Maria Isabel Vargas1, Bénédicte MA Delattre1, José Manuel Baiao Boto1, Karl-Olof Lövblad1, and Sana Boudabbous1
1Geneva University Hospitals, Geneva, Switzerland
Synopsis
Synthetic
MRI was already validated in some cerebral applications. We adapted it to evaluate
the spine and showed that it produces, in general, good image quality and
diagnostic confidence. Efforts will still have to be made to increase image
quality in the dorso-lumbar spine. However, the non-negligible time saving and
the ability to obtain quantitative measurements as well as to generate several
contrasts with a single acquisition should secure an interesting and useful
future for synthetic MRI in clinical routine.
Introduction
Synthetic
MR is a method that produces multiple contrasts as well as quantitative T1, T2
and PD images and that has already been validated in cerebral applications1,2.
The goal of this study was to implement synthetic MR in the spine and to
evaluate its diagnostic confidence compared to conventional sequences.Material and methods
Thirty-one
patients underwent MRI for spine examination (cervical, lumbar or whole spine)
on a Philips Ingenia 1.5T. Conventional T1 TSE, T2 TSE and STIR TSE sequences
were performed as well as the syntac sequence. The syntac sequence is based on
a TSE sequence with a saturation pulse (120°), 4 different inversion times and
2 echo times, which leads to 8 images with different contrasts. These images
are then used by SyMRI v8 software (SyntheticMR AB, Linköping, Sweden) to
generate T1, T2 and PD quantitative images and then synthetically create T1, T2
and STIR contrasts with user chosen TE/TR/TI. Sequence parameters were:
sagittal orientation, FOV 200x321 mm, acquisition (reconstruction) voxel size
0.89x1.48 (0.71x0.72) mm, 15 slices of 4 mm thickness with no gap, TE 11 / 100
ms, TR 2485 ms, TSE factor 12, SENSE acceleration factor 2. The acquisition
time of the synthetic sequence is 5min 40sec. Synthetic images were generated
with matching TE/TR parameters and compared to the conventional sequences (T1,
T2, STIR when available). The TI was chosen to suppress as much as possible the
signal from fat.
For each patient, two sets of images were created (conventional
and synthetic sequences). Random analysis of the sets was performed by a blinded
neuroradiologist who graded the image quality (non-diagnostic, medium, good),
the presence of artifacts and identified the different pathologies.Results
All synthetic images were graded as medium (6.5%)
to good quality (93.5%), none was rated as non-diagnostic quality. Conventional
images were all graded as good quality. The diagnosis was concordant in 87.1%
of cases, which means the lesion was identified with the same confidence in
Synthetic as in conventional images. Figure 1 shows an example of a syrinx that
is clearly visible in synthetic images. The diagnosis was discordant in 3.2%
(1/31) and uncertain in 9.7% (3/31) of cases. Concerning the presence of
artifacts, a majority (84%) of synthetic STIR presented flow artifacts and / or
noise compared to conventional images. In addition, 19.4% of T1 and T2 images
demonstrated either flow artifacts or a noisy appearance. In general, the
images were perceived as of better quality with a higher signal to noise ratio
(SNR) in the cervical region due to the presence of the head-neck coil compared
to dorsal or lumbar regions. Figure 2 shows an example of synthetic images in
postoperative evaluation of the lumbar spine.Discussion
The syntac sequence
was adapted to evaluate the spine and can replace at least 3 conventional sequences
(T1, T2, STIR), that usually require 13min 30sec of acquisition time, and
offers the possibility to generate other contrasts if required. The main current
limitations of this sequence are a certain degree of SNR reduction in the dorso-lumbar
spine and the presence of vascular flow artifacts, particularly in STIR images.
In the cases where the diagnosis was not concordant, the images were affected
by artifacts. A compromise still has to be found to increase the SNR while maintaining
a reasonable acquisition time in the dorso-lumbar spine. However, in our study the
image quality and diagnostic confidence are still very good, especially in the cervical
region where the syntac sequence currently produces the best results. It also
allows the generation of quantitative T1, T2 and PD images (Figure 3), which
are of great interest since they are expected to produce more accurate and
relevant results as suggested by a recent publication of the European society
of radiology3. The next step of this study is to evaluate the added
value of quantitative T1/T2 measurements in the interpretation of the lesions.
We believe that our findings can help in the integration of quantitative MRI in
clinical routine, notably in the context of development of new methods such as fingerprinting.Conclusion
Our study
shows that Synthetic MRI is feasible for spinal imaging and produces, in
general, good image quality and diagnostic confidence. Efforts will still have to
be made to increase image quality in the dorso-lumbar spine. However, the
non-negligible time saving and the ability to obtain quantitative measurements
as well as to generate several contrasts with a single acquisition should secure
an interesting future for synthetic MRI in clinical routine.Acknowledgements
No acknowledgement found.References
1.
Blystad I, Warntjes JB, Smedby O, et al. Synthetic MRI of the brain in a clinical setting. Acta
Radiol. 2012 Dec 1;53(10):1158-63.
2.
Granberg, T, Uppman, M, Hashim, F, et al. Clinical Feasibility of Synthetic MRI
in Multiple Sclerosis: A Diagnostic and Volumetric Validation Study. American Journal of Neuroradiology,
2016 ; 1–7.
3.
European Society of Radiology (ESR). Magnetic Resonance Fingerprinting - a
promising new approach to obtain standardized imaging biomarkers from MRI. Insights
into Imaging, 2015, 6(2), 163–165.