.
INTRODUCTION
Intravoxel
incoherent motion (IVIM) model provides both pure water motion and microcirculation
by using multiple b values. IVIM imaging has been shown to be useful for
assessment of liver diseases, including liver fibrosis and hepatocellular
carcinoma1,2. However, the clinical implementation of IVIM imaging is limited by
long acquisition time. Therefore, the purpose of this study was to determine
the combination of b-values in IVIM imaging that allows the relatively short
acquisition time to obtain reproducible values of the IVIM parameters in
patients with hepatocellular carcinoma.METHODS:
Free-breath
IVIM imaging with 13 b values (Protocol A) and 5 b values (Protocol B) were
performed at 1.5 T MR scanner (MAGNETOM Aera, Siemens Healthcare, Erlangen,
Germany) in 10 patients (2 females, 8 males; mean age 52) with 11
hepatocellular carcinomas. The IVIM imaging parameters of both protocols were
as follows: echo time=4500ms, repetition time=51ms, acquisition matrix=128×128, slices=25, slice thickness=5mm. For Protocol A, b
values at 0, 5, 10, 20, 40, 60, 80, 100,150, 200, 400, 780, 800 s/mm2
were selected, numbers of averages were 0 (b=0 s/mm2), 2 (b≤200 s/mm2) or 4 (b>200
s/mm2), and acquisition time was 7min4s; For Protocol B, b values at
0, 15, 40, 150, 800 s/mm2 were selected, number
of averages was 6, and acquisition time was 6min1s. IVIM parameter maps,
including the diffusion coefficient D, the pseudodiffusion coefficient D*
and the perfusion fraction f, and apparent diffusion coefficient (ADC) were
generated by using a prototype software body diffusion toolbox (Siemens
Healthcare, Germany), and the protocol A was then split into two 10 b-values
protocols: 0, 20, 40, 60, 80, 100, 150, 200, 400, 800 s/mm2
(Protocol C) and
0, 5, 10, 20, 40, 80, 150, 200, 780, 800 s/mm2
(Protocol D) to acquire the corresponding parametric maps in the same software.
A single representative region of interest was traced manually along the margin
of the tumor at the slice containing the largest tumor, excluding areas of
hemorrhage and necrosis. Intrarobserver agreement of the IVIM model parameters
was evaluated by using intraclass correlation coefficient (ICC), and the
differences between these four protocols were assessed by using two-way
analysis of variance followed by Bonferroni's post hoc test.RESULTS
Representative
images of D, D*, f and ADC are provided in Figure 1. Intrarobserver measurements
revealed good agreement (ICC=0.81~0.93). The distribution of IVIM parameters (D and D*) and ADC for the tumor of different
combination of b values are provided in Figure 2, revealing D values (P=0.351) and ADC (P=0.244) were not significantly different between the different
protocols, but D* values (P<0.001) and f values (P <0.001) were different
between Protocol B and other protocols. DISCUSSION
Although IVIM
becomes more and more popular, there is currently no clear consensus on
the number and distribution of b-values to use3. With more b values
being applied, IVIM imaging may not well tolerated by patients due to long scan
time. We acquired IVIM imaging with different b values, the results showed that
5b-values might increase errors in the perfusion-related f and D* values
because of limited number of low b values.CONCLUSION
At least 10b-values should be used in IVIM
imaging for the assessment of hepatocellular carcinoma, and 5b-values IVIM
imaging might increase errors in the perfusion-related f and D* values.Acknowledgements
No.References
1.Wu C H, Ho M C, Jeng Y M, et al. Assessing hepatic fibrosis: comparing the intravoxel incoherent motion in MRI with acoustic radiation force impulse imaging in US[J]. European radiology, 2015, 25(12): 3552-3559.
2.Woo S, Lee J M, Yoon J H, et al. Intravoxel incoherent motion diffusion-weighted MR imaging of hepatocellular carcinoma: correlation with enhancement degree and histologic grade[J]. Radiology, 2013, 270(3): 758-767.
3.Intravoxel Incoherent
Motion Protocol Evaluation and Data Quality in Normal and Malignant Liver
Tissue and Comparison to the Literature