To develop in vivo MR oximetry based on T2-prepared bSSFP constant refocusing-pulse interval ($$$\tau_{180}$$$) T2-preparation, three-parameter signal fitting for T2 extraction and magnetization reset for shortened pulse sequence cycle were implemented and evaluated against a multi-spin echo sequence using static and flowing MnCl2 solutions with known T2 values. The whole-blood T2 was also quantified to investigate the effect of varying . The phantom results support three-parameter fitting, and <5% error incurred in T2 quantification with shortened T2-preparation cycle and presence of constant flow. Approximately 10% longer whole-blood T2 was observed with constant relative to varying .
Protocol and analysis
The semi-empirical Luz-Meiboom model relating to HbO2 demands that be held constant for all T2-preparation durations or TEs. The updated T2-preparation (Fig 1) comprises integer multiples of MLEV-4 cycles to vary TEs [1], and T2 is extracted by three-parameter fitting since the T2-prepared bSSFP transient signal approaches a steady-state instead of zero [6]. Further, immediately following the bSSFP readout the magnetization is reset via saturation and allowed to recover for a period of TSat seconds to reduce the interval between successive T2-preparation and therefore shortened TR for faster T2 quantification.
Phantom study: 2- vs. 3-parameter fitting T2 was quantified with constant refocusing-pulse interval ($$$\tau_{180}$$$-fixed) T2-bSSFP (see below) and multi-spin echo pulse sequences in four MnCl2 solutions of different concentration (nominal T2 values ranging from 105 to 200ms); TSat was varied from 1.5 to 3.5 sec. T2 was also measured in the presence of constant flow speeds (5 to 28cm/s) with one of the prepared MnCl2 solutions (T2=135 ms).
In vivo venous blood T2 quantification: fixed vs. varying $$$\tau_{180}$$$
Venous blood T2 quantification in the superior sagittal sinus (SSS) was performed in four subjects using the T2-bSSFP pulse sequence with fixed and varying . The SSS was targeted for its moderate pulsatility and low average velocity (~20 cm/s).
Imaging parameters
All studies were performed on a 1.5T system (Siemens Avanto) with a 12-channel head coil.
T2-preparation
As shown in Fig 1, after magnetization reset and TSat (varying from1.5 to 3.0 sec) T2-preparation is initiated with a 90o RECT excitation pulse, followed by MLEV-4 type refocusing pulses with =12ms, and composite tip-up (270x360-x). The T2-bSSFP version of [2] has one MLEV-4 and increases to control TE, and TR is set to 4s, resulting in 20s acquisition time for five echoes.
bSSFP
FOV 307mm, slice thickness 8mm, TE/TRbSSFP=2/4 ms, flip angle 60o, in-plane resolution 0.8mm, half-Fourier with linear ramp-up signal stabilization and 14 reference lines for reconstruction phase correction, i.e. the k-space center traversed on 25th pulse cycle.
Multi-spin echo
FOV 170mm, slice thickness 5mm, TEs=6.7, 48, 96, 144, 192ms, TRSE=3000 ms, flip angle 90o, in-plane resolution 0.9mm, partial-Fourier factor 5/8.
T2 estimation
T2 values were quantified by fitting the average ROI pixel intensities from each of the five images to either a two- or three-parameter equation, i.e. Ae-Bt or Ae-Bt+C. Corrected TEs were used to reduce the error associated with T1 decay between RECT pulses [7].
Phantom study
Sample images of MnCl2 solutions acquired with T2-bSSFP are shown in Fig 2. The bias introduced with two-parameter fitting is summarized in Fig 3a. Shorter TSat led to a modest reduction in accuracy as a result of lower SNR (Table 1). Fig 3b suggests signal enhancement from steady flow (as high as 28 cm/s) does not affect the accuracy of T2 quantification.
Fig 4 shows sample in vivo images of the SSS in a healthy subject and the bias in the signal decay profile between fixed and varying . If we use the calibration curve of Wright et al [1] T2=132 and 111 ms translate to blood oxygenation of 62 and 55 %HbO2, respectively. The average bias in estimated T2 was approximately 10% (longer with fixed relative to variable ). Data analysis (not shown) indicates that when is varied two- and three-parameter fit gives the same T2 value.