Emilie Bollache1, Alex J. Barker1, Jeremy D. Collins1, Pim van Ooij2, Rouzbeh Ahmadian1, Alex Powell1, James C. Carr1, Julia Geiger1, and Michael Markl1,3
1Department of Radiology, Northwestern University, Chicago, IL, United States, 2Department of Radiology, Academic Medical Center, Amsterdam, Netherlands, 3Department of Biomedical Engineering, Northwestern University, Chicago, IL, United States
Synopsis
Our
objective was to assess the performance of eight highly k-t accelerated non-gated
free-breathing aortic 4D flow MRI measurements acquired in under 2 minutes
(PEAK GRAPPA R=5; TRes=67.2ms; four ky-kz-space Cartesian
fillings: linear, center-out, out-center-out, random; two spatial resolutions=3.5x2.3x2.6mm3,
4.5x2.3x2.6mm3), both in vitro
and in 10 healthy volunteers. Despite lower image quality, the significantly
shorter k-t accelerated datasets provided aortic hemodynamic indices in
agreement with conventional respiratory-gated 4D flow measurements. Differences
were non-significant when using linear and out-center-out k-space samplings (absolute
differences≤22%). In conclusion, aortic 4D flow MRI in under 2 minutes is
feasible with moderate underestimation of flow indices.
INTRODUCTION
4D flow MRI is often limited by scan times, which are rendered even
longer with respiratory navigator gating1, commonly used for
thoracic applications to correct for breathing-related motion. Thus, typical
scan times for an aortic 4D flow exam range from 8 to 12 minutes when using
regular parallel imaging (acceleration factor R=2). The purpose of this study
was to assess the feasibility of performing aortic 4D flow MRI in under 2
minutes by combining k-t acceleration, free breathing data acquisition using
different ky-kz sampling patterns (linear, centric,
random) to mitigate breathing artifacts, and adapted spatial/temporal
resolutions.METHODS
Eight variants of k-t accelerated non-gated 4D flow MRI (PEAK GRAPPA2
R=5) using 2 spatial resolutions (SRes1 and SRes2) and 4 ky-kz
Cartesian sampling patterns (Figure 1.a) were implemented: 1- conventional linear
line-by-line ky-kz-space filling; 2- centric encoding from the central (ky=kz=0) to the outer k-space positions (‘center-out’); 3- most outer ky-kz-space
positions (k-space corners) initially filled during the first 10 cardiac
cycles, followed by centric encoding (‘out-center-out’); 4- random filling. The
8 acquisition strategies were compared to conventional navigator-gated 4D flow
MRI, which was performed according to consensus recommendations3 (Figure
1.b), using navigator gating of the lung-liver interface with a fixed 16-mm
acceptance window size combined with respiratory-ordered phase encoding4.
The evaluation was performed on a 1.5T MAGNETOM Aera scanner (Siemens, Germany)
both in vitro, on a patient-specific
3D-printed aortic coarctation model connected to a pulsatile pump (Figure 2.a),
and in 10 healthy volunteers (4 women, age: 61±16 [31-77] years). Data analysis
included visual image quality assessment on a 3-point scale in the ascending
aorta (AA), arch and descending aorta (DA) based on the 4D flow magnitude cine
data and the PC-MRA sagittal maximal intensity projection (MIP). Flow indices (regional
peak velocity Vmax and flow Qmax as well as net flow volume Qnet) were further
computed as previously described5,6.RESULTS
Conventional and the 8 k-t accelerated 4D flow MRI datasets were
successfully acquired on the phantom and 10 volunteers (total n=99 datasets). In vitro scan time was 9:27, 2:16 and
1:53 minutes for the conventional and the k-t accelerated acquisitions with SRes1
and SRes2 spatial resolutions, respectively. k-t accelerated aortic Qmax, Qnet
and Vmax were lower than the reference conventional 4D flow indices by ≤4.7%,
≤11%, and ≤22%, respectively, with highest differences observed in the DA
coarctation (Figure 2.b). In vivo k-t
accelerated acquisitions were significantly shorter than conventional 4D flow (Table
1.a, p=0.002). Figure 3 illustrates examples of 4D flow magnitude images and 3D
PC-MRA MIPs. Image quality was overall reduced for k-t accelerated data when
compared against conventional 4D flow (Table 1.b). Finally, results of regional
flow quantification are summarized in Table 1.c. Examples of in vivo peak systolic velocity MIPs are
shown in Figure 4, along with cohort-averaged AA and DA flow waveforms. A
Bland-Altman analysis revealed overall lowest biases compared to conventional
4D flow MRI when using the out-center-out ky-kz sampling
pattern (mean biases [limits of agreement]: from -6.3 [-39;26] to 7.4 [-27;41]
cm/s for Vmax, -12 [-73;49] to -8.9 [-120;103] ml/s for Qmax, -2.7 [-13;7.2] to
1.1 [-13;15] ml for Qnet).DISCUSSION
Our main findings were that: 1) aortic 4D flow under 2 minutes is feasible
but can lead to a moderate underestimation of hemodynamic indices and
particularly of peak systolic velocity; 2) free breathing without controlling
for respiratory motion reduces image quality but flow and velocity indices for
specific k-space ordering schemes (linear, out-center-out) were close to those
obtained using conventional navigator-gated 4D flow; 3) the observed
differences in k-space sampling patterns suggest an opportunity to mitigate
some image artifacts and velocity errors by adapting k-space filling. A
limitation of our work is the lack of inter-exam variability or patient data;
however, recruitment of patients is currently under progress to confirm if our
findings are still valid in case of complex flow dynamics with high velocity
and acceleration. Other previously described acceleration7-9 and
respiratory control10 methods should also be explored in future
studies to help defining the best compromise between short scan time and
breathing motion correction.CONCLUSION
Aortic 4D flow MRI in under 2 minutes is feasible and easy to use, with
no navigator placement and set-up required before the actual acquisition,
providing moderate underestimation of flow indices. Differences reported in
this preliminary study on an in vitro
phantom and in healthy volunteers suggest an opportunity to mitigate image
artifacts by an optimal trade-off between scan time, acceleration, and k-space
sampling.Acknowledgements
This work was supported by the National Institutes of Health grants R01HL115828 and K25HL119608 as well as the American Heart Association Midwest Affiliate grant 16POST27250158.References
1.
van Ooij P, Semaan E, Schnell S, et al. Improved respiratory navigator gating
for thoracic 4D flow MRI. Magn Reson Imaging. 2015;33(8):992-9.
2.
Jung B, Ullmann P, Honal M, et al. Parallel MRI with extended and averaged
GRAPPA kernels (PEAK-GRAPPA): optimized spatiotemporal dynamic imaging. J Magn
Reson Imaging. 2008;28(5):1226-32.
3.
Dyverfeldt P, Bissell M, Barker AJ, et al. 4D flow cardiovascular magnetic
resonance consensus statement. J Cardiovasc Magn Reson. 2015;17:72.
4.
Markl M, Harloff A, Bley TA, et al. Time-resolved 3D MR velocity mapping at 3T:
improved navigator-gated assessment of vascular anatomy and blood flow. J Magn
Reson Imaging. 2007;25(4):824-31.
5.
Rose MJ, Jarvis K, Chowdhary V, et al. Efficient method for volumetric
assessment of peak blood flow velocity using 4D flow MRI. J Magn Reson Imaging.
2016 (in press).
6.
Stalder AF, Russe MF, Frydrychowicz A, et al. Quantitative 2D and 3D phase
contrast MRI: optimized analysis of blood flow and vessel wall parameters. Magn
Reson Med. 2008;60(5):1218-31.
7.
Baltes C, Kozerke S, Hansen MS, et al. Accelerating cine phase-contrast flow
measurements using k-t BLAST and k-t SENSE. Magn Reson Med. 2005;54(6):1430-8.
8.
Giese D, Schaeffter T, Kozerke S. Highly undersampled phase-contrast flow
measurements using compartment-based k-t principal component analysis. Magn
Reson Med. 2013;69(2):434-43.
9.
Tariq U, Hsiao A, Alley M, et al. Venous and arterial flow quantification are
equally accurate and precise with parallel imaging compressed sensing 4D phase
contrast MRI. J Magn Reson Imaging. 2013;37(6):1419-26.
10.
Uribe S, Beerbaum P, Sorensen TS, et al. Four-dimensional (4D) flow of the
whole heart and great vessels using real-time respiratory self-gating. Magn Reson
Med. 2009;62(4):984-92.