Ranajay Mandal1, Nishant Babaria2, Jiayue Cao1, and Zhongming Liu1,2
1Biomedical Engineering, Purdue University, West Lafayette, IN, United States, 2Electrical and Computer Engineering, Purdue University, West Lafayette, IN, United States
Synopsis
The field of multimodal imaging has largely been propelled by two of the most widely used neuroimaging tools, fMRI and EEG, as the complementary nature of the two signals provides a unique avenue to evaluate the brain dynamics and its underlying neural circuitry. However, methodological challenges associated with simultaneous acquisition of EEG and fMRI impedes the method from achieving the full potential. To address these challenges, we present an MR-compatible, active recording system that utilizes surplus MR-hardware and inherent electromagnetic field to synchronize and wirelessly record gradient artifact free multichannel EEG signals and encode these Non-MR signals within the extended FOV of the MR image.
Introduction
Non-invasive functional imaging tools have had a
significant impact on our understanding of brain dynamics as a part of many
clinical and research settings. Although the spatial and temporal resolution of
different modalities have improved significantly over the past decade, major
theoretical limitations on increasing resolution have motivated the need for
integrating multiple neuroimaging modalities. Due to the
complementary nature of the two signals, the integration of fMRI and EEG has
gathered the most attention and thus has opened new avenues to understand large-scale network responses associated with neural events. However,
concurrent acquisition of EEG, MEG or any other electrophysiological signal
within the MR environment poses some stern challenges. Conventional EEG systems suffer from severe
electromagnetic interference induced by static and dynamic magnetic fields due to passive sensing and long wired connections, while
being expensive, bulky, and of potential safety concern when used in high-field
MRI. Here, we demonstrate a novel hardware system to eliminate the effects of
electromagnetic interference, providing low-cost, high-density, and MR-safe EEG
recordings with significant improvement in signal quality.Methods
The MR-compatible recording system functions
through the use of novel methods that maximally utilize existing MR-hardware
and its electromagnetic environment for both synchronization and wireless
transmission of non-MR data (Fig. 1). The system is designed to operate in
either bipolar or unipolar configuration. The active components combined with
minimized cable length and differential signal transmission cancel out
electromagnetic interference, while reducing noise and compensating for the
signal attenuation. Gradient and RF pulse detection circuit, including on-board
pickup coil, amplification, and filtering circuit, reliably detect the ramp
periods of the trapezoidal gradients and RF excitation duration to produce
logic signals that indicate the electromagnetically silent periods. Utilizing
these logical signals stepping stone sampling1 was implemented. Effect of the RF interference and
gradient magnetic field changes was minimized by blocking input signals using
high fidelity single-pole-double-throw (SPDT) analog switching circuits. High
resolution (0.5µV), synchronized digitization circuit (16-bit ADC) provided
necessary dynamic range (33mV) while maintaining very low power consumption. An
ultra-low power transmitting module wirelessly transmits the digitized EEG
signals through FSK demodulation at discrete non-overlapping RF frequencies. Surplus
bandwidth of receiver coil, availed by increasing the imaging field of view
(FOV) in the readout direction2, is used to accommodate these non-MR
signals (Fig.3.b). Upon reception by MR coils, the RF signal is demodulated and
separated into EEG signals in one frequency range and MRI data in a different
range. The transmitted data is then extracted through standard FSK demodulation
algorithm and a locally developed software.Results
Figure 2(a) shows wireless gradient detection
using on-board pickup coil and corresponding switching triggers. A simple
illustration of active sensing and subsequent wireless transmission of rat LFP
is shown in Figure 2(b). Anesthesia induced burst suppression
of spontaneous rat LFP is demonstrated in Figure 2(b)(Top). As concentration isoflurane is increased,
resulting in deeper anesthesia, spontaneous LFP progressively transformed
towards burst suppression. Fore-paw monophasic current stimulus (1mA) of
2ms duration at 5Hz, generated evoked LFP response in somatosensory cortex is
recorded through the device wirelessly (Fig.2.b, bottom). Figure 3(a) displays
reliability of extracted EEG signal demodulated from MR raw data compared with
original transmitted data. Single channel FSK demodulation of EEG signal within
the MR image is demonstrated in Figure 3(b) as non-MR data appears as stripes
in besides the original phantom image.Conclusion
The MR-integrated
system provides a simple solution for high fidelity neural recording in
simultaneous fMRI studies by utilizing MR-surplus hardware. Gradient triggered
sampling and analog switching circuit combined with wireless reception of EEG
signal by MR coil address technical challenges regarding the signal integrity
and electromagnetic artifacts, while reducing the overall complexity by
removing the dependence on bulky synchronization and shielding systems. Additionally,
further refinement and implementation of the system will be focused on future
human applications. Therefore, the success of the current research is expected
to open new avenues for widely accessible, integrative neuroimaging tools.Acknowledgements
No acknowledgement found.References
1. K. Anami et
al., “Stepping stone sampling for retrieving artifact-free electroencephalogram
during functional magnetic resonance imaging,” Neuroimage, vol. 19, no.
2, pp. 281–295, 2003.
2. L. G. Hanson, T. E. Lund, and C. G. Hanson, “Encoding of
electrophysiology and other signals in MR images,” J. Magn. Reson. Imaging,
vol. 25, no. 5, pp. 1059–1066, 2007.