Eva S Peper1, Wouter V Potters2, Abdallah G Motaal3, Pim van Ooij1, Aart J Nederveen1, Gustav J Strijkers4, and Bram F Coolen4
1Department of Radiology, Academic Medical Center, Amsterdam, Netherlands, 2Neurologie en Klinische Neurofysiologie, Academic Medical Center, Amsterdam, Netherlands, 3Imaging Clinical Applications and Platform, Philips Healthcare, Netherlands, 4Department of Biomedical Engineering & Physics, Academic Medical Center, Amsterdam, Netherlands
Synopsis
Measuring
PWV with MRI would provide a useful tool to measure arterial stiffness locally.
However, the limitation of MRI for this implementation is temporal resolution.
This study validates a technique to measure local PWV at up to only 5 cm of the
carotid arteries using 2D PC CINE MRI data of a high temporal resolution
compressed sensing (CS) reconstruction. The method is validated using a
pulsatile flow phantom and two groups of 10 elderly and 10 younger healthy
volunteers. A significant difference between age groups was found.
Introduction
We
developed a method to measure local pulse wave velocity (PWV) in the carotid
arteries using compressed sensing (CS) for a high temporal resolution (4ms)
reconstruction of 2D PC CINE MRI. The method was validated in a flow phantom,
and in two groups of 10 elderly and 10 younger healthy volunteers. Methods
Sequence: 2D PC-MRI
data was acquired with an 8-channel carotid coil with flow encoding in
FH-direction (VENC 150 cm/s). Scans were acquired with a spatial resolution of
0.86x0.86x3 mm3 and TR/TE/FA = 8.0ms/3.9ms/8°. Retrospectively triggering
was done using an ECG signal. For the flow phantom ECG was provided by an
external trigger synchronized with the flow pump. To measure PWV, two slices
perpendicular to the common carotid artery (CCA) and internal carotid artery
(ICA) with an inter-slice distance of typically 7 cm were acquired (figure 1).
Flow phantom: The flow
phantom (LifeTec Group BV, The
Netherlands), consisted of a silicon tube of 8 cm length, mimicking a human
carotid bifurcation (figure 1). A pulsatile water flow was created at a heart
rate of 60 beats/minute and a simulated heart rate variability of 10%. A
temporal mean flow of 300 ml/min was created with a peak systolic flow rate of
1000 ml/min in the CCA. Flow was also measured with a gold standard ultrasound
echo probe (UEP) before and after the phantom connections. In vivo scans: In-vivo scans were performed in 10 young (age = 31±6,
5 women) and 10 elderly (age = 62±10, 5 women) healthy volunteers.
Reconstruction
and PWV analysis: Raw
data was processed using Matlab (MathWorks, USA) and MRecon (Gyrotools, Zürich). Data was
retrospectively binned to cardiac frames of 4 ms, giving on average 250 frames/heartbeat.
Note that the narrow temporal binning led to gaps in k-t space corresponding to
an undersampling factor of 5 on average (figure 2a). CS reconstruction was performed
using the BART toolbox1. A L1 regularization was performed under a total
variation (TV) sparsifying transform over time with a regularization parameter
of 0.2. PWV was calculated and compared for two methods, foot-to-foot (FF) and
correlation, measuring the time difference between flow curves2,3. Additionally
a normalization or a sigmoid fit was applied providing more information on the
accuracy of the measured time shift (figure 3). The path length between the two
slices were determined on 3D black-blood anatomical MR images using OsiriX4.Results
Figure
3 shows the mean flow in the phantom vessel region plotted over time, as well
as the time shift measurements using foot-to-foot (a) and correlation methods
(b). PWV in the pulsatile phantom was in good agreement with PWV measured with
the UEP (figure 4). On average, local carotid PWVs for young (5.7±1.7 m/s) and
elderly (7.9 ±1.5 m/s) healthy volunteers were significantly different
(p=0.031). Comparing the results of each method independently all were
significant, except for the FF method including a sigmoid fit (p=0.053).Discussion
For PWV, high temporal resolution is needed to resolve small time
shifts that occur within a small vessel region. In order to measure flow with a
temporal resolution of 4 ms, k-t space was undersampled randomly by exploiting
heart rate variability5. Variable density sampling patterns were implemented
to ensure sufficient center k-space filling for all reconstructed frames.
For
some methods (FF and FF
normalized) we noticed that the UEP measurement could not be reliably
reproduced in the phantom (figure 4). The idea was however to include all
methods in the calculation of PWV to ensure they on average provide a stable
result. We measured significantly higher PWV in the elderly volunteers than in
the younger ones independent of the method. The standard deviation of measured
PWVs in the elderly group was higher, which might reflect the more
heterogeneous stiffening of the arteries with age. Average PWV values were in
agreement with previously reported numbers (5.7 and 6.9 m/s6,7) determined
from flow curves in aorta and carotids in subjects with similar age.Conclusion
In this study a high temporal resolution
pulse wave velocity measurement method was presented, which enables PWV
quantification in short vessel segments, such as locally in the carotid arteries.
The feasibility of this technique was
demonstrated in a flow phantom and in a comparison of two groups of elderly and
younger healthy volunteers. Results are in line with similar studies on larger
distances in the carotid artery. This new technique opens up
new opportunities for studying and characterizing local vessel disease.Acknowledgements
No acknowledgement found.References
1. M. Uecker et al., Berkeley advanced reconstruction toolbox., Proc. Intl. Soc. Mag. Reson. Med. 2015
2. A.L. Wentland et al., Review of MRI-based measurements of pulse wave
velocity: a biomarker of arterial stiffness. Cardiovasc. Diagn. Ther., 2014,
4:193-206
3. A. Dogui et al., Measurement of Aortic Arch Pulse Wave Velocity in
Cardiovascular MR: Comparison of Transit Time Estimators and Description of a
new Approach. Journal of Magn. Res. Imaging, 2011, 33:1321-1329
4. A. Rosset et al., OsiriX: An Open-Source Software for Navigating in Multidimensional
DICOM Images. Journal of Digit. Imaging, 2004, 17:205-216
5. B. F.
Coolen et al., High resolution frame rate retrospectively triggered Cine MRI
for assessment of murine diastolic function. Magn. Res. Med., 2013, 69:648-656
6. E. Kröner et al., Pulse Wave Velocity and Flow in the Carotid Artery
Versus the Aortic Arch: Effects of Aging. Journal of Magn. Res. Imaging, 2014,
40:287-293
7. W. A. Riley et al., Ultrasonic Measurement of the Elastic Modulus of
the Common Carotid Artery. The Atherosclerosis Risk in Communities Study.
Stroke, 1992, 23:952-956