Johannes Strasser1, Lukas Pirpamer1, Franz Fazekas1, and Stefan Ropele1
1Department of Neurology, Medical University of Graz, Graz, Austria
Synopsis
In
MRE, motion encoded phase images are acquired to calculate mechanical tissue
parameters based on shear wave propagation. We here propose a fast multi
readout MRE imaging concept based on the displacement encoding via stimulated
echo acquisition (DENSE). In this proof of concept study, phantom experiments
yielded excellent clear wave images. The results indicate that the proposed
technique could be used to acquire images using short echo times and accelerate
the total acquisition time of MRE examinations.
Introduction
Magnetic
resonance elastography (MRE) enables the assessment of tissue mechanical
properties via imaging the wave propagation of induced mechanical vibrations [1-5].
In conventional MRE, motion induced phase shifts coming from motion during the
application of bipolar encoding gradients are analyzed. Such gradients
typically oscillate with the same frequency as the mechanical vibration. Consequently,
assessment of low frequency vibrations may result in long echo times and related
susceptibility artifacts. Recently, low frequency MRE has received increased
interest [1,2]. One possibility to keep the echo time short is fractional
motion encoding [1,2,5]. However, tissue motion can alternatively also be
recorded by looking at the temporal displacement of tissue. The displacement information
can be gathered by using a displacement encoding via a stimulated echo (DENSE)
sequence [4,6], which has been proposed for in vivo human heart MRE experiments
[4]. In the DENSE-MRE approach, the displacement sensitivity is
encoded via modulation and demodulation gradients. Since their duration is not
coupled to the vibration wave period, much shorter echo times can be achieved. To
get the wave propagation, a set of time shifted phase offset images has to be
acquired. This prolongs the total acquisition time (TA) by the factor of phase
offset image amount, since the sequence has to be repeated with different
offsets in relation to the driving frequency. We here propose an approach to
shorten TA via using multiple readouts following a single DENSE preparation. The
readouts are consecutive and ordered in a way that the vibration wave can be
sampled at different equidistant points in time. This allows sampling of the wave in a single
run and therefore a much shorter acquisition time. We investigated this
strategy of multi phase offset readout DENSE-MRE as a fast acquisition scheme
for multi-slice MRE.Purpose
In this proof of concept study we investigated the
feasibility of obtaining clear wave images from a multi phase offset readout
DENSE-MRE approach to reduce total acquisition time, echo time and
susceptibility artifacts.Methods
All experiments were performed with an agar phantom
with two stiff inclusions on a 3T MRI scanner (MAGNETOM Prisma fit, Siemens
Healthcare, Erlangen, Germany) using a 20ch head coil. The multi phase offset DENSE-MRE sequence was
implemented with modulation and demodulation gradients in slice selection
direction. Figure 1 illustrates the sequence diagram schematically. The
following imaging parameters were used: matrix=128x128, FOV=300x300mm2, slice
thickness=5mm, 15 slices,
parallel imaging GRAPPA factor 2, α=14°, TE=9.5ms, mixing times=12ms+[0/12.5/25/37.5]ms.
Thus 4 phase offset images were acquired consecutively at one run, by matching
the mixing times to get equidistant sample time points of the vibration. The acquisition order of the sampling points was
set in a way to keep the mixing times as short as possible. Rearrangement of
the images on the time axis was performed afterwards. Thus the needed imaging
time for all phase offsets of one slice during one TR was kept minimal in order
to leave more time for interleaved multi-slice acquisition. Total acquisition
time was 76s per frequency. Shear waves were excited by a piezoelectric
driving unit which was connected to a head cradle for brain MRE. Continuous sine
waves with 20Hz/60Hz/100Hz were used in a series of 3 measurements. Vibration synchronization
was guaranteed by a trigger at each TR of 1000ms. To extract the wave propagation, either
subtraction of the mean phase along time direction or picking the first
harmonic after pixelwise Fourier transform of the 4 phase offset images from
time to frequency domain was performed. To achieve estimates of the complex
shear modulus, a multifrequency reconstruction (MDEV [2]) was used.Results
Using
the proposed approach, 60 images (4 phase offsets, 15 slices) could be acquired
within an acquisition time of 76s per frequency. After background phase
removal, the artifact-free phase images clearly showed the propagation of the
mechanical waves (figures 2 and 3). The magnitude of the estimated complex
shear modulus clearly differentiated the two stiff inclusions from the
background (figure 4).Discussion and conclusion
Our
results indicate, that wave images in MRE can be acquired using DENSE together
with a multi readout approach in a fast way. While the use of stimulated echoes
comes along with an SNR reduction, it offers several advantages. The most
significant ones are reduced susceptibility artifacts and shorter echo times.
Both of them are becoming more relevant when moving towards lower vibration
frequencies.Acknowledgements
The
authors gratefully acknowledge Professor Ingolf Sack for providing the MDEV
inversion algorithm.References
[1] Dittmann F et al., MRM
76:1116-1126 (2016)
[2] Hirsch S et al., MRM 71:267-277
(2014)
[3] Muthupillai R et al., Science
269:1854-1857 (1995)
[4] Robert B et al., MRM 62:1155-1163 (2009)
[5] Rump J et al., MRM 57:388-395 (2007)
[6] Aletras AH et al., J Magn Reson 137:247-252
(1999)