Amir Ali Rahsepar1, Rozann Hansford1, Michael A. Guttman1, Valeria Castro1, Diana McVeigh1, Stefan Zimmermann1, John Kirsch2, Henry R. Halperin1, and Saman Nazarian1
1Johns Hopkins University, Baltimore, MD, United States, 2Siemens Medical Solutions
Synopsis
This large prospective study showed that the
previously suggested SAR limit of 2 W/kg can be unduly restrictive. No other MRI characteristics appear to predict changes in device
system parameters.
Introduction
Magnetic
resonance imaging (MRI) is considered as the modality of choice in many
clinical situations. Unfortunately, MRI is currently unavailable for a large number
of patients who have implanted cardiac devices such as pacemakers, implantable cardioverter defibrillators (ICD) and cardiac resynchronization therapy (CRT) devices. Associated risks with performing
MRI arise from the interaction between the implanted device system and three
essential components of MRI which are the static and gradient magnetic fields, and
radiofrequency (RF) magnetic field. There have been tremendous efforts to
develop MRI safety protocols to reduce complications in patients with cardiac
devices (1-3). Radiofrequency energy absorption may lead to heating
in biologic tissue. Moreover, presence of conductive implants in the body raises
the concern of RF heating in leads and tissues surrounding such devices. Thermal
injury and the subsequent formation of fibrosis may result in increasing capture
thresholds, and deterioration of device function. Specific absorption rate
(SAR) and specific energy dose (SED) are the dosimetric terms used to
characterize the thermogenic aspects of the electromagnetic field. In study we
aimed to explore the effect of MRI field components (gradients and RF magnetic
fields), region of imaging, and cardiac device characteristics to determine
which of these factors has detrimental effects on device parameters. Methods
1,467 patients [age: 66.71±14.98yrs, Male: 930
(63.4%)] who were referred to undergo
medically necessary MRI were prospectively enrolled in this study. A total of 2,027 MRI examinations were performed in these patients with
a total 3,832 leads using
clinical MRI protocols without SAR restrictions. To evaluate appropriate device function atrial
and ventricular sensing, lead impedance, and capture threshold were measured
before and immediately after MRI examination. Two experienced investigators
evaluated the device and lead variables. Imaging was performed at 1.5 Tesla using
an Avanto scanner (Siemens, Erlangen, Germany). Magnetic resonance imaging was
performed according to standard institutional protocols for the region of
interest using usual protocols with standard peak SAR
settings for the scan. Whole body averaged SAR was reported based upon the MR
sequence parameters, applied voltages, the transmitter reference adjustment and
the patient weight. The data regarding the gradient slew rate for y-gradient,
x-gradient and z-gradient coils in addition to the scan duration time for each
sequence were recorded from the scanner.Results
876 patients (59.71%) had a permanent pacemaker,
436 (29.72%) had an ICD, and 155 (10.56%) had a CRT system. Among the patients,
123 (8.4%) were device dependent. Lead length, MRI regions and device
manufacturers have been summarized in table 1. The average number of acquired
sequences during each scan was 18±12.6. The values for peak SAR, peak SED,
during MRI scan of different regions have been summarized in table 2. Following
the previously described safety protocol, no adverse events including device
movement, torque, heating, arrhythmia, or unexpected device activation were
reported during the scans and all devices were functioning properly following scan. Comparison of device interrogation
findings obtained before, immediately and 6 months after MRI showed no
significant changes in battery voltage, lead thresholds, lead impedances, or
sensing signal amplitudes. In multivariable analysis, no associations between
changes in device and lead parameters and scanned region, scan duration, SAR or
dB/dt. However, changes in device right ventricular (RV) sensing and capture thresholds were
associated with RV lead length (P<0.05).Discussion
To
the best of our knowledge, this is the first prospective study to evaluate the
association between cardiac device parameters and MRI field and implantable
device characteristics. The main finding of the current study is that MRI
characteristics including dB/dt, SAR, SED, and scan duration were not
associated with adverse device changes. These findings are crucial particularly
in patients with implanted legacy devices, as the need to scan them will
continue for many years. The current results show that the previously suggested
SAR limit of 2 W/kg can be unduly restrictive. Additionally, no other MRI
characteristics appear to predict changes in device system parameters. However,
the association of lead length with parameter changes warrants further study.Acknowledgements
The study was funded by NIH grants K23HL089333
and R01HL116280 to Dr. Nazarian.References
(1) Nazarian S, et al. Clinical utility and safety of a protocol
for noncardiac and cardiac magnetic resonance imaging of patients with
permanent pacemakers and implantable-cardioverter defibrillators at 1.5 tesla.
Circulation 2006 Sep 19;114(12):1277-84.
(2) Nazarian S, et al. A prospective evaluation of a protocol
for magnetic resonance imaging of patients with implanted cardiac devices. Ann
Intern Med 2011 Oct 4;155(7):415-24.
(3) Roguin A, et al. Modern pacemaker and implantable
cardioverter/defibrillator systems can be magnetic resonance imaging safe: in
vitro and in vivo assessment of safety and function at 1.5 T. Circulation 2004
Aug 3;110(5):475-82.