Franciszek Hennel1, Markus Weiger1, Manuela Rösler1, Roger Luechinger1, Bertram Wilm1, and Klaas P Pruessmann1
1Institute for Biomedical Engineering, ETH, Zurich, Switzerland
Synopsis
MRI with sub-millisecond TE typically uses
non-Cartesian sampling strategies such as radial UTE leading to increased sensitivity to eddy currents and a high
computational burden. We report that TE values in the range so far reserved for
radial UTE can be reached with Cartesian sampling using a gradient coil
dedicated for muscoskeletal applications in humans.
Introduction
Sub-millisecond echo times (TE) are necessary in many
medical MRI applications targeting bones, cartilage, tendons and ligaments, as
well as short-T2 components in white matter such as the myelin water. The
minimum TE in conventional MRI methods, which sample the k-space on a Cartesian
grid, is related to the minimum duration of the phase encoding gradient pulse. With typical gradient systems and the
resolution of 0.5 mm, which is standard in muscoskeletal MRI, TE of Cartesian
sequences is situated well above one millisecond and could be squeezed down to
about 0.8 ms only for the central k-space part by careful sequence optimization
(1). Due to this limitation, MRI with sub-millisecond TE almost exclusively
uses non-Cartesian sampling strategies such as radial UTE (2). This has the drawback of a less efficient
sampling, increased sensitivity to gradient delays and eddy currents and presents
a higher reconstruction burden, especially when combined with sensitivity
encoding. Here we report that TE values in the range so far reserved for radial
UTE can be reached with Cartesian sampling using a gradient coil dedicated for muscoskeletal
applications in humans.Methods
A custom-built gradient coil of 33 cm inner diameter has
been inserted into the whole body gradient coil of a 3T scanner (Achieva,
Philps, The Netherlands). The coil is connected to Copley 787 amplifiers which
are normally used for the standard whole body gradient configuration of this
system. A maximum gradient of 100 mT/m can be reached with a slew rate of
1200T/m/sec. A home-built transmit-receive quadrature birdcage coil of 28 cm
inner diameter, dedicated for head imaging, was used for this study. The
sequence was the manufacturer’s standard implementation of 3D gradient echo imaging
with non-selective RF pulses and partial echo sampling (10% of the k-space on
the “left hand” side). Using a 30-microsecond
block RF pulse of 7° flip angle and TR of 3 ms, the echo time of 0.5 ms could
be used with 0.5 mm isotropic resolution. To maintain a sufficient
signal-to-noise ratio despite a suboptimal filling factor for knee scanning
with the present coil, the in-vivo experiment was carried out with a limited
resolution of 1x1x2 mm and TE=0.45 ms. The actual response of the gradients has
been measured by a field camera (Skope, Switzerland).Results and discussion
Images of a knee of a healthy male adult are presented in
Fig. 1. The image acquired with TE=0.45ms contains strong signal from all types
of tissues except for the cortical bone which remains close to the noise level.
In particular, the quadriceps tendon and the patellar ligament visible in the
central sagittal plane (Fig 1A) have intensity equal to that of the muscle,
which is in contrast to the gradient echo at 4.6ms (Fig 1.B) where these
structures are dark. The images are not affected by aliasing of out-of-FOV
parts of the leg due to frequency encoding of the longitudinal direction. With
radial scanning anti-aliasing would require an increased number of projections
and a longer scan time. It remains to be investigated if the slight enhancement
of contours orthogonal to the z-axis (direction of the readout gradient) is
caused by the slight overshoot of the z gradient, which is visible in the field
camera measurement (Fig.2), and whether this can be corrected by 1-dimensional
k-space gridding. However, a simple delay of the gradient response is certainly
less harmful than in radial UTE, as it only leads to a linear phase ramp across
the image, an effect automatically removed by the reconstruction.
Conclusion
Images with echo time below 0.5 milliseconds having contrast
properties similar to radial UTE can be acquired with a standard Cartesian partial
gradient echo sequence using a dedicated gradient insert providing 100mT/m with
a slew rate of 1200T/m/s. The benefit of this approach compared to radial UTE
is a higher sampling efficiency, reduced sensitivity to gradient response
imperfections, shorter minimum scan time, avoidance of longitudinal aliasing
and the simplified use of sensitivity encoding.
Acknowledgements
No acknowledgement found.References
1.
Deligianni, X., et al. "High-resolution
Fourier-encoded sub-millisecond echo time musculoskeletal imaging at 3 Tesla
and 7 Tesla." Magnetic resonance in medicine 70.5 (2013): 1434-1439.
2.
Robson,
Matthew D., et al. "Magnetic resonance: an introduction to
ultrashort TE (UTE) imaging." Journal of computer assisted tomography 27.6
(2003): 825-846.