Hirohiko Imai1, Kiyotaka Miyake2, and Tetsuya Matsuda2
1Center for the Promotion of Interdisciplinary Education and Research, Kyoto University, Kyoto, Japan, 2Department of Systems Science, Graduate School of Informatics, Kyoto University, Kyoto, Japan
Synopsis
We propose a use of sinc-modulated
RF pulse train instead of the
conventional continuous wave RF irradiation under the presence of constant
gradient as a saturation scheme in chemical exchange
saturation transfer (CEST) MRI aiming at a simultaneous acquisition of multiple
Z-spectra. The proposed method was applied for a glutamic acid solution in
water. The multiple Z-spectra could be observed along the gradient encoding
direction by repetitive application of the saturation scheme. Thus, the present
study shows the potential of the proposed methodology for accelerating the CEST
MRI.
INTRODUCTION
Chemical exchange saturation
transfer (CEST) MRI can sense the presence of low-concentration solutes
containing chemically exchangeable protons.1 To quantify the CEST
effect, a series of images or spectra collected with a different saturation
frequency offset is required for generating Z-spectrum.1 To
accelerate Z-spectrum acquisition, ultrafast Z-spectroscopy (UFZ)2-5
or ultrafast CEST imaging (UCI)6,7 has been developed, in which spectral
information is encoded along a spatial dimension using a magnetic field
gradient during an irradiation of constant wave (CW) RF pulse for saturation. Although
this ultrafast approach requires just two scans for obtaining Z-spectrum, only
a single Z-spectrum can be acquired for the gradient encoding direction, i.e.,
no spatial information about the target molecules could not be obtained along
this direction. In the present study, we propose a use of an RF pulse train instead of the conventional CW for saturation in the UFZ methodology aiming at the simultaneous
acquisition of multiple Z-spectra along the gradient encoding direction.METHODS
The pulse sequence used for
this study is illustrated in Fig.1a. A train of rectangular RF pulses in
conjunction with a constant gradient was applied for saturation transfer (ST) preparation
followed by an image acquisition module. The RF pulse train was modulated by a
three-lobe sinc function8 and its power was adjusted to be a total flip angle
of π/2. The frequency of the RF pulses was set at a
resonance of water proton without a gradient field. The saturation encoding
gradient (Gsat=1.41G/cm) was applied along the readout direction followed
by a crusher gradient. The ST block was repeated N times to enhance the CEST effect. Figure 1b shows the excitation
profile for N=1. For the water
protons, the direct water saturation with a 0.2mm width occurs every 5mm
separation with the position at 0mm as a center (Fig.1b,black). Assuming a
presence of solute exchangeable protons having a resonant frequency difference
of Δω=900Hz with respect to water (chemical
shift difference of 3ppm at 7T), the excitation profile for the solute protons
shifts a distance d=Δω/γGsat,, which is 1.5mm in
the present study (Fig.1b,red). Therefore
the solute protons located at these positions are saturated and exchange with
the water protons, generating a multiple Z-spectra. In the experiments, glutamic acid solution in water (50mM) was used
where the exchangeable amide protons at 3ppm were targeted. 3ml of the solution
was sealed in a cylindrical vessel (20mm i.d.). MR scans were performed on a
Bruker 7T MR system (BioSpec 70/20 USR) using a quadrature transmit-receive
volume coil. The dependence of the CEST effect on N ranging from 1 to 600 was investigated, which corresponds to the ST
preparation period (tsat)
of 19.3ms-11.6s. For image acquisition, RARE sequence was used following ST
preparation with parameters: TR=tsat+2.5s,
effective TE=16.6ms, RARE factor=8, a centric ordered phase encoding, slice
thickness=2mm, matrix size=500×256 and FOV=50×25.6mm2. Depending on the tsat, the acquisition time
for each image took 1.3-7.5min.RESULTS AND DISCUSSION
Figure 2a shows acquired images
with N=1 and N=300. For the N=300, beside
the wide dark bands, which corresponds to the direct water saturation, lines
are seen as slightly reduced signal intensity (indicated by red arrow) as a
result of the chemical exchange saturation transfer. This CEST effect cannot be
found in the image for N=1, which is
due to the insufficient accumulation of saturated protons in water pool. An
intensity profile along the encoding gradient direction extracted from
normalized image of N=300 (Fig.2c)
shows that multiple Z-spectra can be obtained with clear CEST effects at
positions approximately 1.5mm apart from positions of water direct saturation. Figure
3a shows N dependence of normalized
images. Although the sinc-modulated RF pulse train yields sharper-edged
excitation profile, the width of direct water saturation broadened as a result
of the repetitive application of the pulse trains. For increasing the number of
Z-spectra within unit length, i.e., achieving high spatial resolution Z-spectroscopy,
the suppression of this broadening would be required. By using the Z-spectra
extracted from these images, a magnetization transfer ratio asymmetry (MTRasym)
analysis was performed.1 Figure 3b shows
the N dependence of the MTR asymmetry
curves. The values at 3ppm were plotted as a function of N as well as the corresponding tsat
(Fig.3c). The value was reached nearly plateau around N=300(tsat=6s).CONCLUSION
We have demonstrated that the
multiple Z-spectra can be obtained simultaneously by using a sinc-modulated RF
pulse trains in UFZ methodology. While the proposed method requires further
optimization, this study shows its potential for accelerating CEST MRI.Acknowledgements
This work is partly supported by JSPS KAKENHI Grant Number JP15K01283.References
1. van Zijl PCM, et al.
Chemical exchange saturation transfer (CEST): What is in a name and what isn’t.
Magn. Reson. Med. 2011;65:927-948. 2. Xu X, et al. Ultrafast scanning of
exchangeable sites by NMR spectroscopy. Angew. Chem. Int. Ed. 2013;52:8281-8284.
3. Döpfert J, et al. Slice-selective gradient-encoded
CEST spectroscopy for monitoring dynamic parameters and high-throughput sample characterization.
J. Magn. Reson. 2013;237:34-39. 4. Liu Z, et al. UCEPR: Ultrafast localized
CEST-spectroscopy with PRESS in phantom and in vivo. Magn. Reson. Med. 2016;75:1875-1885.
5. Wilson NE, et al. Localized, gradient-reversed ultrafast Z-spectroscopy in
vivo at 7T. Magn. Reson. Med. 2016;76:1039-1046. 6. Döpfert J, et al. Ultrafast CEST imaging. J. Magn.
Reson. 2014;243:47-53. 7. Xu X, et al. Screening CEST contrast agents using ultrafast
CEST imaging. J. Magn. Reson. 2016;265:224-229. 8. Wu EX, et al. MRI cardiac
tagging using a sinc-modulated RF pulse train. Magn. Reson. Med. 2002;48:389-393.