Ning Hua1, Mitchell Horn2, Stephan Anderson3, and Hernan Jara1
1Boston University, Boston, MA, United States, 2Boston University, MA, United States, 3Boston University
Synopsis
Purpose: To
develop a simple and cross-platform pulse sequence that meets the following
criteria: 1) all directly acquired images are clinically useful, 2) achieves
minimalistic scan times, and 3) leads to excellent image quality as well as
accurate MS-qMRI mapping. Methods: Tri-FSE
consists of a single-echo-FSE sequence that is run in temporal
concatenation with a dual-echo FSE
sequence that all together generate T1-, T2, and PD-weighted directly acquired
images. Results: Tri-FSE was
implemented at medium and high spatial resolution at 1.5T and 3.0T. Conclusion: Tri-FSE is a dual purpose
simple pulse sequence that is useful for clinical and scientific purposes.
Introduction
The directly acquired images generated with
multi-spectral (MS) qMRI --proton density (PD) and relaxation times (T1, and T2)
-- pulse sequences generally do not conform to the radiologist’s contrast
expectations and consequently MS-qMRI is largely relegated to research
endeavors. Because scientific research can reduce patient throughput there is
great need for developing dual-purpose MS-qMRI pulse sequences that can simultaneously
be used as clinical as well as scientific tools. The purpose of this work was
to develop and implement a minimalistic and cross-platform pulse sequence --termed
herein Tri-FSE-- that meets the following criteria: 1) any and all directly
acquired images are clinically useful, 2) achieves minimalistic scan times, and
3) leads to excellent directly acquired image quality as well as accurate
MS-qMRI mapping.Methods and Materials
As shown in Figure 1, Tri-FSE
consists of a single echo-FSE sequence that is run in temporal concatenation
--and with the same pre-scan settings-- with a dual-echo FSE sequence that all
together generate T1-, T2, and PD-weighted directly acquired images. Tri-FSE
was implemented at standard clinical spatial resolution (total combined scan
time ≤ 7min. voxel dimensions = 0.5 x 0.5 x 5mm3) at 1.5T (Achieva,
Philips Healthcare, Best, The Netherlands) and 3.0T
(Discovery MR750w, GE Healthcare, Waukesha, WI). Key parameters: 22/100ms
effective echo times, 600/6,000ms repetition times (values depending on
anatomical coverage). Tri-FSE was also implemented at high spatial resolution
(3T: total scantime 10min, 90 slices with reconstructed voxel 0.47 x 0.47 x
1.8mm3). In all cases, the echo train length of the DE-FSE
(ETL=10-14) was twice that of the single echo T1-weighted sequence (ETL=5-7), parallel
imaging was not used, and scanning was performed with the quadrature body coil
and a multi-channel phased array head coil for RF transmission and signal
detection respectively. The matching PD, T1, and T2 qMRI mapping algorithms
were programmed in Mathcad (version 2001i, PTC, Needham MA) and adapted to
handle large datasets. Tri-FSE was implemented clinically at 1.5T and 3.0T and
in addition, one healthy volunteer (male, 24yo) was consented and scanned at 3T
with the high-spatial resolution variant of the Tri-FSE in accordance and
approval by our Institutional Review Board (IRB).Results
The Tri-FSE pulse sequence was implemented in the
scanners of our institution and is run routinely for clinical purposes. An
example of the high spatial resolution Tri-FSE directly acquired images is
illustrated in Figure 2 (top row). The directly acquired images conform to
clinical standards in terms of PD-, T1-, and T2- weighting and meet or exceed
current spatial resolution standards. The corresponding qPD, qT1, and qT2 maps
are shown in Figure 2 (bottom row): ROI measurements for white matter (WM), cortical
gray matter (GM), intraventricular cerebrospinal fluid (CSF) are approximately WM
(0.5ru/600ms/90ms), GM (0.8ru/1,400ms/110ms), and CSF (0.96ru/3,100ms/2,300ms).The
values for GM and CSF are in the range of reported 3T values (1) at
age 24 years. Notably, the T1 values obtained for WM appear underestimated (see
Discussion).Discussion
With the advent of faster MRI pulse sequences and
image processing techniques, MS-qMRI is now possible in routine clinical
practice. Availing MS-qMRI datasets of the main parameters enable several post-processing
applications such as Synthetic MRI, Structural qMRI, and most importantly, it
can enhance diagnostic accuracy by allowing direct intra-patient and
inter-patient comparisons. Tri-FSE described herein is scan time efficient and
can be used with scanners any manufacturer that offer the DE/SE-FSE pulse
sequences; it generates directly acquired images that are immediately available
for clinical assessment and further offer the potential for MS-qMRI processing.
The T1 underestimation for WM likely results from the strong magnetization
transfer effects that are inherent to the RF-intensive Tri-FSE pulse sequence (2,
3).Conclusion
Tri-FSE has
been implemented and tested at 1.5T and 3.0T: it is a scan time efficient and
widely available pulse sequence for MS-qMRI of PD, T1, and T2. This simple and
minimalistic approach to MS-qMRI could lead to broadening the clinical use of
qMRI measures for improved assessment of disease as well as enrich scientific
research. Further theoretical work is necessary for correcting the inherent MT
effects.Acknowledgements
No acknowledgement found.References
1. Stanisz GJ, Odrobina EE, Pun J, et al. T1,
T2 relaxation and magnetization transfer in tissue at 3T. Magn Reson Med 2005;54(3):507-512.
2. Melhem ER, Jara H, Yucel
EK. Multislice
T1-weighted hybrid RARE in CNS imaging: assessment of magnetization transfer
effects and artifacts. J Magn Reson Imaging 1996;6(6):903-908.
3. Meara SJP, Barker GJ. Impact
of incidental magnetization transfer effects on inversion recovery sequences
that use a fast spin echo readout. Magnetic Resonance in Medicine
2007;58(4):825-829.