Robert J Holtackers1, Amedeo Chiribiri1, David M Higgins2, and Rene M Botnar1
1Division of Imaging Sciences and Biomedical Engineering, King's College London, London, United Kingdom, 2Philips, Guildford, United Kingdom
Synopsis
Late
gadolinium enhanced (LGE) MRI often suffers
from poor scar-to-blood contrast when used for detection of endocardial scar due to the bright signal of adjacent blood. We
report a method that significantly reduces left ventricular blood signal by
setting a shorter inversion time in combination with a phase-sensitive
inversion recovery (PSIR) sequence. Nulling the left ventricular blood signal with
PSIR significantly increases scar-to-blood contrast since blood signal and scar
signal no longer have similar signal levels. As no additional magnetization
preparation is used, clinical application on current MR systems is readily
available without the need for software modifications or additional training.
Purpose
Late gadolinium enhancement (LGE) has become
the gold standard in the assessment of myocardial viability, as LGE is able to
excellently depict myocardial infarction (MI) and macroscopic scarring from viable
myocardium[1]. However, due to the bright signal of adjacent blood
within the left ventricle (LV), the apparent volume of the damaged myocardial
tissue can be significantly reduced, or even obscured. Obscuration particularly
occurs in cases of thin subendocardial scarring, mostly caused by coronary
artery disease and subsequent MI. In order to allow improved expert confidence
in assessing the presence of scar and its transmurality, efforts have been made
to reduce blood signal using various additional magnetization preparation
schemes[2-5]. We report a method that significantly reduces LV blood
signal in LGE, without using additional magnetization preparation, which is
readily available on current MR systems.
Phase-sensitive inversion recovery (PSIR) is
commonly used for LGE image acquisition as it avoids the need for precise
selection of the inversion delay time (TI) to null viable myocardium[6].
PSIR also shows the recovered longitudinal magnetization (Mz) differently in
the corrected real image produced for clinical assessment: -Mz is darkest,
nulled tissue is mid-gray, and +Mz is bright; whereas in a magnitude image
nulled tissue is darkest and both -Mz and +Mz are bright. PSIR is used routinely
with nulled viable myocardial tissue, and the clinical observer may adjust
window levels to further darken viable myocardial tissue, mimicking a magnitude
image representation. However, the PSIR gray scale range provides an
opportunity to achieve a darker blood signal whilst preserving bright scar
signal by setting a shorter TI such that the LV blood signal is close to the null
point of recovery.
Methods
Imaging was performed on nine male patients (63±11y) with ischaemic myocardial scar at 1.5T (Ingenia; Philips, The Netherlands) 15-33 minutes
after intravenous injection of 0.2 mmol/kg gadobutrol (Gadovist; Bayer, Germany).
A Look-Locker sequence was performed first to identify the two TIs at which
nulling of the viable myocardium and LV blood signal occurs. Subsequently, single
slice breath-hold LGE images were acquired at those two TIs, in a randomized
order, using a PSIR turbo field echo pulse sequence (TE/TR 3.0/6.1 ms, flip
angle 25°
(PSIR 5°), FOV
350x350 mm, slice thickness 10 mm, acquisition matrix 220x170, reconstructed
voxel size 0.91x0.91 mm2, 19 lines acquired every other RR-interval)
during the mid-diastolic resting period. A dedicated noise scan without any RF
excitations was performed to measure the noise level. The applied scaling in the
stored DICOM data was removed by converting the data to floating point values as
this reflects the true MR signal range directly after reconstruction. Regions
of interest were drawn in the viable myocardium, infarcted myocardium (scar),
and the LV blood pool in each image by an expert observer (>10y cardiovascular
MRI experience) who was blinded to image type. Signal-to-noise ratios (SNRs)
and contrast-to-noise ratios (CNRs) were calculated. CNRs were statistically compared
using a paired sample t-test and visualized using a Bland-Altman bias plot. A
Shapiro-Wilk test was performed to confirm normality of data. Bloch simulations
were performed taking into account all sequence details such as start-up echoes,
reference acquisitions, and flip angle sweeps.Results
Imaging at both inversion times (LV blood nulling
and viable myocardium nulling) was performed successfully in all patients at 22±4.6 min and 23±6.2 min
post-injection, respectively (fig. 1 & 2). The inversion time for LV blood nulling and viable myocardium nulling was
168±28 ms and 259±25 ms, respectively. The scar-to-blood contrast in
the PSIR images was significantly (p<0.001) increased when nulling blood
instead of viable myocardium while a change in scar-to-myocardium contrast was
not detected (p=0.15) (fig. 3). Bland-Altman analysis confirmed these findings
by demonstrating a significant bias of +4.37 (95% CI [+2.53, +6.22], fig. 4) in scar-to-blood
CNR, which is a 99% increase when nulling LV blood. The Shapiro-Wilk test confirmed
normality of data. Bloch simulations confirmed the nulling of viable myocardium
and LV blood when using corresponding sequence and tissue parameters (fig. 5).Discussion
Nulling viable myocardium for PSIR LGE is a
continuation of routine clinical practice from the time non-phase-sensitive (magnitude) images were used for LGE. Nulling the LV blood signal instead for PSIR significantly
increases scar-to-blood contrast since blood signal and scar signal no longer have
similar signal levels. We introduced a novel method that allows visualization
of contrast-enhanced tissues while suppressing the blood pool, thereby
improving subendocardial scar conspicuity in PSIR LGE. As no additional magnetization
preparation is used, clinical application on current MR systems is readily
available without the need for software modifications or additional training.Acknowledgements
References
[1] Kellman et al. Cardiac Imaging Techniques for Physicians: Late Enhancement. JMRI. 2012;36:529-542
[2] Peel et al. Dual Inversion-Recovery MR Imaging Sequence for Reduced Blood Signal on Late Gadolinium-enhanced Images of Myocardial Scar. Radiology. 2012;264:242–249
[3] Liu et al. Improved Delayed Enhanced Myocardial Imaging With T2-Prep Inversion Recovery Magnetization Preparation. JMRI. 2008;28:1280-1286
[4] Basha et al. Black Blood Late Gadolinium Enhanced (BB-LGE) using a joing T2 Magnetization Preparation and Inversion Preparation. Proc Intl Soc Magn Reson Med. 2015;23:0184
[5] Kim et al. Black-Blood Contrast-Enhanced MRI: Validation of a Novel Technique for the Diagnosis of Myocardial Infarction. Proc Intl Soc Magn Reson Med. 2015;23:0662
[6] Kellman et al. Phase-Sensitive Inversion Recovery for Detecting Myocardial Infarction Using Gadolinium-Delayed Hyperenhancement. MRM. 2002;47:372-383