Morgan Mercredi1, Sheryl Herrera1, Richard Buist2, and Melanie Martin1,3
1Physics and Astronomy, University of Manitoba, Winnipeg, MB, Canada, 2Radiology, University of Manitoba, 3Physics, University of Winnipeg, Winnipeg, MB, Canada
Synopsis
There
is an increasing drive to use diffusion spectroscopy to infer the sizes of
structures in samples. Here, we use apodised cosine gradient spin echo
sequences (OGSE) to infer the surface to volume ratio of a collection of packed
capillary tubes. Aiming to reduce imaging times, this study examines how
the number of gradients affects the accuracy and precision of the fitted
parameters. We found that collecting OGSE data with two b-values may be
sufficient to infer information about small structures, especially if higher
gradients strengths are used.
Introduction
There
is an increasing drive to use diffusion spectroscopy to infer the sizes of
structures in samples1-4. Most methods use pulsed gradient spin echo
sequences2,3 which cannot provide short enough diffusion times to
probe very small structures. Using oscillating gradient spin echo sequences
(OGSE)5-7 a diffusion spectrum can be obtained which probes the
“short-time” or high frequency regime allowing for small structures to be
measured. Previous OGSE measurements used long imaging times with multiple
gradient strengths in fit models to infer sample dimensions8. In an
effort to shorten imaging times, this study examines the effects the number of
gradients used to infer the surface to volume ratio of a collection of tubes.
We found that imaging time can be reduced by a factor of 2.5 in exchange for a
decrease in precision of 18 percent.Methods
MRI Each sample
was imaged using a 7 T Bruker Avance III NMR system with Paravision 5.0 with a
BGA6 gradient set with a maximum gradient strength of 1.01 T/m and a 3.5 cm
diameter bird cage RF coil (Bruker Biospin). The number of sinusoidal waves in
each 40 ms apodised cosine6 gradient pulse ranged from n = 1 to 20,
in steps of 1. Five gradient strengths were used for each frequency and the
gradient pulses were separated by 44.52 ms. The b-values were 0,
60, 120, 180, 240 s/mm2. Tubes: Capillary tubing (Polymicro
Technologies), were used with inner and outer diameters of 5 µm and 151 µm,
respectively. Tubing was filled with filtered water, injected using a modified
syringe9. Tubing was cut into
2 cm pieces (~ 500 in total), packed within a 1.5 mL plastic microcentrifuge
tube (Cole Parmer). The volume of water within the tubes was not large enough
to produce a measurable signal. Thus the tubes were soaked in filtered water,
before being placed into a bird cage RF coil. Two 750 µm inner diameter tubes
filled with filtered water were placed near the microcentrifuge tube and used
as a control. A 10 mm thick (2.0 cm)2 slice perpendicular to the
tubes was chosen. A 64 x 64 matrix was used for 312.5 μm in-plane resolution. Analysis
ROIs were created in the water tubes (ROI #2), within the microcentrifuge tube
(ROI #3-6), and in the noise (ROI #1) as shown in Figure 1. The mean ± standard deviation of the signal in the ROIs was
calculated. The log of the signal versus b-value was fitted to a straight line
and the negative of the slope was used as the ADC for each measurement. The
mean ± standard deviation of
the ADC for each frequency was calculated and used in the fit. Calculation of ADCs and subsequent model fitting was repeated using all
possible combinations of the five b-values. Tube fit The diffusion
spectrum D($$$\omega$$$) for the apodised
cosine sequence data were fitted to the short-time surface to volume model
described in Ref. [10] and the surface to volume ratio (S/V) was extracted from
the fit. Assuming a hexagonal array of cylinders, the surface to volume ratio
is related to the cylinder diameter d through the relation $$$d = 4f/[(1-f)(S/V)]$$$, where f is the packing fraction of the tubes. This
relation was used to relate the surface to volume ratio to the tube diameter.Results
Figure
2 shows variation in fitted S/V ± fitted error
when using only two b-values. Note that uncertainty decreases when larger
gradients are used, with fitted S/V values ranging from .09 ± .01 µm-1
with the highest b-value to .15 ± .04 µm-1 with the smallest b-value.
Figure 3 shows fitted S/V ± fitted error
for the best combinations for each number of gradients (these combinations are
shown in Table 1). The fitted values in Figure 3 are all within 13% of each
other, while error when using two gradients is only 18% larger than when using
all five gradients. Shown in Figure 4 are the corresponding tube diameters
found from each combination of gradients from Figure 3 (assuming 80 percent
packing). The estimated diameters are all within 13% of each other, while their estimated errors are within 30% of each other.
Discussion and Conclusion
This
work provides experimental evidence that collecting OGSE data with two b-values
may be sufficient to infer information about small structures, such as axon diameters, especially if
higher gradients strengths are used. This would allow for shorter imaging
times. Future studies will determine the optimal frequency ranges and gradient
strengths for the expected restriction sizes.Acknowledgements
The
authors would like to thank funding from NSERC, CFI and MRIF, and D. Goultz and
D. Craig for assistance filling the tubes.
References
1.
Parsons EC, Does MD, Gore JC. Temporal diffusion spectroscopy: Theory
and implementation in restricted systems using oscillating gradients.
Magn Reson Med 2006;55:75-84
2. Assaf Y, Blumenfeld-Katzir T, Yovel Y,
Basser PJ. AxCaliber: a method for measuring axon diameter distribution
from diffusion MRI. Magn Reson Med 2008;59:1347-1354
3.
Alexander DC, Hubbard PL, Hall MG, Moore EA, Ptito M, Parker GJ, Dyrby
TB. Orientationally invariant indices of axon diameter and density from
diffusion MRI. Neuroimage 2010;52:1374-89
4. Martin M.
Measuring Restriction Sizes Using Diffusion Weighted Magnetic Resonance
Imaging: A Review. Magn Reson Insights 2013;6:59-64
5.
Mercredi M, Vincent TJ, Bidinosti CP, Martin M. Assessing the accuracy
of using oscillating gradient spin echo sequences with AxCaliber to
infer micron-sized axon diameters.
Magn Reson Mater Phy 2016; doi:10.1007/s10334-016-0575-y
6.
Schachter M, Does MD, Anderson AW, Gore
JC. Measurements of restricted diffusion using an oscillating gradient
spin-echo sequence. J Magn Reson 2000;147:232–237
7. Does M, Parsons
E, Gore J. Oscillating gradient measurements of water diffusion in
normal and globally ischemic rat brain. Magn Reson Med
2003;49(2):206–215
8. Herrera SL, Mercredi ME, Vincent TJ, Buist R, Martin M, Using oscillating gradient spin-echo sequences to infer micron-sized bead and pore radii. ISMRM 2015, 3027
9.
Craig DB. A simple system for the measurement of
the distribution of activities of individual molecules of E. coli
beta-galactosidase. Anal Methods 2012;4(1):85-88
10. Reynaud O, Winters
KV, Hoang DM, Wadghiri YZ, Novikov DS, Kim SG. Surface-to-volume ratio
mapping of tumor microstructure using oscillating gradient diffusion
weighted imaging. Magn Reson Med 2016;76:237-247