Shengzhen Tao1, Paul T Weavers1, Joshua D Trzasko1, Yunhong Shu1, Erin M Gray1, John Huston III1, and Matt A Bernstein1
1Radiology, Mayo Clinic, Rochester, MN, United States
Synopsis
Fast-spin-echo
(FSE) acquisitions are routinely used in clinical MRI, but can be affected by concomitant
field (CF)-induced phase errors. The conventional whole-body MR gradient typically
employs symmetric design. On such systems only CFs of 2nd-order
spatial dependence are significant. These CFs can cause ghosting in large-FOV FSE
acquisition, but are typically negligible over a brain scan volume. Recently, a
high-performance, asymmetric gradient system was developed whose CF contains
additional zeroth and first-order spatially-dependent fields. Here, we investigate the effect of CF
in FSE on this system using extended-phase-graphs simulation, and demonstrate a
real-time compensation for them.
Purpose
Carr-Purcell-Meiboom-Gill (CPMG) condition-compliant fast
(or turbo)-spin-echo (FSE) acquisitions are routinely used in clinical MRI, but
can be prone to phase errors from various field perturbations including eddy
current (EC) and concomitant fields (CF)1-4. The conventional
whole-body gradient system typically utilizes symmetric design, and their CF
contains only fields of 2nd-order or higher spatial dependence4.
These CFs were shown to cause ghosting in FSE images with a large FOV, but their
effect is small in FOVs used for brain scanning with the gradient specification
available on most whole-body scanners3. Recently, a high-performance,
asymmetric MRI gradient was developed. This system is capable of 80mT/m
gradient amplitude and 700T/m/s slew rate with greatly reduced peripheral-nerve-stimulation
risk due to its compact size (26-cm diameter-spherical-volume) designed for brain,
infant, and extremity imaging5. The high-performance gradient can substantially
shorten echo-spacing in FSE and improve imaging performance. However, due to its
asymmetric design, the CFs of this system contains terms of zeroth and first-order
spatial dependence, in addition to the second-order terms on symmetric whole-body
gradients6. These additional CFs can cause prominent artifacts even within
a standard brain scan FOV (~18-24cm).
Here, we investigate the effect of CF in FSE on this system using extended-phase-graphs
(EPG) simulation, and demonstrate their correction using a real-time compensation. Methods
The signal strength of a train of echoes in a FSE sequence
was simulated using EPG7-9 with acquisition parameters detailed in
Table1. Only the effect of phase-encoding (PE) gradient was simulated due to its
strong amplitude on our system (78.6mT/m), and the fact that the readout
gradient is symmetric along refocusing pulse and is not expected to cause significant
CF artifact4. A sagittal scan with PE gradient on the physical y-axis
was chosen as the CF artifacts are most prominent in this configuration. Simulation
was performed including all CF terms (0th, 1st, 2nd),
only 1st and 2nd-order, only 0th and 2nd-order,
and only 2nd-order, respectively, to separate the effects from
different terms. In addition, the American College of Radiology (ACR) phantom and
the brain of a healthy volunteer were scanned under an IRB-approved protocol (Table1).
The real-time gradient pre-emphasis-based method10 was utilized to compensate
for the first-order CF. A real-time, zeroth-order CF compensation was performed
by adjusting the RF transmitter/receiver frequency according to the applied gradient
waveforms, similar to real-time B0 EC correction11. These
compensations were switched on and off to observe effects from different terms.
The effect of an optional zeroth/linear EC correction (ECC) module (“phase
correction”) available on GE’s system12 was also tested and its
interaction with CFs was examined. Results
Figure1 shows an example of echo signal strength along the
echo train in a FSE acquisition obtained from simulation. The 2nd-order
CF has negligible effect, while the 0th/1st-order CFs
cause considerable signal loss along the echo train even at 5cm away from
isocenter. Figure2 shows the pixel intensity maps across the 26cm FOV obtained
from simulation after zeroth and/or first-order CF correction (CFC) (see
captions). The value of each position was calculated from the peak of point-spread-functions
separately computed after weighting the k-space by the echo signal strength
along PE direction. The phantom images acquired with matching parameters show
good correlation with simulation. The slight tilting of black band in phantom images is
suspected to be due to EC effect. The black band and signal loss are largely
reduced after 0th/1st-order CFC. Figure3 demonstrates the
interaction between the standard, commercially-available ECC and the 0th/1st-order
CFs. ECC causes image degradation if the CFC is not performed (Figure3a vs 3b).
Figure4 shows in vivo examples confirming these observations (see captions). Discussion
The
zeroth and first-order CF on the asymmetric gradients can cause considerable
signal loss along echo train in a FSE acquisition. These CFs manifest as local
black band or global signal loss, and their effect is expected to be strong for
long TE acquisition, as large phase error generated from high-amplitude PE
gradients during early echoes can propagate throughout the echo train. We also
note interaction between the standard ECC and 0th/1st-order
CFs. This is expected as their phase accumulations can add on each other and
change the appearance of artifacts. Conclusion
We analyze the effect of the zeroth and first-order
CFs from asymmetric gradients in a FSE acquisition assuming satisfied CPMG conditions,
and demonstrate that their correction can improve image quality acquired on
such systems. The effects of the zeroth/first-order CFs are
relatively well-known for EPI and spiral acquisitions. This work demonstrates
the importance of real-time, zeroth and first-order CF compensation
when using asymmetric high-performance gradients even for routine anatomical
imaging.Acknowledgements
This work was supported in part by the NIH grant R01EB010065.References
1. Hennig J, Nauerth A, Friedburg H. RARE imaging: A fast
imaging method for clinical MR. Magn Reson Med 1986;3:823-833.
2. Bernstein MA, King KF, Zhou XJ. Handbook of MRI pulse sequences.
Burlington, MA: Elsevier Academic Press; 2004.
3. Zhou XJ, Tan SG, Bernstein MA. Artifacts induced by
concomitant magnetic field in fast spin-echo imaging. Magn Reson Med
1998;40:582–591.
4. Bernstein MA, Zhou XJ, Polzin JA, King KF, Ganin A, Pelc
NJ, Glover GH. Concomitant gradient terms in phase contrast MR: analysis and
correction. Magn Reson Med 1998;39:300–308.
5. Lee SK, Mathieu JB, Graziani D, et al. Peripheral nerve
stimulation characteristics of an asymmetric head-only gradient coil compatible
with a high-channel-count receiver array. Magn Reson Med 2015; doi:
10.1002/mrm.26044.
6. Meier C, Zwanger M, Feiweier T, Porter D. Concomitant
field terms for asymmetric gradient coils: consequences for diffusion, flow,
and echo-planar imaging. Magn Reson Med 2008;60:128–134.
7. Hennig J. Multiecho imaging sequences with low refocusing
flip angles. J Magn Reson 1988;78:397-407.
8. Hennig J. Echoes - how to generate, recognize, use or
avoid them in MR-imaging sequences. Part I: Fundamental and not so fundamental
properties of spin echoes. Concepts in Magnetic Resonance 1991;3:125-143.
9. Hennig J, Weigel M, Scheffler K. Calculation of flip
angles for echo trains with predefined amplitudes with the extended phase graph
(EPG)-algorithm: principles and applications to hyperecho and TRAPS sequences.
Magn Reson Med 2004;51:68-80.
10. Tao S, Weavers PT, Trzasko JD, Shu Y, Huston J, Lee SK,
Frigo LM, Bernstein MA. Gradient pre-emphasis to counteract first-order
concomitant fields on asymmetric MRI gradient systems. Magn Reson Med 2016;DOI:
10.1002/mrm.26315.
11. Crozier S, Eccles CD, Beckey FA, Field J, Doddrell DM.
Correction of eddy-current-induced B0 shifts by receiver reference-phase
modulation. J Magn Reson 1992;97:661–665.
12. Hinks RS, inventors; General Electric Company, assignee.
Fast spin echo prescan for MRI system. US patent, 5,378,985, 1995.
13. Stanisz GJ, Odrobina EE, Pun J, Escaravage M, Graham SJ,
Bronskill MJ, Henkelman RM. T1, T2 Relaxation and Magnetization Transfer in
Tissue at 3T. Magn Reson Med 2005;54:507-512.