Martin Berzl1,2, Antoine Pfeil1, Craig Meyer3, Adrienne Campbell-Washburn4, Gregor Körzdörfer1, Mathias Nittka1, Andreas Maier2, and Josef Pfeuffer1
1Application Development, Siemens Healthcare, Erlangen, Germany, 2Pattern Recognition Lab, Friedrich-Alexander University Erlangen-Nürnberg (FAU), Erlangen, Germany, 3Biomedical Engineering, University of Virginia, Charlottesville, VA, United States, 4Biochemistry and Biophysics Center, Division of Intramural Research, National Heart Lung and Blood Institute, National Institutes of Health, Bethesda, MD, United States
Synopsis
The purpose of this study is to evaluate
different spiral trajectory prediction models - isotropic, Tan-Meyer and GIRF -
to mitigate image artifacts for spiral MRI and improve accuracy of quantitative
T1/T2 values for MR Fingerprinting. GIRF scan parameters were optimized to
allow a total measurement time of only six minutes for a one-time calibration.
GIRF similarly provided excellent results for vastly different trajectory types,
varying in max. slew rate, gradient amplitude and number of interleaves, and
showed some advantages against Tan-Meyer for trajectory designs with high
k-space center slew rate, both for qualitative and quantitative results.
Purpose
Spiral trajectories are prone to blurring or
image distortion artifacts caused by MR system imperfections like residual
gradient delays and eddy currents. As a reference, trajectories can be measured
and used for reconstruction, which is a time-consuming procedure and needs to
be done for each different trajectory design, i.e. orientation, resolution etc1.
Using a generalized model to predict the trajectory - e.g. as proposed by
Tan-Meyer2 – was shown to improve on this using a one-time
calibration. This work evaluates the Tan-Meyer model and compares it to another
trajectory prediction method using convolution with the gradient impulse
response function (GIRF)3,4,5. The method was optimized to acquire a
3-axis GIRF by MR measurements in a total time of six minutes. Image artifacts
could be reduced significantly; also MR Fingerprinting (MRF) results were shown
with more accurate quantitative T1 and T2 values.Methods
MR measurements were performed on a MAGNETOM
Skyra 3T (Siemens Healthcare, Erlangen, Germany). Different trajectories were designed
using the Hargreaves algorithm (Traj-H)6 and an improved variant of
the dual-density Meyer algorithm (Traj-M)7. Experimental
trajectories were measured with a modified Duyn method1.
Different models were used to calculate the
trajectory for the reconstruction. 1) isotropic model (zero-order gradient
delay), 2) Tan-Meyer model2 (anisotropic zero-order
gradient delays and residual eddy currents), 3) GIRF model3. Assuming
linearity and time-invariance, gradients are predicted by a convolution with
the GIRF. MR measurements8 were thoroughly optimized for minimal
scan time at same performance on a MAGNETOM Aera 1.5T (Siemens Healthcare,
Erlangen, Germany): 4 averages, TR 500 ms, dwell time 2.5 us, offcenter-shift-distance 50 mm
and slice-thickness 2 mm allowed to measure the 3-axis-GIRF within total six
minutes.
Model-calculated trajectories were compared to
the measured ones using Traj-H and Traj-M spiral designs. Deviations were
evaluated in the k-space, the images and in quantitative values from MRF using
a NIST system phantom (High-Precision-Devices, Boulder, CO). 2D-MRF-FISP9
data were acquired using a prototype sequence. Five different trajectory types
(with varying max. slew rate, gradient amplitude and interleaves) were assessed using the
normalized mean absolute error (NMAE) in k-space.Results
Hargreaves and Meyer trajectory designs are
shown in Figure 1. The modified Meyer design (Traj-M) increases the slew
gradually at k-space center, whereas Traj-H starts at maximum slew rate.
The residual k-space error calculated with the
Tan-Meyer and GIRF model is significantly lower compared to the isotropic model
(Figure 2). Specifically, at k-space center, GIRF showed some advantages over
Tan-Meyer. The high slew rate in the Traj-H trajectory design can be better
predicted.
Different trajectory types and models were
evaluated in k-space by NMAE relative to the isotropic model (Figure 3). Both
the Tan-Meyer and GIRF models obtain very low NMAE values for all trajectory
designs, with the GIRF values slightly smaller.
Images were reconstructed with different
trajectories and compared to the image from the measured trajectory (Figure 4).
The isotropic model shows a spatial object scaling (edge) and rotation image
artifact, which is visible at the phantom border. Tan-Meyer was able to correct both artifacts
very well, but left small intensity deviations, which were
not present with GIRF.
Quantitative T1/T2 values obtained with MRF
in the NIST phantom have close to zero deviation from the values obtained by
using the measured trajectory (Figure 5). Isotropic and Tan-Meyer do not differ
significantly from each other and deviate from the zero line. Tan-Meyer
provided better results for Traj-M compared to Traj-H design. GIRF is able to
improve accuracy of quantitative MRF values, especially T2.Discussion
The responsible source of image quality
improvement using the GIRF estimation is the better prediction in k-space
center, where most of the image information is present in MRI. Isotropic and
Tan-Meyer could not reduce all image signal intensity deviations or deviations
from quantitative T1 and T2 values in MRF. As shown, the GIRF is able to cope
with the high slew rate at the center of the Hargreaves trajectory. GIRF is
also a more generic model to predict other trajectory designs in an excellent
manner.Conclusion
The GIRF model works well to correct for
trajectory deviations and image artifacts arising from these. It also improves
accuracy of quantitative values generated by MRF. It is general in terms of
trajectory designs even with near-limit slew rates. A small residuum was still
present in image space, which can potentially be addressed by off-resonance and
Maxwell corrections. A one-time system calibration for a robust GIRF model
takes only six minutes – without any additional equipment – making the approach
very promising to extensively improve spiral imaging quality and quantitative
mapping using MRF.Acknowledgements
We
gratefully thank Prof. Mark Griswold and his group and collaborators at
Case Western Reserve University, Cleveland, OH for support and for many
intensive discussions.References
1. Duyn J, Yang Y, Frank J, et al. Simple
Correction Method for k-Space Trajectory Deviations in MRI. Journal of Magnetic
Resonance. 132, 150–153 (1998).
2. Tan H, Meyer C: Estimation of k-Space
Trajectories in Spiral MRI. Magnetic Resonance in Medicine. 2009;61:1396-1404.
3. Campbell-Washburn A, Xue H, Lederman R, et
al. Real-Time Distortion Correction of Spiral and Echo Planar Images Using the
Gradient System Impulse Response Function. Magnetic Resonance in Medicine.
2016;75:2278-2285.
4. Vannesjo J, Haeberlin M, Kasper L, et al. Gradient
System Characterization by Impulse Response Measurements with a Dynamic Field
Camera. Magnetic Resonance in Medicine. 2013;69:583-593.
5. Vannesjo J, Graedel N, Kapser L, et
al. Image Reconstruction Using a Gradient Impulse
Response Model for Trajectory Prediction. Magnetic Resonance in Medicine.
2016;76:45-58.
6. Hargreaves B. Variable-Density Spiral Design
Functions. Sept 2000.
http://mrsrl.stanford.edu/~brian/vdspiral. Accessed
October 2016.
7. Meyer C, Zhao L, Lustig M, et al:
Dual-density and parallel spiral ASL for motion artifact reduction. Proceedings
ISMRM 2011;3986.
8. Robison R, Pipe J: Linear Gradient System
Characterization Using a Rapid Thin-slice Measurement. Proceedings ISMRM
2016;4283.
9. Jiang Y, Ma D, Seiberlich N, et al. MR fingerprinting
using fast imaging with steady state precession (FISP) with spiral readout.
Magnetic Resonance in Medicine. 2015;74(6):1621-1631.