Kihun Bang1, Jinseong Jang1, and Dosik Hwang1
1School of Electrical and Electronic Engineering, Yonsei University, Seoul, Korea, Republic of
Synopsis
In Magnetic Resonance Imaging system, acquiring fewer measurements is required to reduce scan time, but it leads the aliasing artifact. Compressed Sensing is exploited to reconstruct image from undersampled data without artifacts by solving the optimization problem. However, It has some difficulites in selecting regularization parameters and this abstract propose the way to select regularization parameters by evaluating image quality. The quality of reconstructed image from proposed method is much better than the image from manual parameters. This study also has potential to be helpful in fast MR imaging.
Purpose
It
is very important to reduce scan time of MRI. we hope to get fewer measurements
for fast acquisition. However, insufficient sampling acquisition can degrade image
quality and it leads to aliasing artifact because we cannot meet the Nyquist
sapling ratio. Compressed Sensing (CS) 1 is to reconstruct full-sampling
like images from undersampled and aliased image. CS uses the sparsity of MRI in
sparsifying transform such as wavelet domain and reconstructs the undersampled
MR images into full-sampled-like image which do not have aliasing artifact. To
solve the minimization problem of CS, choosing the proper regularization
parameter is necessary. However, conventional setting of regularization
parameter is manually selected in various experiments or to use the L-curve
based setting which has some limitations (e.g. the smooth solution and the
problem size) 2. In order to overcome the limitation of L-curve
method, this abstract proposes choosing the proper regularization parameter
using no-reference image quality assessment (NR-IQA) not to have ground-truth
images.Methods
The
way to reduce the artifact from the undersampled MR data is solving the
optimization problem. The target image $$$x$$$ is acquired by minimizing the difference
between the Fourier transform of target image $$$x$$$ and the undersampled real MR scan data y,
which is used in 'data consistency'. The problem is made up by adding two terms
of reducing artifacts, one is sparsity term in wavelet domain and another is
total variance term. We need to set the weighting parameters
,
(which are called regularization parameters)
on two terms of reducing artifacts. The solution of the optimization problem
(1) is the target image
which has data consistency with $$$y$$$ and its
artifacts were minimized. The reconstructed image is varying with the
regularization parameters. Setting the regularization parameters exploits the image
quality assessment (IQA) of target image $$$x$$$. $$\min_{x}\parallel{F_{u}x-y}\parallel_{2}+\lambda_{w}\parallel{\Psi{x}}\parallel_{1}+\lambda_{t}TV_{x}\space\space\space\space(1)$$ Aliasing artifacts
have directional structure. NIQE has possibility to detect aliasing artifact.3
but multi-directional filter so as to evaluate the structural information
of image 4. Therefore, Quality evaluation using Multi Directional filters
for MR images (QEMDIM) is good at evaluating image which has undersampled
artifacts and highly correlated with Full Reference (FR)-IQA (e.g. RMSE, SSIM).
(Figure 1). Reconstructed Images which the regularization parameters were set
manually are evaluated using QEMDIM. The best quality image which has the
lowest Q and its regularization parameters are selected. $$Q=\sqrt{(P_{r}-P_{e})^{T}(P_{r}-P_{e})}\space\space\space\space(2)$$ where $$$P_{r}$$$ and
$$$P_{e}$$$ are the quality-aware
features of the undistorted databases and distorted image. Since QEMDIM is
no-reference IQA metric, reconstructed image can be directly evaluated. Therefore,
the best quality images and the parameters can be obtained by repeating IQA
process in various regularization parameters. Figure 1 shows that QEMDIM of CS
results in various regularization parameters has high correlation with FR-IQA,
RMSE. From this result, QEMDIM is great NR-IQA for in vivo CS reconstructed images. Figure
2 shows the process of parameter selecting using NR-IQA. Figure 3 shows that reconstructed images of reduction factor 2 images of 256X256 T2-FLIAR (Cartesian undersampling). Image quality results are various depending on regularization parameters. The trend of QEMDIM is similar to RMSE. From proposed method, we can know which the parameters are the most suitable to CS reconstruction.Conclusion
Selecting
the regularization parameters is very important in MRI reconstruction from
undersampled k-space data. We proposed choosing optimized regularization
parameters using IQA. The reconstructed image quality from proposed method is
better than the conventional solution using the manual parameter. Furthermore,
proposed method has great potential in MRI reconstruction. It is feasible to
solve the limitation of choosing regularization parameters using L-curve based
method. It also has potential to solve the optimization solution faster and
more accurate by using IQA in iterative steps. It is meaningful to study on
tuning of regularization parameter in CS using No Reference IQA.Acknowledgements
This
research was supported by the National Research Foundation of Korea (NRF) grant
funded by the Korean government (MSIP) (2016R1A2R4015016).References
[1] Lustig, Michael, David Donoho, and John M. Pauly.
"Sparse MRI: The application of compressed sensing for rapid MR
imaging." Magnetic resonance in medicine 58.6 (2007): 1182-1195.
[2] Hansen,
Per Christian. The L-curve and its use in the numerical treatment of inverse
problems. IMM, Department of Mathematical Modelling, Technical Universityof
Denmark, 1999.
[3] Mittal, Anish,
Rajiv Soundararajan, and Alan C. Bovik. "Making a “completely blind” image
quality analyzer." IEEE
Signal Processing Letters 20.3
(2013): 209-212.
[4] Jang, Jinseong, Taejoon Eo, and Dosik Hwang,
"Quality Evaluation Scheme for No-Reference MR Images Using Pre-Scanned MR
Big Data." Proceedings of the 25 Annual Meeting of ISMRM. Vol. 0701. 2016.