Edwin E.G.W. ter Voert1,2, Hannes W. Nagel1, Gaspar Delso3, and Irene A. Burger1
1Nuclear Medicine, University Hospital Zurich, Zurich, Switzerland, 2University of Zurich, Zurich, Switzerland, 3GE Healthcare, Waukesha, WI, United States
Synopsis
When
performing simultaneous single station PET and MR scans in the pelvic region,
the PET acquisition time could be increased to match the usually more extensive
MR protocol acquisition time. The gain in detected coincidences could be
applied to decrease the PET tracer dose and thus the patient’s radiation
burden, while maintaining the same image quality. In this study we investigate
the minimal 68Ga-PSMA dose for a 15-minute single station PET(/MR) scan to match
image quality of the standard 2 minutes scan at full dose.
Purpose
The combination of 68Ga-PSMA (prostate-specific
membrane antigen) PET with
multiparametric MRI (mpMRI) has a high potential for staging prostate cancer
(PCa).1 So far mainly high risk PCa were
investigated with PET/MR with the main focus on detection of lymph node or
distant metastasis for the PET component. For those patients a partial body
scan covering 6 bed positions is usually performed. However, first
results suggest an increased detection rate of primary PCa. Since simultaneous
PET/MR has a field of view of 25cm, a single frame 68Ga-PSMA PET/MRI scan might be enough for local tumor assessment (eg.
PET/MR for biopsy guidance, or focal therapy assessment). Therefore, the PET
acquisition time could be extended from the standard 2 minutes to 15 minutes per
bed, thereby matching the acquisition time of the mpMRI protocol. The increase
in detected PET coincidences could be used to decrease the PET tracer dose and
the radiation burden for the patient. The purpose of this study was to
find the minimal 68Ga-PSMA dose for a 15 minute PET(/MR) scan that
still matches the image quality of a 2 minute scan at full dose without
impairment of lesion detectability.
Methods
We retrospectively sequentially included
10 patients (range 53-78 years) with 68Ga-PSMA PET positive pelvic lymph
nodes undergoing routine
PET/MR for biochemical recurrence. The clinical whole-body PET/MR protocol
contains a 15 minute (1 bed) simultaneous PET and MR examination of the pelvic
area as well as a whole body scan with 2 minutes per bed position. The injected
68Ga-PSMA dose was 150 MBq. Scans were started after 60 minutes uptake
time.
For each patient,
the 15 minute PET frame was reconstructed 12 times: a normal reconstruction
with 3 iterations; a reconstruction of the first 2 minutes with 2 iterations
for reference; and reconstructions simulating 40%, 33%, 27%, 20%, and 13% of
the original dose, performed with 2 and 3 iterations.
Dose reductions
were simulated by removing the indicated amount of counts from the listmode
data. This was performed on a second by second basis to include normal effects like
decay, biodistribution and possibly patient motion. The reconstruction
algorithm was 3D-OSEM based and included all default corrections and
time-of-flight information. The number of subsets was 28 and the image grid
256x256.
For each patient and dataset a region
of interest (ROI) around a lymph node was drawn and the SUVmaxLN was
extracted. The background SUVmeanBG and the standard deviation (STD)
were obtained from a 50mm diameter spherical ROI located in homogeneous
tissue. To assess noise, we used the coefficient of variation (CoV=STD/SUVmeanBG).
The contrast recovery was calculated with the 2 minute 2 iterations reconstruction
as a reference (ref): ((SUVmaxLN/SUVmeanBG)-1)/((SUVmaxLN,ref/SUVmeanBG,ref)-1)).
The reconstructed PET images were also visually assessed.
Results
An example dataset with the 12
reconstructions are shown in Figure 1 and 2. Reconstructions with 3 iterations had the highest
contrast recovery (Figure 3)
even at very low doses. However, the low counts in combination with 3
iterations, introduced an excessive amount of noise: mean CoV was
0.47±0.09 for the 2 minute reference, compared to 0.50±0.11 and 0.56±0.12 for 20% and 13%, respectively (Figure 3). As a result, the PET
positive small lymph nodes could no longer be clearly distinguished from other,
noise based, background hotspots, for datasets lower than 33% of the original
dose. The reconstructions with 2 iterations performed better and allowed for a
dose reduction down to 27% of the original dose.
Discussion
In this study we simulated 68Ga-PSMA
dose
reductions for 15 minute PET scans and its effect on contrast recovery and image
quality compared to the full dose with 2 minute acquisition time. The 40% and
33% dose combined with 3 iterations performed well, with high contrast recovery
and acceptable noise. Further dose reduction to 27% while keeping the same
image quality as the 2 minutes reference scan, including acceptable noise, can
be achieved when reconstructing with 2 iterations, while noise increased
substantially for 3 iterations with the same dose.
A limitation of this study is the
assumption that a reduction in the amount of detected PET coincidences is
directly related to the injected tracer dose. In reality a lower
injected dose could e.g. result in a different biodistribution, leading to
even lower counts in some areas. In future work we could use these results for investigating
real dose reductions.
Conclusion
This
study indicates that the (simulated) 68Ga-PSMA dose for 15 minute
PET scans could be decreased down to 27% of the original dose, while the image
quality remains similar to a standard 2 minutes PET scan with normal dose.
Acknowledgements
We would like to thank T. Deller (GE Healthcare) for the technical support.
References
1. Eiber
M, Weirich G, Holzapfel K, et al. Simultaneous 68Ga-PSMA HBED-CC PET/MRI Improves
the Localization of Primary Prostate Cancer. European urology 2016