Mitchell Horn1, Ning Hua1, Stephan Anderson1, and Hernan Jara1
1Boston University, Boston, MA, United States
Synopsis
Purpose: To
measure inaccuracies on T1 qMRI caused by magnetization transfer effects that
are inherent to MRI pulse sequences. Methods:
The mixed-TSE pulse sequence was used at several RF power levels to image a
series of agarose gels, sucrose solutions, and Gd solutions. Results: Measured T1s are increasingly
underestimated as a function of increasing semisolid pool size and RF power
level. Conclusion: Uncorrected
magnetization transfer effects can cause large underestimation of T1
measurements in tissue imaged with RF intensive pulse sequences. This work
could have implications for the design of more accurate qT1 mapping algorithms.
Introduction
Magnetization transfer (MT) effects are known to reduce
the apparent T1s of tissue –particularly tissues with significant semisolid
pools such as white matter (WM)—when using RF-intensive MRI pulse sequences,
such as with multislice T1-weighted turbo spin echo (TSE) sequences. The causal
assumption being that the T1 recovery of WM is slowed down by MT effects caused
by all RF pulses; in particular, by the off-resonance pulses interrogating the
other slices of the package. To our knowledge, this theoretical prediction of
the binary spin bath model (1) has not been demonstrated
directly with T1 qMRI methods.
The mixed-TSE pulse sequence (diagram
in Fig. 1) can be used to map the proton
density (PD) as well as the T1 and T2 relaxation times with a single
acquisition, and can be implemented at various effective RF power levels (see
legend Fig. 2) –specifically, as a
single-slice, multi-slice acquisition, and different power levels of the
initial inversion pulse--. The purpose of this work was to measure the T1
relaxation times of three fundamentally different aqueous substances
–specifically, agarose gels, sucrose solutions, and Gd-DTPA solutions-- with
mixed-TSE pulse sequences implemented at increasing RF power levels and to
evaluate the resulting MT effects relative to the qT1 “gold standard”
measurements performed with single slice serial inversion recovery (IR) with
varying inversion times (TI).Materials and Methods
A
multi-compartment phantom consisting of twenty-five plastic vials containing solutions
of Gd-DTPA in distilled water, sucrose in distilled water, and agarose gels
prepared with distilled water, was constructed (Fig.
3). The vials were
positioned in rows sequentially of 4% agarose gel to reduce vibration effects
during scanning. All MRI scanning was performed using a 1.5T clinical scanner
(Achieva, Philips Healthcare, Best, The Netherlands) using the quadrature body
coil for RF transmission and a pair of surface coils (Flex-L) in phased array
configuration for signal reception. The mixed-TSE pulse sequence generates four
images per slice with different T1- and T2-weightings: the contrast control
variables being two inversion times (TI1&2) and two effective echo times
(TEeff1&2). Each TR-cycle of the mixed-TSE begins with and adiabatic
inversion pulse, which can be implemented at two different power levels
(pw1&2). Mixed-TSE scans were performed at five RF power levels (see Fig. 1 legend) with common parameters:
TE1&2eff = 8/100ms, ETL = 16, TI1&2 = 700/7,438ms, TR = 14,875ms,
matrix 176 x 176, voxel = 0.9 x 0.9 x 3mm3). The qT1 reference scans
were single slice IR-TSE run serially at 18 TI values (ms): 25, 50, 75, 100,
200, 300, 400, 500, 600, 700, 800, 900, 1,000, 1,250, 1,500, 1,750, 2,000, and
3,000. The qT1 algorithms for mix-TSE and IR-TSE pulse sequences were programed
in Mathcad (version 2001i, PTC, Needham, MA) and all measurements were
performed using ROIs placed in qT1 maps
(Fig. 3).
Results: Main qT1 results are shown in Fig. 4 for agarose gels and in Fig. 5 for sucrose solutions. Observed MT
T1 reduction effects were largest for highly concentrated agarose gels,
markedly weaker for sucrose solutions, and negligible for Gd-DTPA solutions
(not shown). MT T1 reductions effects were most pronounced for high RF power
level acquisitions (i.e. 30 slices and
high power inversion pulse) and correlated with RF power level (Figs.
4 and 5).Results
Main qT1 results are shown in Fig.
4 for agarose gels and in Fig.
5 for sucrose solutions. Observed MT T1 reduction effects were largest for
highly concentrated agarose gels, markedly weaker for sucrose solutions, and
negligible for Gd-DTPA solutions (not shown). MT T1 reductions effects were
most pronounced for high RF power level acquisitions (i.e. 30 slices and high power inversion pulse) and correlated with
RF power level (Figs. 4 and 5).Discussion
Accurate qT1 measurements are needed for
standardization across platforms and for an increasing number of qMRI applications
including Synthetic-MRI, structural studies of WM, and WM fibrography. In this
work, we demonstrated direct T1 shortening effects caused by MT effects that
correlate with increasing pulse sequence RF power level and are most pronounced
for tissues with largest semisolid pools.Conclusion
Uncorrected magnetization transfer effects can cause
large underestimation of T1 measurements in tissue imaged with RF intensive
pulse sequences. This work could have implications for the design of more
accurate qT1 mapping algorithms and for improving several applications that use
qT1 measures such as Synthetic MRI, correlation time diffusion MRI, and MR Fibrography.Acknowledgements
No acknowledgement found.References
1. Henkelman
RM, Huang X, Xiang QS, Stanisz GJ, Swanson SD, Bronskill MJ. Quantitative
interpretation of magnetization transfer. Magn Reson Med 1993;29(6):759-766.