The relationship between whole-blood T2 and blood hemoglobin oxygen saturation (HbO2) can be modeled as $$${1\over T_{2b}}={1\over T_{2o}}+K(1-HbO_2)^2$$$, where K and T2o are determined empirically in vitro. The feasibility of estimating K and T2o in vivo is investigated with T2-prepared bSSFP at 1.5T in the superficial femoral vein (SFV) with intermittent cuff occlusion. In this manner a range of HbO2 were achieved allowing quantification of venous oxygen saturation via MR susceptometry, a method that had been validated rigorously against blood gas analysis. Initial result (K=19.6Hz, T2o=185ms) is lower than the literature value (24.6Hz, 254ms) but not unexpected because transient bSSFP signal is acquired while disturbing the T2-prepared magnetization.
SFV of two healthy males (40 and 25 yrs old) were imaged at 1.5T (Siemens Avanto) with a single channel Siemens circularly polarized extremity coil. In the older subject the blood T2 in the superior sagittal sinus was also quantified with T2-prepared bSSFP (see below) using 12-channel Siemens head coil.
Cuff paradigm
The cuff paradigm is summarized in Fig 2. Femoral arterial occlusion was achieved by application of super-systolic pressure (200mmHg) in a cuff secured around the thigh (Hokanson, Inc; Bellevue, WA). During the intermittent cuff occlusions the timing of the protocol was set up so as to delay launching of the pulse sequence by two sec following cuff inflation to ensure suspension of blood flow.
%HbO2 quantification with MR susceptometry
In MR susceptometry [4] HbO2 is quantified by making use of the paramagnetism of venous blood, which scales with the fraction of deoxyhemoglobin, $$$(1-HbO_2)$$$ , which in turn determines the relative magnetic susceptibility, $$$\Delta\chi$$$, of deoxygenated blood. $$$\Delta\chi$$$ is obtained from the measured field shift, $$$\Delta B$$$, relative to the surrounding diamagnetic tissue (typically muscle, which serves as a self-calibrating reference). The field map is obtained from the phase difference image, $$$\Delta\phi$$$, computed from two gradient echoes separated by $$$\Delta TE$$$, $$$\Delta\phi=\gamma\Delta B\Delta TE$$$, where $$$\gamma$$$ gyromagnetic ratio of water protons. Once hematocrit Hct is known $$$HbO_2$$$ can be computed as $$$HbO_2=[1-\frac{2\Delta\phi}{\gamma\Delta\chi_{do}\Delta TE\cdot Hct\cdot B_o(cos^2\theta-1/3)}]$$$, where $$$\Delta\chi_{do}=4\pi(0.27 ppm)$$$ is the magnetic susceptibility between fully deoxygenated and fully oxygenated blood in SI units and $$$\theta$$$ is the vessel tilt angle with respect to $$$B_o$$$. Imaging parameters of RF-spoiled multi-echo GRE: FOV 144mm, matrix size 144, slice thickness 5mm, TE/ΔTE/TR=5.7/8.88/38.5ms, FA=16o, partial Fourier factor 0.83 and acquisition time <5s.
T2 quantification
After RF-spoiled multi-echo GRE acquisition, the magnetization was reset to zero and allowed to recover for 2.5 sec prior to initiating T2-preparation, which consisted of 90o RECT excitation pulse, MLEV-4 type refocusing pulses with $$$\tau_{180}$$$=12ms, and composite tip-up (270x360-x). The duration of MLEV-4 was 48ms and comprises four composite refocusing pulses, each consisting of 90x180y90x RECT pulses separated by 150μsec. The amplitude of all RECT pulses was fixed and the duration was 500μsec/90o, e.g. duration of 180y is 1000μsec . TEs=0, 48, 96, 144 and 192ms correspond to 0, 1, 2, 3 and 4 MLEV-4 refocusing pulses, respectively. Imaging parameters of bSSFP: FOV 144mm, slice thickness 8mm, TE/TR=2/4ms, flip angle 60o, in-plane resolution 0.8mm, half-Fourier with linear ramp-up signal stabilization and 14 reference lines for reconstruction phase correction, i.e. k-space center is collected on 25th pulse cycle; acquisition time ~15s. T2 was quantified by fitting the average pixel intensities S of ROI from each of the five images to the three-parameter equation [5], i.e. S=Ae-Bt+C, and corrected TEs (0, 43.9, 87.7, 131.6, 175ms) were used in the fitting to reduce error associated with T1 decay between RECT pulses [6].