Franz Patzig1, Toralf Mildner1, and Harald Möller1
1Max Planck Institute for Human Cognitive and Brain Sciences, Leipzig, Germany
Synopsis
Geometric distortions caused by magnetic susceptibility variations in
the underlying medium can severely corrupt the image. A novel correction method
is proposed, which uses the prior knowledge of the analytic point spread
function (PSF) of the used imaging sequence and a map of the underlying field inhomogeneities.
From this input, a PSF operator can be devised and applied
to correct the image by performing a deconvolution. Regularization
techniques are used to improve and stabilize the outcome. A significant
reduction in geometric distortions is demonstrated for human brain images as
well as some advantages over existing correction methods.
Purpose
Single-shot acquisition techniques like EPI are the workhorse for functional
and diffusion MRI. Due to their long read-out trains and concomitantly low
bandwidth in phase-encoding (PE) direction, they are particularly prone to
artifacts caused by magnetic susceptibility variations in the object introducing
additional phase offsets. This problem becomes more severe at high magnetic
fields. Correction methods often require additional acquisitions, such as a map
of the underlying field inhomogeneities that is used in the multi-frequency
reconstruction (MFR) approach [1,2] or the experimental point-spread function (PSF)
in each voxel used in a PSF correction approach [3]. While the latter technique
is expected to produce better outcomes, it is also more time consuming. Here,
we present a technique combining a field inhomogeneity map with an analytic PSF
to correct EPI images by a deconvolution technique.Methods
Analytic complex PSFs were obtained for single-shot GE-EPI and a
recently suggested double-shot modification with center-out trajectories (DEPICTING
[4]) by applying the inverse Fourier transform to the modulation transfer
functions in k-space. The PSFs are deduced with a simplified model considering $$$T^*_2$$$ relaxation, main magnetic field inhomogeneities, $$$\Delta B_0$$$, and truncation of k-space
by the acquisition window. They were validated by experiments with water phantom
scans and by simulations (Figure 1). All acquisitions including EPI and
DEPICTING scans in human volunteers as well as a multi-echo-FLASH based
field-map scan were performed at 3T (Verio, Siemens, Erlangen, Germany).
Any measured image is a convolution of the spatially varying PSF and the
underlying object, namely $$$\mbox{IMG} = \mbox{PSF} \ast \mbox{OBJ}$$$. Using an acquired $$$\Delta B_0$$$ map and the analytic PSF of the particular
sequence as input, the PSF operator can be modeled and a deconvolution in image
space with the discrete PSF operator yields a representation of the object that
is distortion- and intensity-corrected. Only effects along PE direction are
considered as they are known to dominate [5]. The deconvolution is performed by
inverting a PSF matrix for every column of the image. This is an ill-posed
problem and requires further regularizations, which is based on singular value
decomposition (SVD). Briefly,
the least-squares solution of a system of linear equations is obtained by the
Moore-Penrose pseudoinverse, which can be computed inverting the SVD matrices.
The inversion of low singular values was regularized as this is an unbounded
operation. Furthermore, in regions of
low signal intensity the field map was interpolated employing discrete cosine
transforms (see figure 2).Results
Figure
3 shows an uncorrected EPI image (A), and EPI images corrected by both the MFR
method (B) and the analytic PSF approach (C). In general, both methods are
capable of correcting the major distortion artifacts. The PSF-based correction
technique, however, additionally achieves a restoration of the signal intensity
distribution and has the potential to reproduce brain structures more clearly
by an inherent consideration of $$$T^*_2$$$ blurring. This is further illustrated in Figure
4, which shows results for the DEPICTING sequence [4], where off-resonance
effects are known to cause simultaneous shifts in opposite directions due to
the double-shot acquisition with opposite phase blips. Figure 5 demonstrates
the effect of applying the regularization strategies necessary for the PSF
deconvolution. While the corrected
image is highly corrupted if only the inverse of the PSF matrices is used (Fig.
5B), the quality is substantially improved employing the SVD regularization (Fig.
5C) and even more after interpolating the field map (Fig. 5D).Discussion
A distortion
correction of EPI-based images was proven to be feasible by the knowledge of an
analytical solution of the imaging PSF, a preparation scan ($$$\Delta B_0$$$/$$$T^*_2$$$ map) as
input, and successive PSF deconvolution. The method yielded improved results when
compared to the MFR method (Figures 3 and 4). In addition, computation times
are reduced for the PSF-based technique. The quality of the obtained corrections
also substantiates that our analytic model of the PSF approximates the real
PSFs quite well and considers the relevant contributions. However, in order to
stabilize the deconvolution procedure towards a meaningful result, a
regularization strategy was required (Figure 5). As in the MFR approach, the quality
of the proposed PSF deconvolution depends on the quality of the underlying
field maps, that is, deviations between the $$$\Delta B_0$$$ distributions used for the
correction and that present during EPI need to be minimized.Conclusion
A novel method for EPI image distortion correction based on analytic PSFs was introduced. Compared to the MFR technique, an improved
image contrast with an inherent intensity correction was achieved. The method might be an alternative to the direct voxel-by-voxel
estimation of the PSF [3].Acknowledgements
No acknowledgement found.References
[1] D.C. Noll et al., A Homogeneity Correction Method
for Magnetic Resonance Imaging with Time-Varying Gradients. IEEE Trans. Med.
Imaging 10: 629-637, 1991.
[2] L.C. Man et al., Multifrequency Interpolation for
Fast Off-resonance Correction. Magn. Reson. Med. 37: 785-792, 1997.
[3] H. Zeng, R.T. Constable. Image Distortion
Correction in EPI: Comparison of Field Mapping With Point Spread Function
Mapping. Mag.
Reson. Med. 48: 137-146, 2002.
[4]
S. Hetzer, T. Mildner, H.E. Möller. A Modified EPI
Sequence for High-Resolution Imaging at Ultra-Short Echo Time. Mag. Reson. Med.
65: 165-175, 2011.
[5] P. Jezzard, R.S. Balaban. Correction for Geometric
Distortions in Echo Planar Images from B0 Field Variations. Mag. Reson. Med.
34: 65-73, 1995.