Ryan Topfer1, Alexandru Foias1, Nikola Stikov1,2, and Julien Cohen-Adad1,3
1NeuroPoly Lab, Institute of Biomedical Engineering, Polytechnique Montreal, Montreal, QC, Canada, 2Montreal Heart Institute, Université de Montréal, Montreal, QC, Canada, 3Functional Neuroimaging Unit, CRIUGM, Université de Montréal, Montreal, QC, Canada
Synopsis
Pathologies of the spinal cord are a primary cause of
functional disability and chronic pain. Although MRI already plays a role in
the evaluation of these pathologies, it continues to be hampered by artifacts due
to magnetic field inhomogeneity. This study reports the first results from applying
a specially designed 24-channel shim array to compensate respiration-induced magnetic
field inhomogeneity in the human spinal cord in real-time. This approach has
the potential to improve the quality of EPI and spectroscopy in the spinal
cord.
Introduction
The sensitivity of MRI to magnetic susceptibility is
both a blessing and a curse: While small-scale changes in the magnetic field
permit useful image contrasts, such as the blood-oxygenation-level dependent
(BOLD) effect, larger scale field distortions, such as those due to air-tissue
interfaces, can cause image artifacts such as signal loss in gradient-echo (GRE)
imaging and geometric distortion in echo-planar imaging (EPI). The proximity of
the spinal cord to the lungs – the body’s most extensive internal air-tissue
interface – makes it particularly susceptible to these field effects, which generally
vary in time, along with respiration (e.g., Figure 1).1–3 Given the extent of these artifacts, along with the inherent
subtlety of the signal changes associated with BOLD activation (about 1%), fMRI
of the spinal cord remains technically challenging, as underscored in a recent
review: “respiration-induced change in B0 is one of the main reasons very few
people dare to venture into fMRI experiments below the cervical level.”4 This study presents the first results updating shims
in real-time to correct for respiration-induced distortions in the spinal cord.Methods
Real-time shimming hardware is exhibited in Figure 2.5 Calculation of the shim updates was informed by the
work of Van Gelderen et al.6 which suggested a possible linear relation between the
time-varying magnetic field and the respiratory cycle as tracked by a pressure
reading from respiratory bellows strapped to the subject’s
abdomen: $$$
\frac{ \partial b_{\chi} (\textbf{r}, t) }{ \partial t } = c_{\chi}(\textbf{r})
\frac{ \partial p(t) }{ \partial t }
$$$, which, when
integrated, yields $$$
b_{\chi} (\textbf{r}, t) = c_{\chi}(\textbf{r})p(t) +
b_{\chi|o}(\textbf{r})
$$$. It follows that
the coupling coefficients $$$c_{\chi}(\bf{r})$$$ and the static
field offset $$$b_{\chi|o}(\textbf{r})$$$ might be
estimated by GRE field maps acquired in as few as two distinct apneas—a sort of
calibration procedure by which all future field perturbations can be inferred
based on the associated pressure reading (Figure
3).
Experiments: Six healthy subjects were scanned on a 3 T system (Siemens
TIM Trio). Standard 2nd-order
static shimming was first performed using the MRI shims. To calibrate the
real-time shim updates, GRE field maps were acquired in inspired and expired
respiratory states (e.g. Figures 1 and 3)
with acquisition and phase processing parameters similar to those described in a
previous work.5 To demonstrate the utility of real-time shimming, a
2D sagittal GRE-EPI sequence typical of fMRI was performed during normal
respiration, with and without real-time shimming, for which shim updates were
issued every 250 ms based on the respiratory trace sampled at 100 Hz. The EPI
sequence used the following parameters: TE=15 ms, TR=500 ms; flip angle 70°; bandwidth=1502
Hz/pixel; R=2 acceleration factor using GRAPPA; partial Fourier=6/8; spatial resolution=1.5x1.5
mm2 in-plane, with 7 sagittal slices of thickness 5.0 mm, for an
effective FOV=35x156x192 mm3; 100 image volumes were acquired, for a
total acquisition time of 50 s. The temporal signal-to-noise ratio (tSNR) of
the EPI data was calculated as the average voxel intensity across the time
series divided by its standard deviation.Results and Discussion
Over the spinal cord, real-time shimming improved the
mean EPI tSNR by 18.9 +/- 13.2 %, while
increasing tSNR over the entire region considered in the shim optimization by 12.2
+/- 10.1%. Figure 4 summarizes
these improvements and Figure 5 shows
visual results for the first two subjects.
This study demonstrates the benefit of real-time
shimming for spinal cord EPI. Regarding improvements to tSNR, a considerable
intersubject variability was observed likely due to anatomical differences that
determine the extent of respiration-induced distortion. Though increased tSNR
was observed across subjects, the real-time protocol could nevertheless be
improved, for example, by updating the shims more frequently, and by accounting
for transmission delays. Furthermore, apart from ignoring other contributions
to the time-varying magnetic field (e.g. eddy currents, cardiac pulsation), the
validity of the linear pressure-to-field relation used to update the shims has been
taken for granted; other models of the time-evolving field,7 or means of tracking it,8–10 may ultimately prove more appropriate.
In addition to signal enhancement and artifact
reduction for EPI, real-time shimming stands to benefit T2*-weighted imaging
generally, as well as spectroscopic studies of the spinal cord.Acknowledgements
We thank the team at Resonance Research Inc. – Piotr Starewicz,
Kai-Ming Lo, Karl Metzemaekers, and Donald Jette – for their work constructing
the shim hardware.
Funded by the Canada
Research Chair in Quantitative Magnetic Resonance Imaging (JCA), the
Canada Foundation for Innovation [32454], the Canadian Institutes
of Health Research [CIHR FDN-143263], the Fonds de Recherche du Québec - Santé
[28826], the Fonds de Recherche du Québec - Nature et Technologies
[2015-PR-182754], the Natural Sciences and Engineering Research Council of
Canada [435897-2013], the Quebec BioImaging Network, and Polytechnique MEDITIS.References
1. Marques, J. P. & Bowtell, R.
Application of a Fourier-based method for rapid calculation of field
inhomogeneity due to spatial variation of magnetic susceptibility. Concepts
Magn Reson Part B Magn Reson Eng 25B, 65–78 (2005).
2. Verma, T.
& Cohen-Adad, J. Effect of respiration on the B0 field in the human spinal
cord at 3T. Magn Reson Med 72, 1629–36 (2014).
3. Vannesjo, S.
J. et al. Breathing-induced B0 field fluctuations in the cervical spinal
cord at 7T. in Proceedings of the 24th Annual Meeting of ISMRM, Singapore,
2016. Abstract 49.
4. Cohen-Adad, J.
Functional MRI of the Spinal Cord: Current Status and Future Developments. Semin.
Ultrasound, CT MRI 5675–5679 (2016). doi:10.1053/j.sult.2016.07.007
5. Topfer, R. et
al. A 24-channel shim array for the human spinal cord: design, evaluation
and application. Magn Reson Med (2016). doi:10.1002/mrm.26354
6. van Gelderen,
P., de Zwart, J. A., Starewicz, P., Hinks, R. S. & Duyn, J. H. Real-time
shimming to compensate for respiration-induced B0 fluctuations. Magn Reson
Med 57, 362–8 (2007).
7. Bianciardi, M.
et al. Evaluation of dynamic off-resonance correction of respiratory
instability in MRI signals for high-order spherical harmonic basis set and
multivariate modeling of respiratory sources. in Proceedings of the 22nd
Annual Meeting of ISMRM, Milan, 2014. Abstract 1623.
8. Boer, V. O.,
van de Bank, B. L., van Vliet, G., Luijten, P. R. & Klomp, D. W. J. Direct
B0 field monitoring and real-time B0 field updating in the human breast at 7 Tesla.
Magn Reson Med 67, 586–91 (2012).
9. Haeberlin, M. et
al. Real-time motion correction using gradient tones and head-mounted NMR
field probes. Magn Reson Med 74, 647–660 (2015).
10. Duerst, Y. et
al. Utility of real-time field control in T2*-Weighted head MRI at 7T. Magn
Reson Med 76, 430–439 (2016).