A new highly flexible RF coil design is described. Individual coil elements are constructed by means of a proprietary process that yields low reactance and low loss conductors while being lightweight, flexible, and durable. Phantom data indicate that for similar coil geometries, the coils provide equivalent SNR profiles but do not suffer from design limitations due to minimum coil overlap requirements of conventional copper coil elements. In vivo data demonstrate that these coils can be fabricated onto highly flexible and thin support materials that closely comply with anatomic contours in challenging imaging sites such as the c-spine.
RF Coil design: One, two and four element ‘AIR Tech’ (GE Healthcare, Waukesha, WI) receive-only RF arrays were constructed. The diameter of each element was fixed and equal to 11 cm and included an associated microelectronic tuning circuit. Individual loops are manufactured using a proprietary process that yields low reactance and low loss conductors, while being exceptionally lightweight, flexible, and durable. Figure 1 shows a four element array and associated miniaturized supporting electronics. A unique feature of this coil design is that the conductor and electronics combination serve to control inter loop interactions. This allows for coil loop overlaps2 that could not be achieved with conventional copper-based elements without the introduction of artifacts and significantly reduced signal-to-noise ratio (SNR).
Phantom experiments: All experiments were conducted on a 3.0T whole body 60 cm bore diameter MR scanner (Signa EXCITE 16.0, GE Healthcare). An electrically non-loading doped rectangular silicone phantom was used to collect SNR and signal intensity profile data. Measurements were obtained using a T1-weighted spin echo pulse sequence (TE/TR = 15 msec/750 msec, FOV = 32 cm, slice thickness = 5mm, signal averages = 1, matrix = 256 x 256 : frequency x phase). Data was collected from double loop coil configurations (Figure 2) and compared to a pair of traditionally constructed coils at the nominal coil center to center distance (86 mm). Data were also collected with the AIR Tech coils over a range of coil center spacings.
In vivo experiments: The four channel array shown in Figure 1 was used for in vivo data collection. 3T (Signa EXCITE 16.0, GE Healthcare) T1- (TE/TR = 9 msec/750 msec, matrix = 352 x 256 : frequency x phase, FOV = 22 cm, slice thickness = 3.5 mm, signal averages = 1, echo train length = 3) and T2-weighted (TE/TR = 107 msec/3200 msec, matrix = 384 x 256 : frequency x phase, FOV = 22 cm, slice thickness = 3.5 mm, signal averages = 1, echo train length = 23) multi echo spin echo images were acquired in both axial and sagittal planes from two normal volunteers. The array was placed underneath the necks of the volunteers resulting in physical coverage of the array from skull base to approximately the C7 vertebral body. The coil was wrapped around the sides of the neck of each volunteer and reached approximately mid line for both.
1. K.P. McGee, Y. Hu, E. Tryggestad, D. Brinkmann, B. Witte, K. Welker, A. Panda, M. Haddock, M.A. Bernstein, "MRI in radiation oncology: Underserved needs," Magn Reson Med 75, 11-14 (2016).
2. P.B. Roemer, W.A. Edelstein, C.E. Hayes, S.P. Souza, O.M. Mueller, "The NMR phased array," Magnetic Resonance in Medicine 16, 192-225 (1990).