James A. Smith1, Olivier E. Mougin1, Maxim Zaitsev2, Benjamin Knowles2, Richard W. Bowtell1, Paul M. Glover1, and Penny A. Gowland1
1Sir Peter Mansfield Imaging Centre, University of Nottingham, Nottingham, United Kingdom, 2Department of Radiology, University Medical Center Freiburg, Freiburg, Germany
Synopsis
For motion correction using an optical tracking camera,
movements must be converted from the camera reference frame to the MR reference
frame. This calibration transform is determined by matching poses recorded in both
reference frames. We investigated the impact that errors in one or more of
these poses has on the resulting calibration. We then investigated the impact
that errors in calibration has upon the apparent motion recorded. Thus we
outline the necessary precision in the calibration poses to achieve motion correction
of different precisions.
Purpose
MRI Motion correction using optical tracking is increasingly
popular [1]. A fundamental step in
this process is transforming camera space co-ordinates to MR space. This
transform is estimated from a calibration involving various poses recorded in
the camera and MR reference frame simultaneously. We investigated the effect that
translational and rotational errors in estimates of the poses, and the number
of poses have upon the resulting calibration, and the impact that the resulting
calibration errors have on recorded motions.Methods
Poses in the camera reference frame and a calibration transform were simulated to produce corresponding poses in the MR reference frame. Random translational and rotational errors were then added to the MR poses. A non-iterative method was used to obtain apparent calibration transforms [2] [3] using these erroneous MR poses. The RMSE of the apparent calibration transform was taken for both the translational and rotational components. Simulations ran for 1000 iterations for random errors and for data sets containing 4 to 12 poses. To establish the impact that calibration errors have on a subsequent motion estimates, a simulation was performed based on a typical calibration for a 7T Magnex/Nova RF coil combination and motions up to 50mm translation and 10° rotation. Translational and rotational errors were then added to the calibration’s x and y components respectively, and the motion estimates resulting from the use of this ‘flawed’ calibration plotted against the true motions. Results
The max RMSE of the translational and rotational components
of the calibration, for calibrations obtained using poses with translational
errors are shown in Figure 1, either with an error in one pose (Figure 1B &
1D), or all poses (Figure 1A and 1C). The equivalent results for rotational
errors in poses are shown in Figure 2.
The use of flawed calibrations resulted in additional motions
in different directions to the true motion, depending on the error in the
calibration and the true motion (Table 1). Typical examples are shown in Figure
3. Rotational movements transformed using a translationally flawed calibrations
resulted in small off-axis apparent translations. When translational movements were
transformed using rotationally flawed calibrations, the resulting errors were
greater than that of the translationally flawed calibration. Rotational movements
resulted in additional apparent translational and rotational motions. Discussion
Figure 1 shows that there was a linear relationship between
the translational error in a pose and the error in the resulting calibration,
with the error decreasing rapidly with increasing number of poses and little
benefit in using more than 8 poses, whether the error is in one pose or all of
them. The results were less clear for rotational errors in pose (Figure 2). In
general increasing the number of poses increased the calibration accuracy
except for the translation error resulting from errors in multiple poses
(Figure 2A). This result was repeatable and suggests that the inclusion of additional
poses with more errors decreases the stability of the calibration.
Errors in a typical calibration resulted in apparent motions
in different directions to the true motion, since motion in one axis is projected
into another axis. Figure 3 shows translational calibration errors generally
result in smaller translational errors than rotational calibration errors. From Figure 3 we can infer the permitted calibration errors necessary
to correct movements to defined accuracies: smaller calibration errors are
required to correct larger motions to the same accuracy as smaller motions.
Table 2 outlines the maximum calibration error consistent with achieving
reasonable motion correction for large movements found in clinical studies, and
smaller motions in neuroscience studies. Table 2 also shows the approximate average
and maximum pose errors to achieve these calibration accuracies, although this
will vary with the number of poses. The calibration errors however have
been approximated assuming either rotational or translational errors in
calibration: comparing Figure 3A and 3E
suggests that errors in calibration rotations can result in larger impacts from
calibration translational errors, as such the numbers provided in Table 2
represent highest errors consistent with the stated accuracy, in practice more accurate
calibrations will be required and this will be investigated by further monte
carlo simulations of possible movements.Conclusion
Errors in camera/MR calibrations leads to errors in motion
estimates and may lead to reduced image quality by correcting for apparent rather
than real motions. We suggest guidelines for necessary calibration error,
assuming only translational or rotational errors in calibration. The
translational errors in pose to achieve these calibration accuracies vary
greatly with the number of poses unlike rotational errors in which more poses
largely increases resistance to singular flawed poses.Acknowledgements
This work was supported by funding from the Engineering and Physical Sciences Research
Council (EPSRC) and Medical
Research Council (MRC) [grant number EP/L016052/1].References
[1]
M. Zaitsev, C. Dold,
G. Sakas, J. Hennig and O. Speck, “Magnetic resonance imaging of freely
moving objects: prospective real-time motion correction using an external
optical motion tracking system,” Neuroimage, vol. 31, no. 3, pp.
1038--1050, 2006.
[2]
K. Daniilidis,
“Hand-eye calibration using dual quaternions,” The International Journal
of Robotics Research, vol. 18, pp. 286--298, 1999.
[3]
B. Zahneisen, C.
Lovell-Smith, M. Herbst, M. Zaitsev, O. Speck, B. Armstrong and T. Ernst,
“Fast noniterative calibration of an external motion tracking device,” Magnetic
resonance in medicine, vol. 71, no. 4, pp. 1489--1500, 2014.