In this
Figure 1c-d shows the simulated signal phase of isochromats (ensembles of spins with a common Larmor frequency) with different off-resonance throughout the acquisition train. For the MRF pattern with varying TR, a dependence of the signal phase on the length of the TR becomes apparent, even though a constant TE was used. The signal phase of the free induction decay (fid) signal path remains constant, but the phase of the echo pathways change. Temporal changes of the phase difference between isochromats will cause signal magnitude modulations as displayed in figure 2a & 3a. Figure 1d shows that switching to an MRF-pattern with constant TR generates a temporally constant phase that provides robustness against intra-voxel dephasing (see Fig. 2b & 3b). The dephasing results in a time-constant reduction of the signal magnitude and therefore does not affect parameter determination except for an increased noise level. Figure 3 further emphasizes this effect in a phantom measurement where intra-voxel dephasing was simulated by averaging over increasing areas along the direction of linearly growing off-resonances.
Figure 4 illustrates the correlation between fingerprints, which can be seen as an indication of the encoding capability [5]. Since the correlation between any two fingerprints, for an acquisition with constant TR does not increase, a loss in the encoding capability is not expected.
[1] Ma D, Gulani V, Seiberlich N, Liu K, Sunshine JL, Duerk JL, Griswold MA. Magnetic Resonance Fingerprinting. Nature 495, 187–192.
[2] Jiang Y, Ma D, Seiberlich N, Gulani V, Griswold MA. MR fingerprinting using fast imaging with steady state precession (FISP) with spiral readout. Magn. Reson. Med., 74: 1621–1631.
[3] Hamilton J, Deshmane A, Griswold MA, and Seiberlich N. MR Fingerprinting with Chemical Exchange (MRF-X) for In Vivo Multi-Compartment Relaxation and Exchange Rate Mapping. In Proceedings of the 24rd Annual Meeting of the ISMRM, Singapore, Singapore, 2016. Abstract 0431
[4] Bernstein MA, King KF, Zhou XJ. Handbook of MRI Pulse Sequences. Academic Pr. Inc
[5] Cohen O, Sarracanie M, Armstrong BD, Ackerman JL, Rosen MS. Magnetic Resonance Fingerprinting Trajectory Optimization. In Proceedings of the 22rd Annual Meeting of the ISMRM, Milan, Italy, 2013. Abstract 0027
[6] Sommer K, Amthor T, Koken P, Doneva M, Börnert P. Towards Judging the Encoding Capability of Magnetic Resonance Fingerprinting Sequences. In Proceedings of the 24rd Annual Meeting of the ISMRM, Singapore, Singapore, 2016. Abstract 0429
Figure 1:
Figure a-b) show the flip angle and TR pattern used, respectively. Both are identical to the pattern published in [2]. Figures c-d) show the signal phase evolution during the acquisition train for a variable TR (c) and a constant TR (d). For all off-resonances the phase of the first acquisition was normalized to -pi/2.
Figure 2:
Figure a) displays the temporal signal magnitude evolution during standard FISP MRF sequence with variable TR for different T2* values (T1 = 1800ms, T2 = 120ms). Figure b) shows the same plot for a FISP MRF sequence with constant TR. The complex signal vectors in both plots were normalized to an identical norm. The off-resonance distribution for the corresponding T2* values can be seen in c). The simulated range of off-resonances was ±100Hz, but only ±60Hz is displayed in c).
Figure 3:
Figure a) displays the normalized signal magnitude of a FISP MRF sequence with variable TR measured in an agarose phantom with a linear off resonance distribution, as seen in c). b) Shows the signal magnitude for a FISP MRF sequence with constant TR. The averaged areas are indicated by the black boxes in c).
Figure 4:
Correlation matrices with respect to T1 and T2 for FISP MRF with variable and fixed TR. These were calculated as presented in [5]. Although subtle, a general decrease in the correlation between any two fingerprints can be observed for constant TR.