Jan Ole Pedersen1,2, Christian Hanson, Rong Xue3, and Lars G. Hanson1,2
1Danish Research Centre for Magnetic Resonance, Centre for Functional and Diagnostic Imaging and Research, Copenhagen University Hospital, Kgs Lyngby, Denmark, 2Centre for Magnetic Resonance, DTU Elektro, Technical University of Denmark, Kgs Lyngby, Denmark, 3State Key Laboratory of Brain and Cognitive Science, Research, Institute of Biophysics, Chinese Academy of Sciences
Synopsis
During MRI, fast
switching of imaging gradients generate loud noise of high intensity
due to vibration of the gradient coils. The in-bore intercom used for
patient communication is therefore typically turned off during
scanning. This has implications for safety and image quality since
patient speech and yells are not heard by the scanner operator. Using
standard sequences, we demonstrate that sound can be recorded by MRI
scanners and extracted from the scanners raw data, thereby enabling
communication with patients for safety or experimental reasons.
Introduction
For most MRI
sequences, acquisition is performed without ramp-sampling and
therefore in silent periods. Using the scanner itself for acquisition
of a sound signal alleviates timing challenges arising when using
external equipment. This is however not directly feasible, as the
filters of the scanner attenuates signal outside a narrow range
around the Larmor frequency. However, by performing an amplitude
modulation of the sound signal to a frequency close to the Larmor
frequency allows the signal to pass the filters of the scanner, and
the demodulation performed by the scanner restores the original sound
signal except for a small frequency offset. The sound can then be
extracted from the raw data of the scan. The constant timing between
sampling and gradient waveforms allow for easy filtration of residual
gradient induced noise in the sound signal. Using a dedicated receive
channel of the scanner allow for sampling of the sound signal with
high SNR and without interference from MR signals.
Method
Two 2”
loudspeakers were sacrificed to make a simple MR-compatible
microphone: A loudspeaker coil mounted on a membrane was placed in a
tube inside the static field (B0) of the scanner. Membrane vibration
thus generated a coil voltage, i.e. a sound signal. An identical coil
with no membrane was mounted in series with similar orientation and
close to the microphone coil to cancel out gradient-induced voltages.
The coils were
connected to a modulator circuitry developed previously[1] for amplitude modulation of non-MR signals for
wireless recording by scanners. Due to a maximum carrier frequency of
130MHz producible by the modulator, the signal was send through a
200MHz mixer to reach the frequency range of the scanner (hydrogen at
7T, Siemens Magnetom). To remove any DC component, the modulated
signal was sent via two inductively coupled coils, before reaching a
receive channel of the scanner. A 50 \ohm resistor in parallel with
the transmission line of the modulator made the setup recognizable by
the scanner as a coil element. The sound signal was send by the
modulator continuously, but only sampled by the scanner during
acquisition periods.
During a GRE
sequence (256x128 receive matrix, 5 volumes) a subject was asked to
repeatedly say out loud “Mary Had a Little Lamb” (chosen for
melodramatic effect, as the verse was famously used by Edison for the
first sound recording ever). Non-linearities from the modulator were
compensated before the DC component was removed. The phase-locked
noise from the read-out gradient and noise from the phase-encoding
gradient were relatively easily estimated and subtracted. The
non-equidistant sampling in time of the GRE sequence left periods
without sound data. Autoregressive modeling across short subsets of
the data, where the noise was assumed to be unchanged, was used to
fill these gaps.Results
A
sound recording with easily
recognizable repetition of
“Mary Had a Little Lamb”,
acquired by an MRI scanner was obtained
after removal of residual
scanner noise. In contrast
to direct audio recording, gradient noise is barely audible.
Discussion & Conclusion
Through the use of a homebuilt MR compatible microphone and amplitude
modulation, we have shown it possible to record sound signals using a
receive channel of an MRI scanner. The limited signal processing
needed to obtain recognizable vocalization allow for fast signal
processing, and thereby real-time oral communication with a patient
during scanning. Here a separate channel of the scanner was used for
receiving the sound signal. As previously shown
[1], it is possible
to transmit a signal wirelessly at a frequency in the oversampled
range of the scan (sampled by the scanner, but not in FOV), and
thereby receive the sound MRI signal on common receive channels.
Acknowledgements
No acknowledgement found.References
[1] Hanson LG;Lund TE;Hanson CG.
Encoding of electrophysiology and other signals in MR images.
J Magn Reson Imaging 2007, 25(5), 1059-1066